To print this text file, a 10-point/12-pitch fixed font (e.g. Courier) is best ============================================================================== BULLETIN FSB-104 - HOW TO PASS NBADCP/BBADCP/WORKHORSE RECEIVER TESTS DATE/REV: 15NOV96 / 0 ============================================================================== ABSTRACT: Explains what steps to take to ensure receiver tests do not FAIL because of external interference signals. INSTRUMENT/PROGRAM AFFECTED: All ADCPs (Narrowband, Broadband, Workhorse) Narrowband - Preamp/Receive Test portion of DRTEST/SCTEST/VMTEST program Broadband - Receive-Path Test (PT3), Receive Bandwidth Test (PT6) Workhorse - Pre-deployment Test (PA) DESCRIPTION: Many users tell us their ADCP reports a FAIL condition during the self-tests that check the ADCP's "receive path." In most cases, the cause of the failure is external interference. This external interference is usually caused by a "noisy" environment, such as in a lab. You can take a few simple steps to find out if the FAIL condition is being caused by external interference or by a problem with the ADCP. SOLUTION/ACTION: The following procedure explains how to conduct the tests listed above in the Instrument/Program Affected section to reduce the likelihood of a false failure. 1. Turn off any nearby equipment (monitors, radios, etc.) that is not needed to conduct the test. 2. The pressure case should be on the ADCP to help shield the circuit boards from external electronic "noise." 3. The ADCP transducer head must be immersed in water. Ensure there are little or no air bubbles on the transducer faces. 4. If the transducer is immersed in a bucket of water that is resting on the floor, noise can be coupled into the ADCP. As such, you should shield the bucket from the floor by inserting a piece of hard foam between the bucket and the floor. 5. Run the appropriate receiver test for your ADCP. 6. If the test fails, "spin" the ADCP 180 degrees. If the failure moves from one beam to another, it is "probably" safe to say the failure was caused by external interference, and it should be all right to deploy the ADCP. 7. If possible, you may also want to move the ADCP to a different room, or at least to a different part of the lab to see if the fail condition goes away. 8. If after following the above procedure, your ADCP still fails the receive tests, contact RDI for assistance. NOTE: Although rare, we had a case where a Vessel-Mounted (VM) system was being tested in a lab. The interconnecting VM cable was apparently picking up interference from the floor. When the cable was isolated from the floor by placing it on a chair, the test passed. ==============================================================================