# MSE instrument readiness review # wed 6/14/2006 13:00 # attending oreilly,salamy chaffey,hamilton, graybeal,gomes,mccann,schlining, ussler,barry,barr,paull, reinecker,lovera,chavez #Workhorse ADCP (ID 1525) - Ussler/O'Reilly - summary data not directly usable to sci - instrument clock must be set before deployment * a service that sets clocks or all instruments would be good (barry) - manually issue TS command (no args) to obtain current ADCP time #Seabird SBE-16 plus (ID 1533) - Ussler/Salamy * summary does not incl transmissometer data * need to know analog ch:instr mappings * xml has only one analog channel #Seahorse (ID 1405) - Hamilton/Headley #Aanderaa oxygen sensor (ID 1444) - Barry/Salamy * warmup start delay (1 s) may be too short #WETLabs Triplet (1535) - Lovera/O'Reilly * how long can the instrument remain in safe mode before it starts overwriting data (60 min interval) * should set safemode interval to 5 min * logging stops when internal storage full... * leave internal logging off when not in safemode #Nortek Aquadopp (ID 1474) - Barry/O'Reilly * better to take sample average over longer period * needs to be configured w/ AquaProHR software * limit of 32 files (can't start and stop sampling) * internal logs wraps when full and keeps logging #Power service (ID 1554) - Mellinger/Headley * if we park in linux in safemode, WDT will cause reset w/in 24h and take everything out of safemode. OK if safemode is due to power failure (nodes will be halted when power runs out and no power to restart) #Triaxys (ID 1286) - Hamilton/O'Reilly * check x,y,z offsets for this deployment, may need to change * sample schedule 20 min? takes 25+ min to sample and process data * verify and set 30 min schedule * need to erase mem? no #Medusa (ID 1526) - Chaffey/Headley #MSP430 (ID 1527) - Chaffey/Headley