From: Peltzer, Edward Sent: Wednesday, May 26, 2010 14:21 To: 'Bob Herlien'; Maughan, Thom Subject: FW: [foce] ph sensor calibration complete - the pH cal info Bob and Thom: It took some looking but ... here are the assignments from our previous deployment on 4/26 (FOCE/MARS4). Ed. -----Original Message----- From: Maughan, Thom Sent: Friday, April 02, 2010 11:04 To: Maughan, Thom; Peltzer, Edward; Free Ocean CO2 Experiment Discussion List Cc: Headley, Kent; Gomes, Kevin Subject: RE: [foce] ph sensor calibration complete - the pH cal info Bob, Here are the pH calibration coefficients, along with serial number and SSDS (ISSI) ID's as well as location on the FOCE frame. Please let Ed know when this has been built into the FOCE app. Let me know if I can help. Location Serial # SSDS Calibration ---------- -------- ------ ----------------------------- FOCE pH #1 180358A 1724 slope= 9.5287, offset= 2.6129 FOCE pH #2 180512A 1735 slope= 9.8988, offset= 2.6233 FOCE pH #3 180513A 1722 slope= 9.8170, offset= 2.6404 FOCE pH #4 180514A 1725 slope= 9.7577, offset= 2.6290 FOCE pH #5 180511A 1734 slope= 9.8334, offset= 2.6086 FOCE pH #6 180507A 1750 slope= 9.9380, offset= 2.6287 -------- backup spares, not on FOCE frame -------------- Spare 1 180506A 1749 slope= 9.8211, offset= 2.5793 Spare 2 180508A 1751 slope= 9.8575, offset= 2.6704 Thom -----Original Message----- From: Maughan, Thom [mailto:tm@mbari.org] Sent: Thursday, April 01, 2010 3:48 PM To: Peltzer, Edward; Free Ocean CO2 Experiment Discussion List Cc: Headley, Kent; Gomes, Kevin Subject: RE: [foce] ph sensor calibration complete Ed, I still need to get ISSI id's (those 4 digit numbers for SSDS) for the new sensors. I'll work with PeterW on agreeing to a map for the sensors. I'll work with Tom O on getting the coefficients built into the FOCE application (by tomorrow). I will also write the calibration coefficients into the flash memory on each of the pH sensors as backup traceability. Thom -----Original Message----- From: Peltzer, Edward [mailto:etp3@mbari.org] Sent: Thursday, April 01, 2010 2:18 PM To: Free Ocean CO2 Experiment Discussion List Subject: [foce] ph sensor calibration complete FOCErs: I've completed the number crunching on the HRpH sensors for our next FOCE deployment. They all looked very good and had very fast response times -- got to love new electrodes... Anyway, the calibration constants are attached for those who need to know. Now, does someone have a map of which sensors went where so we know how to assign these constants? Ed. -------------------------------- Dr. Edward T. Peltzer Monterey Bay Aquarium Research Institute 7700 Sandholdt Road Moss Landing, CA  95039-9644 Phone: 831-775-1851 Fax:   831-775-1620 Email: etp3@mbari.org http://www.mbari.org/staff/etp3/ --------------------------------