BRS midweek valve/system test package ===================================== Same OASIS.CFG / STIR.SCP / MODLOG.SCP as BRS_3day_valve_test. Only RESP.SCP differs. Contents (deploy to OASIS top-level filesystem): OASIS.CFG - same as Jun 2026 deployment (valves already defined) STIR.SCP - same as deployment MODLOG.SCP - same as deployment RESP.SCP - midweek schedule + flush-valve open/close + syslog Boot scripts (from OASIS.CFG): script = modlog.scp script = stir.scp script = resp.scp Intended window: Tuesday morning start -> Thursday afternoon recovery (no fixed wall-clock start time) RESP timing (~52 h / ~2.2 days): Start delay: 10 min Cycle interval: 4 h Period 1: 1 flush-only warmup cycle (900 s flush, no inject) Period 2: 12 cycles, O2 250 ml all chambers, 900 s flush Valve syslog (consistent; period banners are separate): open valves N-N close valves N-N Pairs: open/close_1 (ch1-2), _3 (ch3-4), _5 (ch5-6), _7 (ch7-8)