# $Id: FormatNotes,v 1.15 2012/04/27 16:57:51 swift Exp $
#~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
# RCS Log:
#
# $Log: FormatNotes,v $
# Revision 1.15  2012/04/27 16:57:51  swift
# Update the format spec for FLBB, Isus, and Optode.
#
# Revision 1.14  2012/03/28 22:41:58  swift
# Modified format spec to match data as transmitted.
#
# Revision 1.13  2007/07/31 15:20:00  swift
# Modifications to reflect the integration of the ISUS sensor.
#
# Revision 1.12  2007/07/03 22:52:18  swift
# Minor change to facilitate automatic management of FormatNotes.
#
# Revision 1.11  2006/01/30 17:43:05  swift
# Eliminate "dbars" and "C" from the ParkPt entries.
#
# Revision 1.10  2006/01/28 17:11:41  swift
# Integrated the Aanderaa optode into iridium firmware.
#
# Revision 1.9  2006/01/06 23:25:41  swift
# Minor syntax fixes.
#
# Revision 1.8  2005/10/25 17:02:46  swift
# Change message extension from 'prf' to 'msg'.
#
# Revision 1.7  2005/09/05 16:04:16  swift
# Fleshed out formatting documentation for iridium messages.
#
# Revision 1.6  2005/03/30 21:11:09  swift
# Description of the encoding of data.
#
# Revision 1.5  2004/12/29 23:12:17  swift
# Deleted format notes that pertained only to ARGOS firmware.
#
# Revision 1.4  2004/04/10 18:38:27  swift
# Format description for use with 28-bit Argos ids.
#
# Revision 1.3  2004/04/10 18:36:54  swift
# Format description for use with 20-bit Argos ids.
#
# Revision 1.2  2003/11/21 02:46:23  swift
# Added documentation for the SBE41 status buts.
#
# Revision 1.1  2003/11/12 23:05:54  swift
# Initial revision
#
#~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~

Format specification for APF9 firmware revision: $FwName$-$FwRev$ 
===================================================================================

Iridium message files end with a ".msg" extension.  Each iridium message
file consists of blocks of similar data presented in the order that they
were collected during the profile cycle.  This firmware revision includes
five blocks of data:

   1) Park-phase PT samples: These are hourly low-power PT samples collected
      during the park phase of the profile cycle.

   2) Low resolution PTS/Optode/ISUS/FLBB samples: This firmware implements
      low-resolution sampling over the whole water column of the Aanderaa
      Optode oxygen sensor, the MBARI ISUS nitrate sensor, and the Wetlabs
      FLBB sensor.

      In addition, the deep parts of the profile can be represented using
      low-resolution PTS spot samples collected at predetermined pressures.
      Low resolution spot sampling in the deep water was implemented as an
      energy savings measure.

   3) High resolution PTS samples: The shallower parts of the profile can be
      represented with high resolution (ie., 2 decibar) bin-averaged PTS
      samples.  In continuous profiling mode,  the CTD samples at 1Hz and
      stores the data for later binning and averaging.

   4) GPS fixes: After the profile is executed and the float reaches the
      surface, the location of the profile is determined via GPS.

   5) Biographical & engineering data: Various kinds of biographical and
      engineering data are collected at various times during the profile
      cycle.  

Usually, only one telemetry cycle is required to upload the data to the
remote host computer.  However, sometimes the iridium connection is broken
or the quality of the connection is so poor that the float will abort the
telemetry attempt, wait a few minutes, and then try again.  Data blocks 4
and 5 will be repeated for each telemetry cycle of a given profile.

A description of the format for each of these blocks of data follows.


1) Park-phase PT sample format.
-------------------------------

   Hourly low-power PT samples are collected during the park phase of the
   profile cycle.  The park phase is also when active ballasting is done.
   Each sample includes the date & time of the sample, the unix epoch (ie.,
   the number of seconds since 00:00:00 on Jan 1, 1970), the mission time
   (ie., the number of seconds since the start of the current profile
   cycle), the pressure (decibars), and the temperature (C).  For example:

                 |------- date -----|   UnixEpoch   MTime      P      T
      ParkPt:    Aug 27 2005 13:28:01  1125149281   21615  999.8 4.1024
      ParkPt:    Aug 27 2005 14:27:57  1125152877   25212 1006.8 4.1554
      ParkPt:    Aug 27 2005 15:27:57  1125156477   28812 1004.6 4.1710
      ParkPt:    Aug 27 2005 16:27:57  1125160077   32412 1004.0 4.1775
      ParkPt:    Aug 27 2005 17:27:57  1125163677   36012 1000.2 4.1525
      ParkPt:    Aug 27 2005 18:27:57  1125167277   39612 1001.0 4.1381
      ParkPt:    Aug 27 2005 19:27:57  1125170877   43212  998.6 4.1030



2) Low resolution PTS/Optode/ISUS/FLBB sample format.
-----------------------------------------------------
   
   The SBE41CP that is used on iridium floats has features that enable
   subsampling of the water column (similar to the SBE41) as well as the
   ability to bin-average a continuous sampling of the water column.  For
   subsampled data, the values of pressure, temperature, and salinity are
   not encoded but are given in conventional units (decibars, C, PSU).

   The Aanderaa optode does not report the actual oxygen concentration.
   Instead, it reports various quantities that require additional processing.
   In this case, the TPhase, RPhase, and optode temperature (C) are reported.
   The oxygen concentration is obtained by applying the calibration equations
   to the TPhase and optode temperature.  The calibration equations are
   expressed in terms of calibration coefficients that are specific to
   individual optodes.

   The MBARI ISUS nitrate sensor reports nitrate in micromolar units.

   The Wetlabs FLBB reports three ~12-bit channels that represent the output
   of an analogue-to-digital measurement.  The quantities FSig, BbSig, and
   TSig represent the fluorometer, back-scatter, and temperature channels,
   respectively.  A calibration equation must be applied to these channels in
   order to obtain chlorophyl concentration, back-scatter, and temperature in
   engineering units. 

   For example:

      $ Discrete samples: 61
      $       p        t        s     Topt   TPhase    RPhase      no3  FSig  BbSig  TSig
        1074.49   3.1208  34.4275   19.371   28.721     1.913      nan    52    106   569 (Park Sample)
         999.35   3.1810  34.3676   19.372   28.724     1.940     33.1    50    106   569
         949.57   3.2660  34.3401   19.385   28.723     1.944     33.1    50    111   569
         901.15   3.3348  34.3111   19.398   28.718     1.946     32.9    50    110   569
         851.05   3.4960  34.2855   19.402   28.715     1.950     32.1    50    108   569


3) High resolution PTS sample format.
-------------------------------------   

   For continuously sampled data, 2-decibar bins are used for bin-averaging.
   These data are encoded as three 16-bit integers (PTS) and then an 8-bit
   integer that represents the number of samples in the bin:
   
      # Nov 05 2006 23:38:59 Sbe41cpSerNo[1520] NSample[11134] NBin[495]
      00000000000000[2]
      002B5CE0885115
      003C5CE188511F
      00505CE088511E
      00645CE088511D
      00785CDF88521C
      008C5CDF88521B
      00A05CDF88511A
      00B45CDC88511A
      00C85CAD885A1A
      00DC5C8D886418
      00F05C88886919
      01045C88886A19
      01185C74887618
      012C5C4E88BB15
      [snippage...]
      2684144D874814
      2698145C874D14
   
   The first 4-bytes of the encoded sample represents the pressure in
   centibars.  The second 4-bytes represents the temperature in
   millidegrees.  The third 4-bytes represent the salinity in parts per
   million.  The final 2-bytes represent the number of samples collected in
   the 2dbar pressure bin.
   
   For example, the encoding: 2698145C874D14 represents a bin with (0x14=)
   20 samples where the mean pressure was (0x2698=) 988.0dbars, the mean
   temperature was (0x145C=) 5.212C, and the mean salinity was (0x874D=)
   34.637PSU.  The PTS values were encoded as 16-bit hex integers according
   to the functions below.
   
   Integers in square brackets '[]' indicate replicates of the same encoded
   line.  For example, a line that looks like: 00000000000000[2] indicates
   that there were 2 adjacent lines with the same encoding....all zeros in
   this case.
   
   /*------------------------------------------------------------------------*/
   /* function to encode the number of samples as an 8-bit unsigned integer  */
   /*------------------------------------------------------------------------*/
   /**
      This function encodes the number samples in the bin average as an 8-bit
      unsigned integer with protection against overflow.  The encoding accounts
      for the full range of 16-bit unsigned integers but only values in the
      open range: 0<NSample<255 are representable.  This encoding makes full
      use of all 8-bits.
   
         \begin{verbatim}
         input:
            NSample ... The number of samples in the bin-average.
   
         output:
            1) Values greater than or equal to 255 are mapped to 0xff.
   
            2) All other values are expressed as an 8-bit unsigned integer.
         \end{verbatim}
   */
   unsigned char EncodeN(unsigned int NSample)
   {
      /* prevent overflow of the sample counter */
      unsigned int N = (NSample>=255) ? 0xff : NSample;
      
      return N;
   }
   
   /*------------------------------------------------------------------------*/
   /* function to encode pressure as a 2-byte unsigned integer               */
   /*------------------------------------------------------------------------*/
   /**
      This function implements the hex-encoding of IEEE-formattted floating
      point pressure data into 2-byte signed integers with 2's complement
      representation.  The encoding formula accounts for the full range of
      32-bit IEEE floating point values but only values in the open range:
      -3276.7<p<3276.7 are representable.  This encoding makes full use of all
      16-bits.
   
         \begin{verbatim}
         input:
            p ... The pressure (decibars) expressed as a floating point value.
   
         output:
            1) Nonfinite values (Inf, -Inf, NaN) are mapped to the sentinel hex
               value: 0x8000.
   
            2) Pressure values less than -3276.7 are mapped to the sentinel
               value: 0x8001.
   
            3) Pressure values greater than 3276.7 are mapped to the sentinel
               value: 0x7fff.
   
            4) All other values are expressed in millibars rounded to the
               nearest integer and expressed as a 16-bit signed integer in
               2's-complement form.
         \end{verbatim}
   */
   unsigned int EncodeP(float p)
   {
      /* initialize with the mapping for a nonfinite pressure */
      long int P = 0x8000L;
   
      /* make sure long ints are at least 3 bytes long */
      assert(sizeof(long int)>=3);
      
      if (finite(p))
      {
         /* assign out-of-range values to sentinel values */
         if (p>=3276.7) P=0x7fffL; else if (p<=-3276.7) P=0x8001L;
   
         /* encode the pressure as the number of centibars (rounded) */
         else P = (long int)(10*(p + ((p<0) ? -0.05 : 0.05)));
   
         /* express in 16-bit 2's-complement form */
         if (P<0) P+=0x10000L;
      }
   
      return P;
   } 
   
   /*------------------------------------------------------------------------*/
   /* function to encode salinity as a 2-byte unsigned long integer          */
   /*------------------------------------------------------------------------*/
   /**
      This function implements the hex-encoding of IEEE-formattted floating
      point salinity data into 16-bit unsigned integers with 2's complement
      representation.  The encoding formula accounts for the full range of
      32-bit IEEE floating point values but only values in the open range:
      -4.095<s<61.439 are representable.  This encoding makes full use of all
      16-bits.
   
         \begin{verbatim}
         input:
            s ... The salinity (PSU) expressed as a floating point value.
   
         output:
            1) Nonfinite values (Inf, -Inf, NaN) are mapped to the sentinel hex
               value: 0xf000.
   
            2) Salinity values less than -4.095 are mapped to the sentinel
               value: 0xf001.
   
            3) Salinity values greater than 61.439 are mapped to the sentinel
               value: 0xefff.
   
            4) All other values are expressed in parts-per-million
               rounded to the nearest integer and expressed as a 16-bit
               signed integer in 2's-complement form.
         \end{verbatim}
   */
   unsigned int EncodeS(float s)
   {
      /* initialize with the mapping for a nonfinite salinity */
      long int S = 0xf000L;
   
      /* make sure that long integers have at least three bytes */
      assert(sizeof(long int)>=3);
   
      if (finite(s))
      {
         /* assign out-of-range values to sentinel values */
         if (s>=61.439) S=0xefffL; else if (s<=-4.095) S=0xf001L;
   
         /* encode the salinity as the number of parts-per-million (rounded) */
         else S = (long int)(1000*(s + ((s<0) ? -0.0005 : 0.0005)));
   
         /* express in 16-bit 2's-complement form */
         if (S<0) S+=0x10000L;
      }
      
      return S;
   }
   
   /*------------------------------------------------------------------------*/
   /* function to encode temperature as a 2-byte unsigned integer            */
   /*------------------------------------------------------------------------*/
   /**
      This function implements the hex-encoding of IEEE-formattted floating
      point temperature data into 16-bit unsigned integers with 2's complement
      representation.  The encoding formula accounts for the full range of
      32-bit IEEE floating point values but only values in the open range:
      -4.095<t<61.439 are representable.  This encoding makes full use of all
      16-bits.
   
         \begin{verbatim}
         input:
            t ... The temperature (C) expressed as a floating point value.
   
         output:
            1) Nonfinite values (Inf, -Inf, NaN) are mapped to the sentinel hex
               value: 0xf000.
   
            2) Temperature values less than -4.095 are mapped to the sentinel
               value: 0xf001.
   
            3) Temperature values greater than 61.439 are mapped to the sentinel
               value: 0xefff.
   
            4) All other values are expressed in millidegrees Celsius rounded
               to the nearest integer and expressed as a 16-bit signed integer
               in 2's-complement form.
         \end{verbatim}
   */
   unsigned int EncodeT(float t)
   {
      /* initialize with the mapping for a nonfinite temperature */
      long int T = 0xf000L;
   
      /* make sure that long integers have at least three bytes */
      assert(sizeof(long int)>=3);
   
      if (finite(t))
      {
         /* assign out-of-range values to sentinel values */
         if (t>=61.439) T=0xefffL; else if (t<=-4.095) T=0xf001L;
   
         /* encode the temperature as the number of millidegrees (rounded) */
         else T = (long int)(1000*(t + ((t<0) ? -0.0005 : 0.0005)));
   
         /* express in 16-bit 2's-complement form */
         if (T<0) T+=0x10000L;
      }
      
      return T;
   }
   


4) GPS fix format.
------------------
   
   Each telemetry cycle begins with the float attempting to acquire a GPS
   fix.  The fix includes the amount of time required to acquire the fix,
   the longitude and latitude (degrees), the date & time of the fix, and the
   number of satellites used to determine the fix.  For example:

      # GPS fix obtained in 98 seconds.
      #         lon     lat mm/dd/yyyy hhmmss nsat
      Fix: -152.945  22.544 09/01/2005 104710    8

   Positive values of longitude, latitude represent east, north hemispheres,
   respectively. Negative values of longitude, latitude represent west,
   south hemispheres, respectively.  The date is given in month-day-year
   format and the time is given in hours-minutes-seconds format.

   If no fix was acquired then the following note is entered into the
   iridium message:

      # Attempt to get GPS fix failed after 600 seconds.



5) Biographical & engineering data.
-----------------------------------

   These data have the format, "key"="value", as shown in the following
   examples:

      ActiveBallastAdjustments=5
      AirBladderPressure=119
      AirPumpAmps=91
      AirPumpVolts=192
      BuoyancyPumpOnTime=1539

   Interpretation of these data requires detailed knowledge of firmware
   implementations.  