$	Date	Time	State	Comment	Pressure	Bel Pos	Batt Volt	Batt Amps	CTD Temp	CTD Sal	Raw Press	LPF Press	LPF Vel	Diff Vel	Diff 2 Vel	Slope of LPF Press.	CTD Sample Period	LPS331 Press	LPS311 Temp	SHT21 Humid	SHT21 Temp	PID SP	PID PV	PID Y1	PID Y2	PID I	PID v	PID u	PID uCPS	PID h	FLBB	Topt	Tphase	Rphase	NO3	pH	OCR	GPS	Irid Modem	
$	12-Dec-2018	18:55:09.966	Surface Ops	Opened new engr log file: \SD\data\2018-12-12T18-55-8.eng

$	12-Dec-2018	18:55:10.069	Surface Ops	Profile Number = 2

$	12-Dec-2018	18:55:10.657	Surface Ops	Writing Header to Sci File

$	12-Dec-2018	18:55:10.778	Surface Ops	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.231716	30.450000	190.500000																																

$	12-Dec-2018	18:55:10.890	Surface Ops	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:03.0307102

$	12-Dec-2018	18:55:10.895	AB Dec Buoycy Fast	State Entry

$	12-Dec-2018	18:55:11.100	AB Dec Buoycy Fast	Decreasing Buoyancy to pressure = 0.29999999999999999 at 200000 cps

$	12-Dec-2018	18:55:11.107	AB Dec Buoycy Fast	Initializing SBE 41 LPF

$	12-Dec-2018	18:55:11.312	AB Dec Buoycy Fast	Waiting 00:00:15 for sampling to finish before ABRetractFast

$	12-Dec-2018	18:55:09.860	AB Dec Buoycy Fast	SBE41 Message						   -0.06  	-0.060000	-0.100000	0.000000	-0.013333	-0.006667	-0.000000	3.000000																							

$	12-Dec-2018	18:55:11.743	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.060000	141.237065	30.450000	190.500000																																

$	12-Dec-2018	18:55:12.098	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.060000	141.216338	30.450000	190.500000																																

$	12-Dec-2018	18:55:12.860	AB Dec Buoycy Fast	SBE41 Message						   -0.10  	-0.100000	-0.099915	0.000001	0.013333	0.000000	-0.000000	3.000000																							

$	12-Dec-2018	18:55:13.222	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.233057	30.450000	190.500000																																

$	12-Dec-2018	18:55:14.099	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.224466	30.450000	190.500000																																

$	12-Dec-2018	18:55:15.099	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.241156	30.450000	190.500000																																

$	12-Dec-2018	18:55:15.860	AB Dec Buoycy Fast	SBE41 Message						   -0.10  	-0.100000	-0.099039	0.000008	0.000000	0.006667	-0.000001	3.000000																							

$	12-Dec-2018	18:55:16.126	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.227029	30.450000	190.500000																																

$	12-Dec-2018	18:55:17.099	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.220436	30.450000	190.500000																																

$	12-Dec-2018	18:55:18.100	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.207591	30.450000	190.500000																																

$	12-Dec-2018	18:55:18.860	AB Dec Buoycy Fast	SBE41 Message						   -0.09  	-0.090000	-0.095915	0.000007	-0.003333	-0.001667	-0.000006	3.000000																							

$	12-Dec-2018	18:55:19.226	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.090000	141.229384	30.450000	190.500000																																

$	12-Dec-2018	18:55:20.100	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.090000	141.229913	30.450000	190.500000																																

$	12-Dec-2018	18:55:21.100	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.090000	141.229682	30.450000	190.500000																																

$	12-Dec-2018	18:55:21.860	AB Dec Buoycy Fast	SBE41 Message						   -0.10  	-0.100000	-0.090940	-0.000095	0.003333	0.000000	-0.000017	3.000000																							

$	12-Dec-2018	18:55:22.225	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.234189	30.450000	190.500000																																

$	12-Dec-2018	18:55:23.100	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.227737	30.450000	190.500000																																

$	12-Dec-2018	18:55:24.100	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.225889	30.450000	190.500000																																

$	12-Dec-2018	18:55:24.861	AB Dec Buoycy Fast	SBE41 Message						   -0.10  	-0.100000	-0.088099	-0.000399	0.000000	0.001667	-0.000030	3.000000																							

$	12-Dec-2018	18:55:25.129	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.218439	30.450000	190.500000																																

$	12-Dec-2018	18:55:26.100	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.240925	30.450000	190.500000																																

$	12-Dec-2018	18:55:27.101	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	-0.100000	141.226940	30.450000	190.500000																																

$	12-Dec-2018	18:55:27.899	AB Dec Buoycy Fast	Disabling unneeded power

$	12-Dec-2018	18:55:27.903	AB Dec Buoycy Fast	Powering down MSC

$	12-Dec-2018	18:55:28.005	AB Dec Buoycy Fast	Powering down GPS

$	12-Dec-2018	18:55:28.106	AB Dec Buoycy Fast	Powering down MSC

$	12-Dec-2018	18:55:28.305	AB Dec Buoycy Fast	Powering down FLBB

$	12-Dec-2018	18:55:28.407	AB Dec Buoycy Fast	Powering down 4330 Optode

$	12-Dec-2018	18:55:28.414	AB Dec Buoycy Fast	Powering down OCR504

$	12-Dec-2018	18:55:28.613	AB Dec Buoycy Fast	Powering down Bluetooth

$	12-Dec-2018	18:55:31.618	AB Dec Buoycy Fast	Powering down BT Module

$	12-Dec-2018	18:55:27.859	AB Dec Buoycy Fast	SBE41 Message						   -0.10  	-0.100000	-0.089940	-0.000794	0.000000	0.000000	-0.000041	3.000000																							

$	12-Dec-2018	18:55:30.860	AB Dec Buoycy Fast	SBE41 Message						    0.84  	0.840000	-0.093647	-0.000968	-0.313333	-0.156667	-0.000047	3.000000																							

$	12-Dec-2018	18:55:32.172	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	0.840000	141.212247	30.450000	190.500000																																

$	12-Dec-2018	18:55:32.283	AB Dec Buoycy Fast	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:04.3840999

$	12-Dec-2018	18:55:32.386	AB Dec Buoycy Fast	Starting valve open

$	12-Dec-2018	18:55:33.860	AB Dec Buoycy Fast	SBE41 Message						    0.85  	0.850000	-0.094126	-0.000714	-0.003333	-0.158333	-0.000052	3.000000																							

$	12-Dec-2018	18:55:34.320	AB Dec Buoycy Fast	RecMode true, PID Off, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.850000	141.252324	30.450000	190.500000																																

$	12-Dec-2018	18:55:34.434	AB Dec Buoycy Fast	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.1449055

$	12-Dec-2018	18:55:34.740	AB Dec Buoycy Fast	RecMode true, PID Off, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.850000	141.273543	30.450000	190.500000																																

$	12-Dec-2018	18:55:35.450	AB Dec Buoycy Fast	AB Decreasing Buoyancy Fast Exiting: Depth > pressure threshold

$	12-Dec-2018	18:55:35.658	AB Dec Buoycy Fast	Normal State Exit

$	12-Dec-2018	18:55:35.662	AB Dec Buoycy Fast	RecMode true, PID Off, Motor Stopped, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.850000	141.244188	30.450000	190.500000																																

$	12-Dec-2018	18:55:36.441	Descend	State Entry

$	12-Dec-2018	18:55:36.445	Descend	Starting valve open

$	12-Dec-2018	18:55:36.853	Descend	RecMode true, PID On, Motor Stopped, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.850000	141.263127	30.450000	190.500000																																

$	12-Dec-2018	18:55:36.964	Descend	Vel PID Values																		-0.200000	0.000000	0.000000	0.000000	-0.960000	-16.000000	-16.000000	-131072.000000	3.000000										

$	12-Dec-2018	18:55:36.964	Descend	SBE41 Message						    0.85  	0.850000	-0.073062	0.000012	0.000000	-0.001667	-0.000082	3.000000																							

$	12-Dec-2018	18:55:37.987	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.850000	141.255058	30.450000	190.500000																																

$	12-Dec-2018	18:55:38.645	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.850000	141.274981	30.450000	190.500000																																

$	12-Dec-2018	18:55:39.645	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.850000	141.286813	30.450000	190.500000																																

$	12-Dec-2018	18:55:39.860	Descend	Vel PID Values																		-0.200000	0.206667	0.205979	0.205979	-2.908699	-33.493243	-33.493243	-274376.000000	3.000000										

$	12-Dec-2018	18:55:39.860	Descend	SBE41 Message						    0.23  	0.230000	0.021842	0.000766	0.206667	0.103333	-0.000231	3.000000																							

$	12-Dec-2018	18:55:40.985	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.230000	141.301848	30.450000	190.500000																																

$	12-Dec-2018	18:55:41.646	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.230000	141.675696	30.450000	190.500000																																

$	12-Dec-2018	18:55:42.647	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.230000	141.825981	30.450000	190.500000																																

$	12-Dec-2018	18:55:42.859	Descend	Vel PID Values																		-0.200000	0.006667	0.007331	-0.198648	-3.903888	-19.442213	-19.442213	-159270.000000	3.000000										

$	12-Dec-2018	18:55:42.859	Descend	SBE41 Message						    0.21  	0.210000	0.233185	-0.001098	0.006667	0.106667	-0.000641	3.000000																							

$	12-Dec-2018	18:55:43.987	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.210000	142.129257	30.450000	190.500000																																

$	12-Dec-2018	18:55:44.648	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.210000	142.439768	30.450000	190.500000																																

$	12-Dec-2018	18:55:45.648	Descend	PTH Message														787.663574	23.600000	54.100555	24.482625																			

$	12-Dec-2018	18:55:45.852	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.210000	142.569236	30.125000	139.500000																																

$	12-Dec-2018	18:55:45.962	Descend	Vel PID Values																		-0.200000	0.000000	0.000023	-0.007308	-4.864000	-19.903803	-19.903803	-163051.000000	3.000000										

$	12-Dec-2018	18:55:45.962	Descend	SBE41 Message						    0.21  	0.210000	0.494334	-0.010650	0.000000	0.003333	-0.001367	3.000000																							

$	12-Dec-2018	18:55:46.989	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.210000	142.780393	30.125000	139.500000																																

$	12-Dec-2018	18:55:47.649	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.210000	142.989539	30.125000	139.500000																																

$	12-Dec-2018	18:55:48.649	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.210000	143.119849	30.125000	139.500000																																

$	12-Dec-2018	18:55:48.860	Descend	Vel PID Values																		-0.200000	0.000000	0.000000	-0.000023	-5.824000	-20.863997	-20.863997	-170917.000000	3.000000										

$	12-Dec-2018	18:55:48.860	Descend	SBE41 Message						    0.21  	0.210000	0.645849	-0.029789	0.000000	0.000000	-0.002252	3.000000																							

$	12-Dec-2018	18:55:49.987	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.210000	143.274136	30.125000	139.500000																																

$	12-Dec-2018	18:55:50.649	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.210000	143.543892	30.125000	139.500000																																

$	12-Dec-2018	18:55:51.649	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	0.210000	143.740296	30.125000	139.500000																																

$	12-Dec-2018	18:55:51.860	Descend	Vel PID Values																		-0.200000	0.100000	0.099667	0.099667	-7.262403	-29.823956	-29.823956	-244317.000000	3.000000										

$	12-Dec-2018	18:55:51.860	Descend	SBE41 Message						   -0.09  	-0.090000	0.601952	-0.050931	0.100000	0.050000	-0.003038	3.000000																							

$	12-Dec-2018	18:55:52.989	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	-0.090000	143.865578	30.125000	139.500000																																

$	12-Dec-2018	18:55:53.650	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	-0.090000	144.105978	30.125000	139.500000																																

$	12-Dec-2018	18:55:54.651	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	-0.090000	144.280054	30.125000	139.500000																																

$	12-Dec-2018	18:55:54.860	Descend	Vel PID Values																		-0.200000	-0.390000	-0.388370	-0.488037	-6.358227	7.937337	7.937337	65022.000000	3.000000										

$	12-Dec-2018	18:55:54.860	Descend	SBE41 Message						    1.08  	1.080000	0.439653	-0.060019	-0.390000	-0.145000	-0.003572	3.000000																							

$	12-Dec-2018	18:55:55.990	Descend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	1.080000	144.537427	30.125000	139.500000																																

$	12-Dec-2018	18:55:56.651	Descend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	1.080000	144.789718	30.125000	139.500000																																

$	12-Dec-2018	18:55:57.650	Descend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	1.080000	144.786023	30.125000	139.500000																																

$	12-Dec-2018	18:55:57.859	Descend	Vel PID Values																		-0.200000	-0.146667	-0.147474	0.240896	-6.610353	-10.624572	-10.624572	-87036.000000	3.000000										

$	12-Dec-2018	18:55:57.859	Descend	SBE41 Message						    1.52  	1.520000	0.291225	-0.047776	-0.146667	-0.268333	-0.003878	3.000000																							

$	12-Dec-2018	18:55:58.995	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	1.520000	144.780599	30.125000	139.500000																																

$	12-Dec-2018	18:55:59.651	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	1.520000	144.794092	30.125000	139.500000																																

$	12-Dec-2018	18:56:00.651	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	1.520000	144.774385	30.125000	139.500000																																

$	12-Dec-2018	18:56:00.860	Descend	Vel PID Values																		-0.200000	-0.236667	-0.236369	-0.088895	-6.435784	-3.677166	-3.677166	-30123.000000	3.000000										

$	12-Dec-2018	18:56:00.860	Descend	SBE41 Message						    2.23  	2.230000	0.221821	-0.017657	-0.236667	-0.191667	-0.004068	3.000000																							

$	12-Dec-2018	18:56:01.990	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	2.230000	144.784026	30.125000	139.500000																																

$	12-Dec-2018	18:56:02.652	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	2.230000	144.847490	30.125000	139.500000																																

$	12-Dec-2018	18:56:03.652	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	2.230000	144.883342	30.125000	139.500000																																

$	12-Dec-2018	18:56:03.860	Descend	Vel PID Values																		-0.200000	-0.163333	-0.163577	0.072792	-6.610615	-9.369046	-9.369046	-76751.000000	3.000000										

$	12-Dec-2018	18:56:03.860	Descend	SBE41 Message						    2.72  	2.720000	0.268651	0.016014	-0.163333	-0.200000	-0.004296	3.000000																							

$	12-Dec-2018	18:56:04.990	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	2.720000	144.906275	30.125000	139.500000																																

$	12-Dec-2018	18:56:05.651	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	2.720000	144.992888	30.125000	139.500000																																

$	12-Dec-2018	18:56:06.651	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	2.720000	145.048939	30.125000	139.500000																																

$	12-Dec-2018	18:56:06.860	Descend	Vel PID Values																		-0.200000	-0.210000	-0.209845	-0.046268	-6.563359	-5.810672	-5.810672	-47601.000000	3.000000										

$	12-Dec-2018	18:56:06.860	Descend	SBE41 Message						    3.35  	3.350000	0.506814	0.035714	-0.210000	-0.186667	-0.004810	3.000000																							

$	12-Dec-2018	18:56:07.992	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	3.350000	145.142905	30.125000	139.500000																																

$	12-Dec-2018	18:56:08.652	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	3.350000	145.220362	30.125000	139.500000																																

$	12-Dec-2018	18:56:09.652	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	3.350000	145.244472	30.125000	139.500000																																

$	12-Dec-2018	18:56:09.859	Descend	Vel PID Values																		-0.200000	-0.203333	-0.203355	0.006490	-6.547254	-6.296666	-6.296666	-51582.000000	3.000000										

$	12-Dec-2018	18:56:09.859	Descend	SBE41 Message						    3.96  	3.960000	0.965106	0.028892	-0.203333	-0.206667	-0.005883	3.000000																							

$	12-Dec-2018	18:56:10.992	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	3.960000	145.312853	30.125000	139.500000																																

$	12-Dec-2018	18:56:11.653	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	3.960000	145.397037	30.125000	139.500000																																

$	12-Dec-2018	18:56:12.653	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	3.960000	145.407900	30.125000	139.500000																																

$	12-Dec-2018	18:56:12.859	Descend	Vel PID Values																		-0.200000	-0.206667	-0.206656	-0.003300	-6.515307	-6.013928	-6.013928	-49266.000000	3.000000										

$	12-Dec-2018	18:56:12.859	Descend	SBE41 Message						    4.58  	4.580000	1.554039	-0.005792	-0.206667	-0.205000	-0.007655	3.000000																							

$	12-Dec-2018	18:56:13.992	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	4.580000	145.480379	30.125000	139.500000																																

$	12-Dec-2018	18:56:14.653	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	4.580000	145.568222	30.125000	139.500000																																

$	12-Dec-2018	18:56:15.653	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	4.580000	145.570934	30.125000	139.500000																																

$	12-Dec-2018	18:56:15.860	Descend	Vel PID Values																		-0.200000	-0.200000	-0.200022	0.006633	-6.515201	-6.515305	-6.515305	-53373.000000	3.000000										

$	12-Dec-2018	18:56:15.860	Descend	SBE41 Message						    5.18  	5.180000	2.163760	-0.058354	-0.200000	-0.203333	-0.010113	3.000000																							

$	12-Dec-2018	18:56:16.992	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	5.180000	145.641796	30.125000	139.500000																																

$	12-Dec-2018	18:56:17.653	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	5.180000	145.704933	30.125000	139.500000																																

$	12-Dec-2018	18:56:18.653	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	5.180000	145.758137	30.125000	139.500000																																

$	12-Dec-2018	18:56:18.860	Descend	Vel PID Values																		-0.200000	-0.200000	-0.200000	0.000022	-6.515200	-6.515194	-6.515194	-53372.000000	3.000000										

$	12-Dec-2018	18:56:18.860	Descend	SBE41 Message						    5.78  	5.780000	2.760268	-0.114450	-0.200000	-0.200000	-0.013194	3.000000																							

$	12-Dec-2018	18:56:19.994	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	5.780000	145.822935	30.125000	139.500000																																

$	12-Dec-2018	18:56:20.654	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	5.780000	145.921364	30.125000	139.500000																																

$	12-Dec-2018	18:56:21.654	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	5.780000	145.922706	30.125000	139.500000																																

$	12-Dec-2018	18:56:21.860	Descend	Vel PID Values																		-0.200000	-0.250000	-0.249834	-0.049833	-6.275999	-2.515222	-2.515222	-20604.000000	3.000000										

$	12-Dec-2018	18:56:21.860	Descend	SBE41 Message						    6.53  	6.530000	3.357050	-0.161958	-0.250000	-0.225000	-0.016853	3.000000																							

$	12-Dec-2018	18:56:22.995	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	6.530000	145.984873	30.125000	139.500000																																

$	12-Dec-2018	18:56:23.654	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	6.530000	146.072693	30.125000	139.500000																																

$	12-Dec-2018	18:56:24.654	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	6.530000	146.111473	30.125000	139.500000																																

$	12-Dec-2018	18:56:24.861	Descend	Vel PID Values																		-0.200000	-0.140000	-0.140366	0.109468	-6.562244	-11.075939	-11.075939	-90734.000000	3.000000										

$	12-Dec-2018	18:56:24.861	Descend	SBE41 Message						    6.95  	6.950000	3.962655	-0.193108	-0.140000	-0.195000	-0.021054	3.000000																							

$	12-Dec-2018	18:56:26.016	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	6.950000	146.126732	30.125000	139.500000																																

$	12-Dec-2018	18:56:26.676	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	6.950000	146.144323	30.125000	139.500000																																

$	12-Dec-2018	18:56:27.676	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	6.950000	146.230064	30.125000	139.500000																																

$	12-Dec-2018	18:56:27.860	Descend	Vel PID Values																		-0.200000	-0.203333	-0.203123	-0.062758	-6.547252	-6.295650	-6.295650	-51573.000000	3.000000										

$	12-Dec-2018	18:56:27.860	Descend	SBE41 Message						    7.56  	7.560000	4.574620	-0.206513	-0.203333	-0.171667	-0.025756	3.000000																							

$	12-Dec-2018	18:56:29.014	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	7.560000	146.365047	30.125000	139.500000																																

$	12-Dec-2018	18:56:29.676	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	7.560000	146.461897	30.125000	139.500000																																

$	12-Dec-2018	18:56:30.677	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	7.560000	146.495722	30.125000	139.500000																																

$	12-Dec-2018	18:56:30.861	Descend	Vel PID Values																		-0.200000	-0.230000	-0.229911	-0.026788	-6.403680	-4.147235	-4.147235	-33974.000000	3.000000										

$	12-Dec-2018	18:56:30.861	Descend	SBE41 Message						    8.25  	8.250000	5.191666	-0.208511	-0.230000	-0.216667	-0.030915	3.000000																							

$	12-Dec-2018	18:56:32.019	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	8.250000	146.524444	30.125000	139.500000																																

$	12-Dec-2018	18:56:32.678	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	8.250000	146.627195	30.125000	139.500000																																

$	12-Dec-2018	18:56:33.678	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	8.250000	146.624595	30.125000	139.500000																																

$	12-Dec-2018	18:56:33.860	Descend	Vel PID Values																		-0.200000	-0.183333	-0.183489	0.046422	-6.482935	-7.736976	-7.736976	-63381.000000	3.000000										

$	12-Dec-2018	18:56:33.860	Descend	SBE41 Message						    8.80  	8.800000	5.809956	-0.208061	-0.183333	-0.206667	-0.036481	3.000000																							

$	12-Dec-2018	18:56:35.018	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	8.800000	146.650776	30.125000	139.500000																																

$	12-Dec-2018	18:56:35.679	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	8.800000	146.741435	30.125000	139.500000																																

$	12-Dec-2018	18:56:36.679	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	8.800000	146.788806	30.125000	139.500000																																

$	12-Dec-2018	18:56:36.859	Descend	Vel PID Values																		-0.200000	-0.203333	-0.203267	-0.019778	-6.467253	-6.216307	-6.216307	-50923.000000	3.000000										

$	12-Dec-2018	18:56:36.859	Descend	SBE41 Message						    9.41  	9.410000	6.416589	-0.209332	-0.203333	-0.193333	-0.042386	3.000000																							

$	12-Dec-2018	18:56:38.020	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	9.410000	146.872357	30.125000	139.500000																																

$	12-Dec-2018	18:56:38.680	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	9.410000	146.928646	30.125000	139.500000																																

$	12-Dec-2018	18:56:39.680	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	9.410000	147.015721	30.125000	139.500000																																

$	12-Dec-2018	18:56:39.860	Descend	Vel PID Values																		-0.200000	-0.200000	-0.200011	0.003256	-6.467200	-6.467233	-6.467233	-52979.000000	3.000000										

$	12-Dec-2018	18:56:39.860	Descend	SBE41 Message						   10.01  	10.010000	7.008323	-0.211030	-0.200000	-0.201667	-0.048560	3.000000																							

$	12-Dec-2018	18:56:41.019	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	10.010000	147.018977	30.125000	139.500000																																

$	12-Dec-2018	18:56:41.680	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	10.010000	147.042900	30.125000	139.500000																																

$	12-Dec-2018	18:56:42.680	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	10.010000	147.099428	30.125000	139.500000																																

$	12-Dec-2018	18:56:42.861	Descend	Vel PID Values																		-0.200000	-0.250000	-0.249834	-0.049823	-6.227999	-2.467224	-2.467224	-20211.000000	3.000000										

$	12-Dec-2018	18:56:42.861	Descend	SBE41 Message						   10.76  	10.760000	7.601354	-0.211064	-0.250000	-0.225000	-0.054953	3.000000																							

$	12-Dec-2018	18:56:44.019	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	10.760000	147.160329	30.125000	139.500000																																

$	12-Dec-2018	18:56:44.681	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	10.760000	147.280946	30.125000	139.500000																																

$	12-Dec-2018	18:56:45.682	Descend	PTH Message														789.947266	23.733333	54.100555	24.536250																			

$	12-Dec-2018	18:56:45.884	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	10.760000	147.268184	30.350000	130.000000																																

$	12-Dec-2018	18:56:45.883	Descend	Vel PID Values																		-0.200000	-0.200000	-0.200166	0.049668	-6.227202	-6.227970	-6.227970	-51019.000000	3.000000										

$	12-Dec-2018	18:56:45.883	Descend	SBE41 Message						   11.36  	11.360000	8.205113	-0.209470	-0.200000	-0.225000	-0.061530	3.000000																							

$	12-Dec-2018	18:56:47.021	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	11.360000	147.286326	30.350000	130.000000																																

$	12-Dec-2018	18:56:47.681	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	11.360000	147.296503	30.350000	130.000000																																

$	12-Dec-2018	18:56:48.682	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	11.360000	147.341050	30.350000	130.000000																																

$	12-Dec-2018	18:56:48.861	Descend	Vel PID Values																		-0.200000	-0.180000	-0.180067	0.020099	-6.322882	-7.827217	-7.827217	-64120.000000	3.000000										

$	12-Dec-2018	18:56:48.861	Descend	SBE41 Message						   11.90  	11.900000	8.813751	-0.207555	-0.180000	-0.190000	-0.068246	3.000000																							

$	12-Dec-2018	18:56:50.022	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	11.900000	147.406884	30.350000	130.000000																																

$	12-Dec-2018	18:56:50.682	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	11.900000	147.491142	30.350000	130.000000																																

$	12-Dec-2018	18:56:51.682	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	11.900000	147.550151	30.350000	130.000000																																

$	12-Dec-2018	18:56:51.861	Descend	Vel PID Values																		-0.200000	-0.193333	-0.193289	-0.013222	-6.355094	-6.856232	-6.856232	-56166.000000	3.000000										

$	12-Dec-2018	18:56:51.861	Descend	SBE41 Message						   12.48  	12.480000	9.429610	-0.206295	-0.193333	-0.186667	-0.075063	3.000000																							

$	12-Dec-2018	18:56:53.025	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	12.480000	147.593163	30.350000	130.000000																																

$	12-Dec-2018	18:56:53.683	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	12.480000	147.698373	30.350000	130.000000																																

$	12-Dec-2018	18:56:54.683	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	12.480000	147.702642	30.350000	130.000000																																

$	12-Dec-2018	18:56:54.861	Descend	Vel PID Values																		-0.200000	-0.183333	-0.183367	0.009922	-6.434934	-7.688412	-7.688412	-62983.000000	3.000000										

$	12-Dec-2018	18:56:54.861	Descend	SBE41 Message						   13.03  	13.030000	10.061357	-0.206398	-0.183333	-0.188333	-0.081950	3.000000																							

$	12-Dec-2018	18:56:56.028	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	13.030000	147.818953	30.350000	130.000000																																

$	12-Dec-2018	18:56:56.682	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	13.030000	147.872388	30.350000	130.000000																																

$	12-Dec-2018	18:56:57.682	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	13.030000	147.917584	30.350000	130.000000																																

$	12-Dec-2018	18:56:57.861	Descend	Vel PID Values																		-0.200000	-0.183333	-0.183333	0.000033	-6.514934	-7.768267	-7.768267	-63637.000000	3.000000										

$	12-Dec-2018	18:56:57.861	Descend	SBE41 Message						   13.58  	13.580000	10.698384	-0.208466	-0.183333	-0.183333	-0.088865	3.000000																							

$	12-Dec-2018	18:56:59.022	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	13.580000	147.975422	30.350000	130.000000																																

$	12-Dec-2018	18:56:59.683	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	13.580000	148.079172	30.350000	130.000000																																

$	12-Dec-2018	18:57:00.683	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	13.580000	148.113690	30.350000	130.000000																																

$	12-Dec-2018	18:57:00.861	Descend	Vel PID Values																		-0.200000	-0.193333	-0.193300	-0.009967	-6.547094	-7.048274	-7.048274	-57739.000000	3.000000										

$	12-Dec-2018	18:57:00.861	Descend	SBE41 Message						   14.16  	14.160000	11.315507	-0.211895	-0.193333	-0.188333	-0.095736	3.000000																							

$	12-Dec-2018	18:57:02.024	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	14.160000	148.166485	30.350000	130.000000																																

$	12-Dec-2018	18:57:02.683	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	14.160000	148.250207	30.350000	130.000000																																

$	12-Dec-2018	18:57:03.683	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	14.160000	148.321666	30.350000	130.000000																																

$	12-Dec-2018	18:57:03.861	Descend	Vel PID Values																		-0.200000	-0.196667	-0.196656	-0.003356	-6.563147	-6.813754	-6.813754	-55818.000000	3.000000										

$	12-Dec-2018	18:57:03.861	Descend	SBE41 Message						   14.75  	14.750000	11.902485	-0.214725	-0.196667	-0.195000	-0.102481	3.000000																							

$	12-Dec-2018	18:57:05.023	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	14.750000	148.323506	30.350000	130.000000																																

$	12-Dec-2018	18:57:05.682	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	14.750000	148.424834	30.350000	130.000000																																

$	12-Dec-2018	18:57:06.682	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	14.750000	148.480214	30.350000	130.000000																																

$	12-Dec-2018	18:57:06.861	Descend	Vel PID Values																		-0.200000	-0.186667	-0.186700	0.009956	-6.626987	-7.629797	-7.629797	-62503.000000	3.000000										

$	12-Dec-2018	18:57:06.861	Descend	SBE41 Message						   15.31  	15.310000	12.469792	-0.215195	-0.186667	-0.191667	-0.109037	3.000000																							

$	12-Dec-2018	18:57:08.022	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	15.310000	148.550965	30.350000	130.000000																																

$	12-Dec-2018	18:57:08.682	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	15.310000	148.631386	30.350000	130.000000																																

$	12-Dec-2018	18:57:09.682	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	15.310000	148.639560	30.350000	130.000000																																

$	12-Dec-2018	18:57:09.862	Descend	Vel PID Values																		-0.200000	-0.170000	-0.170055	0.016645	-6.770721	-9.027003	-9.027003	-73949.000000	3.000000										

$	12-Dec-2018	18:57:09.862	Descend	SBE41 Message						   15.82  	15.820000	13.031220	-0.212602	-0.170000	-0.178333	-0.116232	3.000000																							

$	12-Dec-2018	18:57:11.021	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	15.820000	148.729473	30.350000	130.000000																																

$	12-Dec-2018	18:57:11.683	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	15.820000	148.819491	30.350000	130.000000																																

$	12-Dec-2018	18:57:12.683	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	15.820000	148.901634	30.350000	130.000000																																

$	12-Dec-2018	18:57:12.861	Descend	Vel PID Values																		-0.200000	-0.133333	-0.133455	0.036600	-7.090136	-12.104056	-12.104056	-99156.000000	3.000000										

$	12-Dec-2018	18:57:12.861	Descend	SBE41 Message						   16.22  	16.219999	13.595169	-0.207313	-0.133333	-0.151666	-0.123294	3.000000																							

$	12-Dec-2018	18:57:14.023	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	16.219999	148.962967	30.350000	130.000000																																

$	12-Dec-2018	18:57:14.684	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	16.219999	149.052754	30.350000	130.000000																																

$	12-Dec-2018	18:57:15.685	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	16.219999	149.129972	30.350000	130.000000																																

$	12-Dec-2018	18:57:15.861	Descend	Vel PID Values																		-0.200000	-0.146667	-0.146623	-0.013167	-7.346347	-11.356811	-11.356811	-93034.000000	3.000000										

$	12-Dec-2018	18:57:15.861	Descend	SBE41 Message						   16.66  	16.660000	14.163396	-0.201116	-0.146667	-0.140000	-0.130193	3.000000																							

$	12-Dec-2018	18:57:17.026	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	16.660000	149.252586	30.350000	130.000000																																

$	12-Dec-2018	18:57:17.684	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	16.660000	149.426132	30.350000	130.000000																																

$	12-Dec-2018	18:57:18.684	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	16.660000	149.461791	30.350000	130.000000																																

$	12-Dec-2018	18:57:18.862	Descend	Vel PID Values																		-0.200000	-0.146667	-0.146667	-0.000044	-7.602347	-11.612994	-11.612994	-95133.000000	3.000000										

$	12-Dec-2018	18:57:18.862	Descend	SBE41 Message						   17.10  	17.100000	14.726239	-0.196563	-0.146667	-0.146667	-0.136892	3.000000																							

$	12-Dec-2018	18:57:20.026	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	17.100000	149.556272	30.350000	130.000000																																

$	12-Dec-2018	18:57:20.685	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	17.100000	149.708584	30.350000	130.000000																																

$	12-Dec-2018	18:57:21.685	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	17.100000	149.737656	30.350000	130.000000																																

$	12-Dec-2018	18:57:21.860	Descend	Vel PID Values																		-0.200000	-0.156666	-0.156633	-0.009966	-7.810507	-11.069036	-11.069036	-90677.000000	3.000000										

$	12-Dec-2018	18:57:21.860	Descend	SBE41 Message						   17.57  	17.570000	15.264025	-0.194776	-0.156666	-0.151667	-0.143325	3.000000																							

$	12-Dec-2018	18:57:23.025	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	17.570000	149.829522	30.350000	130.000000																																

$	12-Dec-2018	18:57:23.695	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	17.570000	149.973735	30.350000	130.000000																																

$	12-Dec-2018	18:57:24.696	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	17.570000	150.041148	30.350000	130.000000																																

$	12-Dec-2018	18:57:24.861	Descend	Vel PID Values																		-0.200000	-0.160000	-0.159989	-0.003356	-8.002562	-11.010517	-11.010517	-90198.000000	3.000000										

$	12-Dec-2018	18:57:24.861	Descend	SBE41 Message						   18.05  	18.049999	15.761487	-0.194215	-0.160000	-0.158333	-0.149409	3.000000																							

$	12-Dec-2018	18:57:26.036	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	18.049999	150.157407	30.350000	130.000000																																

$	12-Dec-2018	18:57:26.696	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	18.049999	150.257014	30.350000	130.000000																																

$	12-Dec-2018	18:57:27.697	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	18.049999	150.313549	30.350000	130.000000																																

$	12-Dec-2018	18:57:27.861	Descend	Vel PID Values																		-0.200000	-0.303333	-0.302856	-0.142868	-7.508852	0.264039	0.264039	2163.000000	3.000000										

$	12-Dec-2018	18:57:27.861	Descend	SBE41 Message						   18.96  	18.959999	16.222607	-0.191993	-0.303333	-0.231667	-0.155062	3.000000																							

$	12-Dec-2018	18:57:29.037	Descend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	18.959999	150.429972	30.350000	130.000000																																

$	12-Dec-2018	18:57:29.697	Descend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	18.959999	150.515109	30.350000	130.000000																																

$	12-Dec-2018	18:57:30.697	Descend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	18.959999	150.518961	30.350000	130.000000																																

$	12-Dec-2018	18:57:30.861	Descend	Vel PID Values																		-0.200000	-0.273334	-0.273433	0.029423	-7.156374	-1.642066	-1.642066	-13451.000000	3.000000										

$	12-Dec-2018	18:57:30.861	Descend	SBE41 Message						   19.78  	19.780001	16.668940	-0.186174	-0.273334	-0.288334	-0.160235	3.000000																							

$	12-Dec-2018	18:57:32.037	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	19.780001	150.523558	30.350000	130.000000																																

$	12-Dec-2018	18:57:32.697	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	19.780001	150.530249	30.350000	130.000000																																

$	12-Dec-2018	18:57:33.698	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	19.780001	150.556110	30.350000	130.000000																																

$	12-Dec-2018	18:57:33.861	Descend	Vel PID Values																		-0.200000	-0.203333	-0.203566	0.069867	-7.139257	-6.889724	-6.889724	-56440.000000	3.000000										

$	12-Dec-2018	18:57:33.861	Descend	SBE41 Message						   20.39  	20.389999	17.129749	-0.177047	-0.203333	-0.238333	-0.164920	3.000000																							

$	12-Dec-2018	18:57:35.038	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	20.389999	150.563188	30.350000	130.000000																																

$	12-Dec-2018	18:57:35.697	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	20.389999	150.577933	30.350000	130.000000																																

$	12-Dec-2018	18:57:36.698	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	20.389999	150.588341	30.350000	130.000000																																

$	12-Dec-2018	18:57:36.862	Descend	Vel PID Values																		-0.200000	-0.296667	-0.296357	-0.092791	-6.676744	0.594039	0.594039	4866.000000	3.000000										

$	12-Dec-2018	18:57:36.862	Descend	SBE41 Message						   21.28  	21.280001	17.644208	-0.167098	-0.296667	-0.250000	-0.169154	3.000000																							

$	12-Dec-2018	18:57:38.038	Descend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	21.280001	150.680423	30.350000	130.000000																																

$	12-Dec-2018	18:57:38.699	Descend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	21.280001	150.757559	30.350000	130.000000																																

$	12-Dec-2018	18:57:39.699	Descend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	21.280001	150.739133	30.350000	130.000000																																

$	12-Dec-2018	18:57:39.861	Descend	Vel PID Values																		-0.200000	-0.256666	-0.256799	0.039558	-6.404110	-2.143390	-2.143390	-17558.000000	3.000000										

$	12-Dec-2018	18:57:39.861	Descend	SBE41 Message						   22.05  	22.049999	18.252615	-0.160560	-0.256666	-0.276667	-0.173169	3.000000																							

$	12-Dec-2018	18:57:41.039	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	22.049999	150.775261	30.350000	130.000000																																

$	12-Dec-2018	18:57:41.700	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	22.049999	150.784865	30.350000	130.000000																																

$	12-Dec-2018	18:57:42.700	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	22.049999	150.790751	30.350000	130.000000																																

$	12-Dec-2018	18:57:42.861	Descend	Vel PID Values																		-0.200000	-0.233334	-0.233411	0.023387	-6.243735	-3.737430	-3.737430	-30617.000000	3.000000										

$	12-Dec-2018	18:57:42.861	Descend	SBE41 Message						   22.75  	22.750000	18.956475	-0.162355	-0.233334	-0.245000	-0.177175	3.000000																							

$	12-Dec-2018	18:57:44.042	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	22.750000	150.784403	30.350000	130.000000																																

$	12-Dec-2018	18:57:44.701	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	22.750000	150.819123	30.350000	130.000000																																

$	12-Dec-2018	18:57:45.701	Descend	PTH Message														792.870605	23.583333	54.077667	24.568425																			

$	12-Dec-2018	18:57:45.904	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	22.750000	150.847316	30.350000	151.500000																																

$	12-Dec-2018	18:57:45.902	Descend	Vel PID Values																		-0.200000	-0.256667	-0.256589	-0.023178	-5.972106	-1.710410	-1.710410	-14011.000000	3.000000										

$	12-Dec-2018	18:57:45.902	Descend	SBE41 Message						   23.52  	23.520000	19.711118	-0.175231	-0.256667	-0.245000	-0.181226	3.000000																							

$	12-Dec-2018	18:57:47.048	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	23.520000	150.857009	30.350000	151.500000																																

$	12-Dec-2018	18:57:47.701	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	23.520000	150.886193	30.350000	151.500000																																

$	12-Dec-2018	18:57:48.701	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	23.520000	150.915816	30.350000	151.500000																																

$	12-Dec-2018	18:57:48.860	Descend	Vel PID Values																		-0.200000	-0.256667	-0.256667	-0.000077	-5.700106	-1.438750	-1.438750	-11786.000000	3.000000										

$	12-Dec-2018	18:57:48.860	Descend	SBE41 Message						   24.29  	24.290001	20.477902	-0.197042	-0.256667	-0.256667	-0.185241	3.000000																							

$	12-Dec-2018	18:57:50.042	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.290001	150.921233	30.350000	151.500000																																

$	12-Dec-2018	18:57:50.701	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.290001	150.966838	30.350000	151.500000																																

$	12-Dec-2018	18:57:51.701	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.290001	150.984570	30.350000	151.500000																																

$	12-Dec-2018	18:57:51.860	Descend	Vel PID Values																		-0.200000	-0.253333	-0.253344	0.003323	-5.444055	-1.433483	-1.433483	-11743.000000	3.000000										

$	12-Dec-2018	18:57:51.860	Descend	SBE41 Message						   25.05  	25.049999	21.247433	-0.221859	-0.253333	-0.255000	-0.188898	3.000000																							

$	12-Dec-2018	18:57:53.042	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.049999	151.007704	30.350000	151.500000																																

$	12-Dec-2018	18:57:53.701	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.049999	150.989659	30.350000	151.500000																																

$	12-Dec-2018	18:57:54.702	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.049999	150.992706	30.350000	151.500000																																

$	12-Dec-2018	18:57:54.861	Descend	Vel PID Values																		-0.200000	-0.150000	-0.150344	0.103000	-5.682403	-9.443991	-9.443991	-77365.000000	3.000000										

$	12-Dec-2018	18:57:54.861	Descend	SBE41 Message						   25.50  	25.500000	22.013863	-0.243457	-0.150000	-0.201667	-0.191991	3.000000																							

$	12-Dec-2018	18:57:56.042	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.500000	151.009470	30.350000	151.500000																																

$	12-Dec-2018	18:57:56.702	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.500000	151.016608	30.350000	151.500000																																

$	12-Dec-2018	18:57:57.703	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.500000	151.044287	30.350000	151.500000																																

$	12-Dec-2018	18:57:57.861	Descend	Vel PID Values																		-0.200000	0.000000	-0.000500	0.149844	-6.640005	-21.682383	-21.682383	-177622.000000	3.000000										

$	12-Dec-2018	18:57:57.861	Descend	SBE41 Message						   25.50  	25.500000	22.771051	-0.257607	0.000000	-0.075000	-0.194467	3.000000																							

$	12-Dec-2018	18:57:59.043	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.500000	151.166856	30.350000	151.500000																																

$	12-Dec-2018	18:57:59.703	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.500000	151.345052	30.350000	151.500000																																

$	12-Dec-2018	18:58:00.704	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.500000	151.475169	30.350000	151.500000																																

$	12-Dec-2018	18:58:00.861	Descend	Vel PID Values																		-0.200000	0.006667	0.006644	0.007144	-7.631895	-23.173414	-23.173414	-189836.000000	3.000000										

$	12-Dec-2018	18:58:00.861	Descend	SBE41 Message						   25.48  	25.480000	23.517433	-0.263533	0.006667	0.003333	-0.196359	3.000000																							

$	12-Dec-2018	18:58:02.044	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.480000	151.670322	30.350000	151.500000																																

$	12-Dec-2018	18:58:02.704	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.480000	151.911922	30.350000	151.500000																																

$	12-Dec-2018	18:58:03.705	Descend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.480000	152.052842	30.350000	151.500000																																

$	12-Dec-2018	18:58:04.502	Descend	Bottom detected based on Platform Velocity near zero

$	12-Dec-2018	18:58:03.861	Descend	SBE41 Message						   25.48  	25.480000	24.225429	-0.264153	0.000000	0.003333	-0.197718	3.000000																							

$	12-Dec-2018	18:58:04.939	Descend	Normal State Exit

$	12-Dec-2018	18:58:04.943	Descend	RecMode true, PID Off, Motor Stopped, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.480000	152.210452	30.350000	151.500000																																

$	12-Dec-2018	18:58:05.502	Anchor	State Entry

$	12-Dec-2018	18:58:05.506	Anchor	Starting valve close

$	12-Dec-2018	18:58:05.612	Anchor	Powering up MSC

$	12-Dec-2018	18:58:05.713	Anchor	Powering up FLBB

$	12-Dec-2018	18:58:07.315	Anchor	Sending FLBB stop command

$	12-Dec-2018	18:58:08.822	Anchor	Sending FLBB pkt command: $pkt 1  

$	12-Dec-2018	18:58:09.926	Anchor	Sending FLBB ave command: $ave 10  

$	12-Dec-2018	18:58:10.932	Anchor	Powering up 4330 Optode

$	12-Dec-2018	18:58:12.436	Anchor	Sending Optode Wakeup

$	12-Dec-2018	18:58:16.453	Anchor	Powering up OCR504

$	12-Dec-2018	18:58:16.561	Anchor	Starting anchor period = 00:04:00

$	12-Dec-2018	18:58:16.995	Anchor	Error in I2CPortBase.Execute

$	12-Dec-2018	18:58:18.006	Anchor	ERROR: Missed PH message 24

$	12-Dec-2018	18:58:18.010	Anchor	ERROR: Missed NO3 message 20

$	12-Dec-2018	18:58:18.212	Anchor	Sending Optode Do Sample

$	12-Dec-2018	18:58:18.313	Anchor	Sending Optode Wakeup

$	12-Dec-2018	18:58:18.317	Anchor	Error executing 2485 Transaction = 0

$	12-Dec-2018	18:58:18.420	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.480000	250.000000	30.350000	151.500000																																

$	12-Dec-2018	18:58:18.631	Anchor	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:13.1282992

$	12-Dec-2018	18:58:06.861	Anchor	SBE41 Message						   25.46  	25.459999	24.819281	-0.262516	0.006667	0.003333	-0.198608	3.000000																							

$	12-Dec-2018	18:58:18.859	Anchor	Sending FLBB run command

$	12-Dec-2018	18:58:18.876	Anchor	Sending radiometer pollSample

$	12-Dec-2018	18:58:18.982	Anchor	MSC: RX: 

$	12-Dec-2018	18:58:19.087	Anchor	MSC: RX: 

$	12-Dec-2018	18:58:19.189	Anchor	MSC: RX:   MBARI Sensor Controller (ISUS/DURA) Initializing... 

$	12-Dec-2018	18:58:09.861	Anchor	SBE41 Message						   25.46  	25.459999	25.218737	-0.256984	0.000000	0.003333	-0.199052	3.000000																							

$	12-Dec-2018	18:58:11.208	Anchor	Optode Message																												% %  								

$	12-Dec-2018	18:58:11.319	Anchor	Optode Message																												StartupInfo,4330,1532,Mode,AADI Smart Sensor Terminal Protocol,RS232 Protocol Version,3,Config Version,11  								

$	12-Dec-2018	18:58:19.760	Anchor	MSC: RX:   MBARI Serial Number : 0777 

$	12-Dec-2018	18:58:20.064	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.459999	152.475141	30.350000	151.500000																																

$	12-Dec-2018	18:58:20.177	Anchor	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.5411502

$	12-Dec-2018	18:58:20.183	Anchor	MSC: RX:   CF2 App: APF_MAIN.C Date: Sep 30 2015 14:08:04 

$	12-Dec-2018	18:58:20.384	Anchor	MSC: RX:   CF2 S/N: 14968 - BIOS:4.03 - PicoDOS:4.03 

$	12-Dec-2018	18:58:12.861	Anchor	SBE41 Message						   25.13  	25.129999	25.413161	-0.241263	0.110000	0.055000	-0.198972	3.000000																							

$	12-Dec-2018	18:58:20.514	Anchor	MSC: RX:   CF Card: Size=256614400 Avail=99.13% 

$	12-Dec-2018	18:58:20.714	Anchor	MSC: RX:   CF2 Power Up  (GMT): Tue Jan 31 01:23:16 2012 

$	12-Dec-2018	18:58:20.719	Anchor	MSC: RX:   CF2 RTC Time  (GMT): Tue Jan 31 01:23:20 2012 

$	12-Dec-2018	18:58:20.921	Anchor	MSC: RX:   Adjusted Local Time: Mon Jan 30 17:23:20 2012 

$	12-Dec-2018	18:58:21.024	Anchor	MSC: RX:   TZ Offset (GMT - LOCAL)  : +8.00 hrs. 

$	12-Dec-2018	18:58:21.029	Anchor	MSC: RX:   RTC delta (CF2 - DS3234) : -1.00 secs. 

$	12-Dec-2018	18:58:21.133	Anchor	MSC: RX:   System Batt Voltage = 11.764 

$	12-Dec-2018	18:58:21.374	Anchor	MSC: RX:   System Load Current = 2.62e-02 

$	12-Dec-2018	18:58:13.591	Anchor	Optode Message																												!#  								

$	12-Dec-2018	18:58:21.595	Anchor	MSC: RX:   Housing Temperature = +22.85C = +73.13F 

$	12-Dec-2018	18:58:21.697	Anchor	MSC: RX:   Housing    Humidity =  56.0%  =  13.6C (DP) 

$	12-Dec-2018	18:58:21.801	Anchor	MSC: RX:   Power Cycle   Count = 145 

$	12-Dec-2018	18:58:22.007	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.129999	152.471997	30.350000	151.500000																																

$	12-Dec-2018	18:58:22.119	Anchor	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.9363428

$	12-Dec-2018	18:58:22.224	Anchor	MSC: RX:   System Reset  Count = 149 

$	12-Dec-2018	18:58:22.326	Anchor	MSC: RX:   System Error  Count = 64 

$	12-Dec-2018	18:58:22.428	Anchor	MSC: RX:   Last System   Error = -110 

$	12-Dec-2018	18:58:22.533	Anchor	MSC: RX:   Last Error     Time = Tue Jan 31 01:07:48 2012 

$	12-Dec-2018	18:58:22.539	Anchor	MSC: RX:   ISUS Error    Count = 1 

$	12-Dec-2018	18:58:22.641	Anchor	MSC: RX:   Last ISUS     Error = -202 

$	12-Dec-2018	18:58:22.842	Anchor	MSC: RX:   Last Error     Time = Thu Jan 26 03:13:44 2012 

$	12-Dec-2018	18:58:22.847	Anchor	MSC: RX:   DURA Error    Count = 0 

$	12-Dec-2018	18:58:22.949	Anchor	MSC: RX:   Last DURA     Error = 0 

$	12-Dec-2018	18:58:23.051	Anchor	MSC: RX:   Last Error     Time = Thu Jan  1 00:00:00 1970 

$	12-Dec-2018	18:58:23.251	Anchor	MSC: RX:   Fiberlite  Odometer = 100:10:56 = 360656 secs 

$	12-Dec-2018	18:58:23.257	Anchor	MSC: RX:   Fiberlite    Cycles = 227 

$	12-Dec-2018	18:58:23.458	Anchor	MSC: RX:   Spectrometer Status = MBARI-ISUS 1.1, 04-Mar-15, 10:16:08 

$	12-Dec-2018	18:58:23.560	Anchor	MSC: RX:   Zeiss Coef File Found 4 coefs loaded. 

$	12-Dec-2018	18:58:23.663	Anchor	MSC: RX:   Ecoef File Found 256 coefs loaded. 

$	12-Dec-2018	18:58:23.967	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.129999	152.453817	30.350000	151.500000																																

$	12-Dec-2018	18:58:24.078	Anchor	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.9522915

$	12-Dec-2018	18:58:24.183	Anchor	MSC: RX:   There are 70 records in the error file! 

$	12-Dec-2018	18:58:24.285	Anchor	MSC: RX:   MBARI Sensor Controller (ISUS/DURA) Initialized 

$	12-Dec-2018	18:58:14.624	Anchor	Optode Message																												#  								

$	12-Dec-2018	18:58:24.406	Anchor	MSC: RX:   Auto Shutdown in 10 hours at: Tue Jan 31 11:23:21 2012 

$	12-Dec-2018	18:58:24.607	Anchor	MSC: RX:   Entering Suspend Mode, Serial Input (/WAKE) will wake me, send M for Menu Mode  

$	12-Dec-2018	18:58:15.494	Anchor	Optode Message																												#  								

$	12-Dec-2018	18:58:15.860	Anchor	SBE41 Message						   25.03  	25.030001	25.470431	-0.209660	0.033333	0.071666	-0.198251	3.000000																							

$	12-Dec-2018	18:58:24.849	Anchor	Radiometer Message																																	Satlantic OCR-504 Multispectral Radiometer 			

$	12-Dec-2018	18:58:25.072	Anchor	Radiometer Message																																	Copyright (C) 2011, Satlantic Inc. All rights reserved. 			

$	12-Dec-2018	18:58:25.085	Anchor	Radiometer Message																																	Firmware version: 5.1.0 - SatNet Type B 			

$	12-Dec-2018	18:58:25.193	Anchor	Radiometer Message																																	Instrument: SATFI4 			

$	12-Dec-2018	18:58:25.301	Anchor	Radiometer Message																																	S/N: 0287 			

$	12-Dec-2018	18:58:25.412	Anchor	Radiometer Message																																	Reset Source: Power 			

$	12-Dec-2018	18:58:25.522	Anchor	Radiometer Message																																	Press <Ctrl+C> for command console. 			

$	12-Dec-2018	18:58:25.630	Anchor	Radiometer Message																																	Initializing system. Please wait... 			

$	12-Dec-2018	18:58:25.941	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.030001	152.461834	30.350000	151.500000																																

$	12-Dec-2018	18:58:26.053	Anchor	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.9679294

$	12-Dec-2018	18:58:19.632	Anchor	SBE41 Message						   25.03  	25.030001	25.465678	-0.162548	0.000000	0.016666	-0.196628	3.000000																							

$	12-Dec-2018	18:58:19.949	Anchor	Optode Message																												4330,1532,265.881,22.270,29.198,29.129,36.165,7.035,539.0,468.4,117.8  								

$	12-Dec-2018	18:58:19.953	Anchor	Optode Message																												#  								

$	12-Dec-2018	18:58:26.408	Anchor	FLBB Message																											/99/99,99:99:99,695,2309,700,4130,536  									

$	12-Dec-2018	18:58:26.615	Anchor	Radiometer Message																																	Autonomous operation enabled. 			

$	12-Dec-2018	18:58:22.532	Anchor	Unparsed SBE41 Message						S>  																														

$	12-Dec-2018	18:58:24.947	Anchor	SBE41 Message						   25.02  	25.020000	25.422663	-0.107807	0.003333	0.001667	-0.193660	3.000000																							

$	12-Dec-2018	18:58:27.071	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.020000	152.479365	30.350000	151.500000																																

$	12-Dec-2018	18:58:27.182	Anchor	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.1239925

$	12-Dec-2018	18:58:27.487	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.020000	152.483650	30.350000	151.500000																																

$	12-Dec-2018	18:58:27.861	Anchor	SBE41 Message						   25.66  	25.660000	25.332868	-0.056846	-0.213333	-0.105000	-0.190060	3.000000																							

$	12-Dec-2018	18:58:28.418	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.660000	152.451105	30.350000	151.500000																																

$	12-Dec-2018	18:58:29.390	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.660000	152.448811	30.350000	151.500000																																

$	12-Dec-2018	18:58:30.393	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.660000	152.459934	30.350000	151.500000																																

$	12-Dec-2018	18:58:30.861	Anchor	SBE41 Message						   25.66  	25.660000	25.213096	-0.017826	0.000000	-0.106667	-0.185828	3.000000																							

$	12-Dec-2018	18:58:31.417	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.660000	152.448170	30.350000	151.500000																																

$	12-Dec-2018	18:58:32.391	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.660000	152.439788	30.350000	151.500000																																

$	12-Dec-2018	18:58:33.393	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.660000	152.473047	30.350000	151.500000																																

$	12-Dec-2018	18:58:33.861	Anchor	SBE41 Message						   25.64  	25.639999	25.123239	0.007976	0.006667	0.003333	-0.181064	3.000000																							

$	12-Dec-2018	18:58:34.421	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.639999	152.465686	30.350000	151.500000																																

$	12-Dec-2018	18:58:35.393	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.639999	152.462028	30.350000	151.500000																																

$	12-Dec-2018	18:58:36.394	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.639999	152.464539	30.350000	151.500000																																

$	12-Dec-2018	18:58:36.861	Anchor	SBE41 Message						   24.98  	24.980000	25.129949	0.023427	0.220000	0.113333	-0.175937	3.000000																							

$	12-Dec-2018	18:58:37.519	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.980000	152.469434	30.350000	151.500000																																

$	12-Dec-2018	18:58:38.395	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.980000	152.478382	30.350000	151.500000																																

$	12-Dec-2018	18:58:39.396	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.980000	152.473576	30.350000	151.500000																																

$	12-Dec-2018	18:58:39.860	Anchor	SBE41 Message						   24.84  	24.840000	25.251641	0.029656	0.046666	0.133333	-0.170614	3.000000																							

$	12-Dec-2018	18:58:40.521	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.840000	152.485952	30.350000	151.500000																																

$	12-Dec-2018	18:58:41.396	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.840000	152.465135	30.350000	151.500000																																

$	12-Dec-2018	18:58:42.396	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.840000	152.465940	30.350000	151.500000																																

$	12-Dec-2018	18:58:42.860	Anchor	SBE41 Message						   24.82  	24.820000	25.414672	0.023983	0.006667	0.026667	-0.165154	3.000000																							

$	12-Dec-2018	18:58:43.519	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.476735	30.350000	151.500000																																

$	12-Dec-2018	18:58:44.396	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.438141	30.350000	151.500000																																

$	12-Dec-2018	18:58:45.397	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.456917	30.350000	151.500000																																

$	12-Dec-2018	18:58:45.861	Anchor	SBE41 Message						   25.73  	25.730000	25.482745	0.005593	-0.303333	-0.148333	-0.159468	3.000000																							

$	12-Dec-2018	18:58:46.522	Anchor	PTH Message														794.487061	23.650000	54.016632	24.600601																			

$	12-Dec-2018	18:58:46.725	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.467400	30.425000	140.500000																																

$	12-Dec-2018	18:58:47.396	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.466781	30.425000	140.500000																																

$	12-Dec-2018	18:58:48.397	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.468473	30.425000	140.500000																																

$	12-Dec-2018	18:58:48.862	Anchor	SBE41 Message						   25.71  	25.709999	25.380036	-0.017176	0.006667	-0.148333	-0.153390	3.000000																							

$	12-Dec-2018	18:58:49.424	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.476795	30.425000	140.500000																																

$	12-Dec-2018	18:58:50.397	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.466230	30.425000	140.500000																																

$	12-Dec-2018	18:58:51.399	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.460627	30.425000	140.500000																																

$	12-Dec-2018	18:58:51.861	Anchor	SBE41 Message						   25.71  	25.709999	25.186587	-0.028813	0.000000	0.003333	-0.146869	3.000000																							

$	12-Dec-2018	18:58:52.425	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.479544	30.425000	140.500000																																

$	12-Dec-2018	18:58:53.398	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.469724	30.425000	140.500000																																

$	12-Dec-2018	18:58:54.398	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.475804	30.425000	140.500000																																

$	12-Dec-2018	18:58:54.862	Anchor	SBE41 Message						   24.79  	24.790001	25.083863	-0.019068	0.306666	0.153333	-0.140066	3.000000																							

$	12-Dec-2018	18:58:55.425	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.456664	30.425000	140.500000																																

$	12-Dec-2018	18:58:56.399	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.467534	30.425000	140.500000																																

$	12-Dec-2018	18:58:57.400	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.452253	30.425000	140.500000																																

$	12-Dec-2018	18:58:57.861	Anchor	SBE41 Message						   24.78  	24.780001	25.180168	0.005792	0.003333	0.155000	-0.133277	3.000000																							

$	12-Dec-2018	18:58:58.426	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.461812	30.425000	140.500000																																

$	12-Dec-2018	18:58:59.400	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.462080	30.425000	140.500000																																

$	12-Dec-2018	18:59:00.401	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.472727	30.425000	140.500000																																

$	12-Dec-2018	18:59:00.861	Anchor	SBE41 Message						   25.42  	25.420000	25.384125	0.025473	-0.213333	-0.105000	-0.126662	3.000000																							

$	12-Dec-2018	18:59:01.428	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.420000	152.466238	30.425000	140.500000																																

$	12-Dec-2018	18:59:02.401	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.420000	152.464449	30.425000	140.500000																																

$	12-Dec-2018	18:59:03.401	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.420000	152.457245	30.425000	140.500000																																

$	12-Dec-2018	18:59:03.862	Anchor	SBE41 Message						   25.76  	25.760000	25.484296	0.022530	-0.113333	-0.163333	-0.120093	3.000000																							

$	12-Dec-2018	18:59:04.525	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.760000	152.466968	30.425000	140.500000																																

$	12-Dec-2018	18:59:05.402	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.760000	152.463891	30.425000	140.500000																																

$	12-Dec-2018	18:59:06.402	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.760000	152.464263	30.425000	140.500000																																

$	12-Dec-2018	18:59:06.860	Anchor	SBE41 Message						   25.73  	25.730000	25.379095	-0.000473	0.010000	-0.051667	-0.113319	3.000000																							

$	12-Dec-2018	18:59:07.429	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.477920	30.425000	140.500000																																

$	12-Dec-2018	18:59:08.402	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.478978	30.425000	140.500000																																

$	12-Dec-2018	18:59:09.403	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.486391	30.425000	140.500000																																

$	12-Dec-2018	18:59:09.861	Anchor	SBE41 Message						   24.78  	24.780001	25.201844	-0.021755	0.316666	0.163333	-0.106267	3.000000																							

$	12-Dec-2018	18:59:10.431	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.454734	30.425000	140.500000																																

$	12-Dec-2018	18:59:11.403	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.476147	30.425000	140.500000																																

$	12-Dec-2018	18:59:12.403	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.475171	30.425000	140.500000																																

$	12-Dec-2018	18:59:12.860	Anchor	SBE41 Message						   24.78  	24.780001	25.169091	-0.021586	0.000000	0.158333	-0.099139	3.000000																							

$	12-Dec-2018	18:59:13.430	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.467616	30.425000	140.500000																																

$	12-Dec-2018	18:59:14.404	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.446315	30.425000	140.500000																																

$	12-Dec-2018	18:59:15.404	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.476765	30.425000	140.500000																																

$	12-Dec-2018	18:59:15.861	Anchor	SBE41 Message						   24.77  	24.770000	25.310381	-0.002642	0.003333	0.001667	-0.092171	3.000000																							

$	12-Dec-2018	18:59:16.433	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.770000	152.485862	30.425000	140.500000																																

$	12-Dec-2018	18:59:17.405	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.770000	152.477905	30.425000	140.500000																																

$	12-Dec-2018	18:59:18.007	Anchor	ERROR: Missed PH message 25

$	12-Dec-2018	18:59:18.011	Anchor	ERROR: Missed NO3 message 21

$	12-Dec-2018	18:59:18.213	Anchor	Sending Optode Do Sample

$	12-Dec-2018	18:59:18.217	Anchor	Sending Optode Wakeup

$	12-Dec-2018	18:59:18.618	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.770000	152.476095	30.425000	140.500000																																

$	12-Dec-2018	18:59:18.929	Anchor	Sending FLBB run command

$	12-Dec-2018	18:59:18.935	Anchor	Sending radiometer pollSample

$	12-Dec-2018	18:59:19.417	Anchor	Radiometer Message																																	SATFI40287,-0.0042,-0.0016,-0.0053,-0.1871 			

$	12-Dec-2018	18:59:19.733	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.770000	152.459294	30.425000	140.500000																																

$	12-Dec-2018	18:59:19.631	Anchor	SBE41 Message						   25.76  	25.760000	25.427204	0.013797	-0.330000	-0.163333	-0.085245	3.000000																							

$	12-Dec-2018	18:59:20.006	Anchor	Optode Message																												4330,1532,265.812,22.282,29.196,29.127,36.162,7.035,538.9,468.2,117.4  								

$	12-Dec-2018	18:59:20.009	Anchor	Optode Message																												#  								

$	12-Dec-2018	18:59:20.672	Anchor	FLBB Message																											/99/99,99:99:99,695,2312,700,4130,535  									

$	12-Dec-2018	18:59:21.081	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.760000	152.475111	30.425000	140.500000																																

$	12-Dec-2018	18:59:21.619	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.760000	152.463727	30.425000	140.500000																																

$	12-Dec-2018	18:59:22.394	Anchor	Unparsed SBE41 Message						S>  																														

$	12-Dec-2018	18:59:22.639	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.760000	152.459547	30.425000	140.500000																																

$	12-Dec-2018	18:59:23.620	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.760000	152.457923	30.425000	140.500000																																

$	12-Dec-2018	18:59:24.620	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.760000	152.464643	30.425000	140.500000																																

$	12-Dec-2018	18:59:24.862	Anchor	SBE41 Message						   24.98  	24.980000	25.345632	0.011877	0.260000	-0.035000	-0.077964	3.000000																							

$	12-Dec-2018	18:59:25.746	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.980000	152.447984	30.425000	140.500000																																

$	12-Dec-2018	18:59:26.621	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.980000	152.470410	30.425000	140.500000																																

$	12-Dec-2018	18:59:27.621	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.980000	152.479239	30.425000	140.500000																																

$	12-Dec-2018	18:59:27.861	Anchor	SBE41 Message						   24.78  	24.780001	25.133059	-0.001870	0.066666	0.163333	-0.070563	3.000000																							

$	12-Dec-2018	18:59:28.648	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.470142	30.425000	140.500000																																

$	12-Dec-2018	18:59:29.622	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.453311	30.425000	140.500000																																

$	12-Dec-2018	18:59:30.622	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.459495	30.425000	140.500000																																

$	12-Dec-2018	18:59:30.861	Anchor	SBE41 Message						   24.78  	24.780001	25.004524	-0.007513	0.000000	0.033333	-0.063128	3.000000																							

$	12-Dec-2018	18:59:31.651	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.457148	30.425000	140.500000																																

$	12-Dec-2018	18:59:32.623	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.472176	30.425000	140.500000																																

$	12-Dec-2018	18:59:33.623	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.479254	30.425000	140.500000																																

$	12-Dec-2018	18:59:33.860	Anchor	SBE41 Message						   25.74  	25.740000	25.043866	0.004428	-0.320000	-0.160000	-0.055839	3.000000																							

$	12-Dec-2018	18:59:34.650	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.740000	152.464949	30.425000	140.500000																																

$	12-Dec-2018	18:59:35.624	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.740000	152.450137	30.425000	140.500000																																

$	12-Dec-2018	18:59:36.624	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.740000	152.454153	30.425000	140.500000																																

$	12-Dec-2018	18:59:36.860	Anchor	SBE41 Message						   25.72  	25.719999	25.112697	0.022607	0.006667	-0.156666	-0.048722	3.000000																							

$	12-Dec-2018	18:59:37.749	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.474932	30.425000	140.500000																																

$	12-Dec-2018	18:59:38.625	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.446449	30.425000	140.500000																																

$	12-Dec-2018	18:59:39.624	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.468838	30.425000	140.500000																																

$	12-Dec-2018	18:59:39.861	Anchor	SBE41 Message						   25.65  	25.650000	25.083800	0.029600	0.023333	0.015000	-0.041687	3.000000																							

$	12-Dec-2018	18:59:40.651	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.650000	152.466185	30.425000	140.500000																																

$	12-Dec-2018	18:59:41.625	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.650000	152.478077	30.425000	140.500000																																

$	12-Dec-2018	18:59:42.625	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.650000	152.472094	30.425000	140.500000																																

$	12-Dec-2018	18:59:42.861	Anchor	SBE41 Message						   24.80  	24.799999	25.047483	0.021532	0.283333	0.153333	-0.034807	3.000000																							

$	12-Dec-2018	18:59:43.653	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.458422	30.425000	140.500000																																

$	12-Dec-2018	18:59:44.626	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.450278	30.425000	140.500000																																

$	12-Dec-2018	18:59:45.627	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.467996	30.425000	140.500000																																

$	12-Dec-2018	18:59:45.861	Anchor	SBE41 Message						   24.81  	24.809999	25.157800	0.007933	-0.003333	0.140000	-0.028382	3.000000																							

$	12-Dec-2018	18:59:46.653	Anchor	PTH Message														795.949219	23.833333	53.963226	24.611326																			

$	12-Dec-2018	18:59:46.856	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.809999	152.458422	30.450000	138.500000																																

$	12-Dec-2018	18:59:47.626	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.809999	152.453661	30.450000	138.500000																																

$	12-Dec-2018	18:59:48.627	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.809999	152.467556	30.450000	138.500000																																

$	12-Dec-2018	18:59:48.861	Anchor	SBE41 Message						   25.68  	25.680000	25.368559	-0.003306	-0.290000	-0.146667	-0.022658	3.000000																							

$	12-Dec-2018	18:59:49.653	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.680000	152.461052	30.450000	138.500000																																

$	12-Dec-2018	18:59:50.627	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.680000	152.476460	30.450000	138.500000																																

$	12-Dec-2018	18:59:51.627	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.680000	152.449347	30.450000	138.500000																																

$	12-Dec-2018	18:59:51.861	Anchor	SBE41 Message						   25.71  	25.709999	25.472680	-0.013212	-0.010000	-0.150000	-0.017593	3.000000																							

$	12-Dec-2018	18:59:52.654	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.468726	30.450000	138.500000																																

$	12-Dec-2018	18:59:53.628	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.486756	30.450000	138.500000																																

$	12-Dec-2018	18:59:54.628	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.468584	30.450000	138.500000																																

$	12-Dec-2018	18:59:54.860	Anchor	SBE41 Message						   25.72  	25.719999	25.375258	-0.024161	-0.003333	-0.006667	-0.012997	3.000000																							

$	12-Dec-2018	18:59:55.654	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.476139	30.450000	138.500000																																

$	12-Dec-2018	18:59:56.627	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.458377	30.450000	138.500000																																

$	12-Dec-2018	18:59:57.627	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.476341	30.450000	138.500000																																

$	12-Dec-2018	18:59:57.861	Anchor	SBE41 Message						   24.83  	24.830000	25.224485	-0.030789	0.296666	0.146667	-0.008861	3.000000																							

$	12-Dec-2018	18:59:58.654	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.830000	152.477838	30.450000	138.500000																																

$	12-Dec-2018	18:59:59.628	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.830000	152.459383	30.450000	138.500000																																

$	12-Dec-2018	19:00:00.628	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.830000	152.461864	30.450000	138.500000																																

$	12-Dec-2018	19:00:00.861	Anchor	SBE41 Message						   24.82  	24.820000	25.227743	-0.024249	0.003333	0.150000	-0.005409	3.000000																							

$	12-Dec-2018	19:00:01.753	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.460814	30.450000	138.500000																																

$	12-Dec-2018	19:00:02.629	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.465850	30.450000	138.500000																																

$	12-Dec-2018	19:00:03.629	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.450815	30.450000	138.500000																																

$	12-Dec-2018	19:00:03.862	Anchor	SBE41 Message						   24.81  	24.809999	25.381525	-0.006362	0.003333	0.003333	-0.002845	3.000000																							

$	12-Dec-2018	19:00:04.753	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.809999	152.466282	30.450000	138.500000																																

$	12-Dec-2018	19:00:05.631	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.809999	152.448967	30.450000	138.500000																																

$	12-Dec-2018	19:00:06.630	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.809999	152.460024	30.450000	138.500000																																

$	12-Dec-2018	19:00:06.860	Anchor	SBE41 Message						   25.72  	25.719999	25.478448	0.007477	-0.303333	-0.150000	-0.001132	3.000000																							

$	12-Dec-2018	19:00:07.756	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.446963	30.450000	138.500000																																

$	12-Dec-2018	19:00:08.632	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.463943	30.450000	138.500000																																

$	12-Dec-2018	19:00:09.632	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.458519	30.450000	138.500000																																

$	12-Dec-2018	19:00:09.860	Anchor	SBE41 Message						   25.71  	25.709999	25.374975	0.005188	0.003333	-0.150000	-0.000040	3.000000																							

$	12-Dec-2018	19:00:10.756	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.472317	30.450000	138.500000																																

$	12-Dec-2018	19:00:11.633	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.470663	30.450000	138.500000																																

$	12-Dec-2018	19:00:12.634	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.461283	30.450000	138.500000																																

$	12-Dec-2018	19:00:12.861	Anchor	SBE41 Message						   25.64  	25.639999	25.160293	-0.006254	0.023333	0.013333	0.000643	3.000000																							

$	12-Dec-2018	19:00:13.663	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.639999	152.473070	30.450000	138.500000																																

$	12-Dec-2018	19:00:14.635	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.639999	152.459092	30.450000	138.500000																																

$	12-Dec-2018	19:00:15.636	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.639999	152.465329	30.450000	138.500000																																

$	12-Dec-2018	19:00:15.861	Anchor	SBE41 Message						   24.82  	24.820000	25.050118	-0.008149	0.273333	0.148333	0.000982	3.000000																							

$	12-Dec-2018	19:00:16.679	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.460977	30.450000	138.500000																																

$	12-Dec-2018	19:00:17.635	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.451649	30.450000	138.500000																																

$	12-Dec-2018	19:00:18.008	Anchor	ERROR: Missed PH message 26

$	12-Dec-2018	19:00:18.110	Anchor	ERROR: Missed NO3 message 22

$	12-Dec-2018	19:00:18.214	Anchor	Sending Optode Do Sample

$	12-Dec-2018	19:00:18.218	Anchor	Sending Optode Wakeup

$	12-Dec-2018	19:00:18.635	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.464882	30.450000	138.500000																																

$	12-Dec-2018	19:00:18.820	Anchor	Sending FLBB run command

$	12-Dec-2018	19:00:18.935	Anchor	Sending radiometer pollSample

$	12-Dec-2018	19:00:19.434	Anchor	Radiometer Message																																	SATFI40287,-0.0056,0.0005,-0.0058,-0.2216 			

$	12-Dec-2018	19:00:19.546	Anchor	SBE41 Message						   24.82  	24.820000	25.155566	0.007073	0.000000	0.136667	0.000948	3.000000																							

$	12-Dec-2018	19:00:19.779	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.493484	30.450000	138.500000																																

$	12-Dec-2018	19:00:19.911	Anchor	Optode Message																												4330,1532,265.636,22.303,29.195,29.126,36.161,7.035,538.8,467.8,116.7  								

$	12-Dec-2018	19:00:19.914	Anchor	Optode Message																												#  								

$	12-Dec-2018	19:00:20.564	Anchor	FLBB Message																											/99/99,99:99:99,695,2309,700,4130,534  									

$	12-Dec-2018	19:00:20.974	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.448401	30.450000	138.500000																																

$	12-Dec-2018	19:00:21.636	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.481057	30.450000	138.500000																																

$	12-Dec-2018	19:00:22.394	Anchor	Unparsed SBE41 Message						S>  																														

$	12-Dec-2018	19:00:22.656	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.464017	30.450000	138.500000																																

$	12-Dec-2018	19:00:23.636	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.457729	30.450000	138.500000																																

$	12-Dec-2018	19:00:24.637	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.465098	30.450000	138.500000																																

$	12-Dec-2018	19:00:24.861	Anchor	SBE41 Message						   25.73  	25.730000	25.367134	0.024754	-0.303333	-0.151667	0.000571	3.000000																							

$	12-Dec-2018	19:00:25.761	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.469642	30.450000	138.500000																																

$	12-Dec-2018	19:00:26.638	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.459636	30.450000	138.500000																																

$	12-Dec-2018	19:00:27.639	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.476974	30.450000	138.500000																																

$	12-Dec-2018	19:00:27.860	Anchor	SBE41 Message						   25.75  	25.750000	25.469080	0.023447	-0.006667	-0.155000	0.000075	3.000000																							

$	12-Dec-2018	19:00:28.668	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.750000	152.493894	30.450000	138.500000																																

$	12-Dec-2018	19:00:29.640	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.750000	152.479135	30.450000	138.500000																																

$	12-Dec-2018	19:00:30.640	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.750000	152.457871	30.450000	138.500000																																

$	12-Dec-2018	19:00:30.861	Anchor	SBE41 Message						   24.80  	24.799999	25.375662	0.000715	0.316667	0.155000	-0.000246	3.000000																							

$	12-Dec-2018	19:00:31.762	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.461663	30.450000	138.500000																																

$	12-Dec-2018	19:00:32.639	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.453728	30.450000	138.500000																																

$	12-Dec-2018	19:00:33.639	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.470402	30.450000	138.500000																																

$	12-Dec-2018	19:00:33.861	Anchor	SBE41 Message						   24.81  	24.809999	25.232997	-0.022228	-0.003333	0.156667	-0.000292	3.000000																							

$	12-Dec-2018	19:00:34.765	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.809999	152.468234	30.450000	138.500000																																

$	12-Dec-2018	19:00:35.641	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.809999	152.471647	30.450000	138.500000																																

$	12-Dec-2018	19:00:36.641	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.809999	152.468525	30.450000	138.500000																																

$	12-Dec-2018	19:00:36.862	Anchor	SBE41 Message						   24.80  	24.799999	25.227122	-0.024505	0.003333	0.000000	-0.000194	3.000000																							

$	12-Dec-2018	19:00:37.765	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.487434	30.450000	138.500000																																

$	12-Dec-2018	19:00:38.640	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.472667	30.450000	138.500000																																

$	12-Dec-2018	19:00:39.640	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.465299	30.450000	138.500000																																

$	12-Dec-2018	19:00:39.861	Anchor	SBE41 Message						   25.73  	25.730000	25.314717	-0.008369	-0.310000	-0.153333	-0.000090	3.000000																							

$	12-Dec-2018	19:00:40.669	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.468041	30.450000	138.500000																																

$	12-Dec-2018	19:00:41.641	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.480952	30.450000	138.500000																																

$	12-Dec-2018	19:00:42.641	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.464457	30.450000	138.500000																																

$	12-Dec-2018	19:00:42.862	Anchor	SBE41 Message						   25.72  	25.719999	25.294418	0.007537	0.003333	-0.153333	0.000041	3.000000																							

$	12-Dec-2018	19:00:43.667	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.470268	30.450000	138.500000																																

$	12-Dec-2018	19:00:44.641	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.456544	30.450000	138.500000																																

$	12-Dec-2018	19:00:45.641	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.460858	30.450000	138.500000																																

$	12-Dec-2018	19:00:45.861	Anchor	SBE41 Message						   25.50  	25.500000	25.136723	0.012105	0.073333	0.038333	0.000218	3.000000																							

$	12-Dec-2018	19:00:46.669	Anchor	PTH Message														796.848877	23.625000	53.940338	24.622051																			

$	12-Dec-2018	19:00:46.872	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.500000	152.461812	30.450000	139.000000																																

$	12-Dec-2018	19:00:47.642	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.500000	152.458891	30.450000	139.000000																																

$	12-Dec-2018	19:00:48.642	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.500000	152.462311	30.450000	139.000000																																

$	12-Dec-2018	19:00:48.860	Anchor	SBE41 Message						   24.80  	24.799999	25.040852	0.010477	0.233334	0.153333	0.000302	3.000000																							

$	12-Dec-2018	19:00:49.766	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.476944	30.450000	139.000000																																

$	12-Dec-2018	19:00:50.643	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.451582	30.450000	139.000000																																

$	12-Dec-2018	19:00:51.644	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.456120	30.450000	139.000000																																

$	12-Dec-2018	19:00:51.860	Anchor	SBE41 Message						   24.82  	24.820000	25.147959	0.011551	-0.006667	0.113333	0.000158	3.000000																							

$	12-Dec-2018	19:00:52.673	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.466461	30.450000	139.000000																																

$	12-Dec-2018	19:00:53.644	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.462795	30.450000	139.000000																																

$	12-Dec-2018	19:00:54.644	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.484305	30.450000	139.000000																																

$	12-Dec-2018	19:00:54.860	Anchor	SBE41 Message						   25.71  	25.709999	25.353594	0.014475	-0.296666	-0.151667	-0.000117	3.000000																							

$	12-Dec-2018	19:00:55.671	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.465932	30.450000	139.000000																																

$	12-Dec-2018	19:00:56.644	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.478591	30.450000	139.000000																																

$	12-Dec-2018	19:00:57.644	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.484745	30.450000	139.000000																																

$	12-Dec-2018	19:00:57.860	Anchor	SBE41 Message						   25.71  	25.709999	25.439209	0.009825	0.000000	-0.148333	-0.000259	3.000000																							

$	12-Dec-2018	19:00:58.674	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.473383	30.450000	139.000000																																

$	12-Dec-2018	19:00:59.645	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.472720	30.450000	139.000000																																

$	12-Dec-2018	19:01:00.645	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.458295	30.450000	139.000000																																

$	12-Dec-2018	19:01:00.861	Anchor	SBE41 Message						   25.73  	25.730000	25.332791	-0.006226	-0.006667	-0.003333	-0.000139	3.000000																							

$	12-Dec-2018	19:01:01.672	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.471870	30.450000	139.000000																																

$	12-Dec-2018	19:01:02.646	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.478524	30.450000	139.000000																																

$	12-Dec-2018	19:01:03.646	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.457871	30.450000	139.000000																																

$	12-Dec-2018	19:01:03.861	Anchor	SBE41 Message						   24.79  	24.790001	25.195736	-0.021912	0.313333	0.153333	0.000035	3.000000																							

$	12-Dec-2018	19:01:04.675	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.471453	30.450000	139.000000																																

$	12-Dec-2018	19:01:05.648	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.476929	30.450000	139.000000																																

$	12-Dec-2018	19:01:06.648	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.473308	30.450000	139.000000																																

$	12-Dec-2018	19:01:06.862	Anchor	SBE41 Message						   24.82  	24.820000	25.222254	-0.021263	-0.010000	0.151667	-0.000069	3.000000																							

$	12-Dec-2018	19:01:07.675	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.468480	30.450000	139.000000																																

$	12-Dec-2018	19:01:08.649	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.466811	30.450000	139.000000																																

$	12-Dec-2018	19:01:09.649	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	152.473100	30.450000	139.000000																																

$	12-Dec-2018	19:01:09.861	Anchor	SBE41 Message						   24.80  	24.799999	25.388989	-0.005563	0.006667	-0.001666	-0.000498	3.000000																							

$	12-Dec-2018	19:01:10.775	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.469225	30.450000	139.000000																																

$	12-Dec-2018	19:01:11.651	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.488485	30.450000	139.000000																																

$	12-Dec-2018	19:01:12.651	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.468532	30.450000	139.000000																																

$	12-Dec-2018	19:01:12.862	Anchor	SBE41 Message						   25.73  	25.730000	25.484085	0.006458	-0.310000	-0.151667	-0.000914	3.000000																							

$	12-Dec-2018	19:01:13.775	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.477868	30.450000	139.000000																																

$	12-Dec-2018	19:01:14.666	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.475476	30.450000	139.000000																																

$	12-Dec-2018	19:01:15.666	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.461223	30.450000	139.000000																																

$	12-Dec-2018	19:01:15.862	Anchor	SBE41 Message						   25.73  	25.730000	25.371935	0.002006	0.000000	-0.155000	-0.000979	3.000000																							

$	12-Dec-2018	19:01:16.696	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.456574	30.450000	139.000000																																

$	12-Dec-2018	19:01:17.667	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.473442	30.450000	139.000000																																

$	12-Dec-2018	19:01:18.008	Anchor	ERROR: Missed PH message 27

$	12-Dec-2018	19:01:18.111	Anchor	ERROR: Missed NO3 message 23

$	12-Dec-2018	19:01:18.215	Anchor	Sending Optode Do Sample

$	12-Dec-2018	19:01:18.219	Anchor	Sending Optode Wakeup

$	12-Dec-2018	19:01:18.668	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.469091	30.450000	139.000000																																

$	12-Dec-2018	19:01:18.821	Anchor	Sending FLBB run command

$	12-Dec-2018	19:01:18.838	Anchor	Sending radiometer pollSample

$	12-Dec-2018	19:01:19.466	Anchor	Radiometer Message																																	SATFI40287,-0.0051,0.0008,-0.0061,-0.2163 			

$	12-Dec-2018	19:01:19.578	Anchor	SBE41 Message						   24.88  	24.879999	25.150888	-0.010091	0.283333	0.141667	-0.000736	3.000000																							

$	12-Dec-2018	19:01:19.810	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.879999	152.452789	30.450000	139.000000																																

$	12-Dec-2018	19:01:19.922	Anchor	Optode Message																												4330,1532,265.627,22.329,29.185,29.116,36.156,7.040,538.6,467.5,115.9  								

$	12-Dec-2018	19:01:19.922	Anchor	Optode Message																												#  								

$	12-Dec-2018	19:01:20.596	Anchor	FLBB Message																											/99/99,99:99:99,695,2311,700,4130,534  									

$	12-Dec-2018	19:01:21.006	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.879999	152.480900	30.450000	139.000000																																

$	12-Dec-2018	19:01:21.669	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.879999	152.469270	30.450000	139.000000																																

$	12-Dec-2018	19:01:22.394	Anchor	Unparsed SBE41 Message						S>  																														

$	12-Dec-2018	19:01:22.689	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.879999	152.475968	30.450000	139.000000																																

$	12-Dec-2018	19:01:23.670	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.879999	152.474388	30.450000	139.000000																																

$	12-Dec-2018	19:01:24.670	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.879999	152.465709	30.450000	139.000000																																

$	12-Dec-2018	19:01:24.861	Anchor	SBE41 Message						   24.78  	24.780001	25.039796	-0.010432	0.033333	0.158333	-0.000530	3.000000																							

$	12-Dec-2018	19:01:25.699	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.464889	30.450000	139.000000																																

$	12-Dec-2018	19:01:26.670	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.468301	30.450000	139.000000																																

$	12-Dec-2018	19:01:27.670	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.780001	152.465023	30.450000	139.000000																																

$	12-Dec-2018	19:01:27.861	Anchor	SBE41 Message						   25.73  	25.730000	25.134780	0.006994	-0.316666	-0.141667	-0.000599	3.000000																							

$	12-Dec-2018	19:01:28.794	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.473457	30.450000	139.000000																																

$	12-Dec-2018	19:01:29.671	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.473383	30.450000	139.000000																																

$	12-Dec-2018	19:01:30.671	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.475618	30.450000	139.000000																																

$	12-Dec-2018	19:01:30.861	Anchor	SBE41 Message						   25.73  	25.730000	25.294725	0.025884	0.000000	-0.158333	-0.000769	3.000000																							

$	12-Dec-2018	19:01:31.700	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.464673	30.450000	139.000000																																

$	12-Dec-2018	19:01:32.673	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.460612	30.450000	139.000000																																

$	12-Dec-2018	19:01:33.673	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.453542	30.450000	139.000000																																

$	12-Dec-2018	19:01:33.860	Anchor	SBE41 Message						   25.71  	25.709999	25.322093	0.026146	0.006667	0.003333	-0.000718	3.000000																							

$	12-Dec-2018	19:01:34.796	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.467854	30.450000	139.000000																																

$	12-Dec-2018	19:01:35.673	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.463593	30.450000	139.000000																																

$	12-Dec-2018	19:01:36.674	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.469933	30.450000	139.000000																																

$	12-Dec-2018	19:01:36.859	Anchor	SBE41 Message						   24.80  	24.799999	25.243203	0.008048	0.303333	0.155000	-0.000410	3.000000																							

$	12-Dec-2018	19:01:37.703	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.460590	30.450000	139.000000																																

$	12-Dec-2018	19:01:38.674	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.451955	30.450000	139.000000																																

$	12-Dec-2018	19:01:39.675	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.480081	30.450000	139.000000																																

$	12-Dec-2018	19:01:39.861	Anchor	SBE41 Message						   24.79  	24.790001	25.256533	-0.010185	0.003333	0.153333	-0.000152	3.000000																							

$	12-Dec-2018	19:01:40.799	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.460806	30.450000	139.000000																																

$	12-Dec-2018	19:01:41.675	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.443618	30.450000	139.000000																																

$	12-Dec-2018	19:01:42.675	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.476348	30.450000	139.000000																																

$	12-Dec-2018	19:01:42.861	Anchor	SBE41 Message						   25.56  	25.559999	25.402147	-0.016073	-0.256666	-0.126667	-0.000133	3.000000																							

$	12-Dec-2018	19:01:43.704	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.559999	152.473122	30.450000	139.000000																																

$	12-Dec-2018	19:01:44.676	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.559999	152.458362	30.450000	139.000000																																

$	12-Dec-2018	19:01:45.677	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.559999	152.472295	30.450000	139.000000																																

$	12-Dec-2018	19:01:45.861	Anchor	SBE41 Message						   25.72  	25.719999	25.490149	-0.013779	-0.053333	-0.155000	-0.000164	3.000000																							

$	12-Dec-2018	19:01:46.800	Anchor	PTH Message														795.679688	23.666667	53.909821	24.632776																			

$	12-Dec-2018	19:01:47.002	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.453437	NaN	138.500000																																

$	12-Dec-2018	19:01:47.676	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.469955	NaN	138.500000																																

$	12-Dec-2018	19:01:48.676	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.462475	NaN	138.500000																																

$	12-Dec-2018	19:01:48.860	Anchor	SBE41 Message						   25.72  	25.719999	25.389303	-0.012497	0.000000	-0.026667	-0.000011	3.000000																							

$	12-Dec-2018	19:01:49.811	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.463265	NaN	138.500000																																

$	12-Dec-2018	19:01:50.676	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.463064	NaN	138.500000																																

$	12-Dec-2018	19:01:51.676	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.719999	152.455226	NaN	138.500000																																

$	12-Dec-2018	19:01:51.861	Anchor	SBE41 Message						   24.86  	24.860001	25.222292	-0.012938	0.286666	0.143333	0.000222	3.000000																							

$	12-Dec-2018	19:01:52.800	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.860001	152.470753	NaN	138.500000																																

$	12-Dec-2018	19:01:53.676	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.860001	152.467445	NaN	138.500000																																

$	12-Dec-2018	19:01:54.676	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.860001	152.462535	NaN	138.500000																																

$	12-Dec-2018	19:01:54.861	Anchor	SBE41 Message						   24.80  	24.799999	25.202227	-0.008295	0.020000	0.153333	0.000229	3.000000																							

$	12-Dec-2018	19:01:55.705	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.456708	NaN	138.500000																																

$	12-Dec-2018	19:01:56.677	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.468853	NaN	138.500000																																

$	12-Dec-2018	19:01:57.676	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.483597	NaN	138.500000																																

$	12-Dec-2018	19:01:57.860	Anchor	SBE41 Message						   24.79  	24.790001	25.346421	0.003521	0.003333	0.011667	-0.000065	3.000000																							

$	12-Dec-2018	19:01:58.801	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.465396	NaN	138.500000																																

$	12-Dec-2018	19:01:59.677	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.461603	NaN	138.500000																																

$	12-Dec-2018	19:02:00.677	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.474627	NaN	138.500000																																

$	12-Dec-2018	19:02:00.861	Anchor	SBE41 Message						   25.73  	25.730000	25.454873	0.012942	-0.313333	-0.155000	-0.000373	3.000000																							

$	12-Dec-2018	19:02:01.802	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.469188	NaN	138.500000																																

$	12-Dec-2018	19:02:02.678	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.457818	NaN	138.500000																																

$	12-Dec-2018	19:02:03.678	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.473360	NaN	138.500000																																

$	12-Dec-2018	19:02:03.860	Anchor	SBE41 Message						   25.73  	25.730000	25.368779	0.009174	0.000000	-0.156666	-0.000396	3.000000																							

$	12-Dec-2018	19:02:04.803	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.475432	NaN	138.500000																																

$	12-Dec-2018	19:02:05.679	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.477801	NaN	138.500000																																

$	12-Dec-2018	19:02:06.679	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.730000	152.458481	NaN	138.500000																																

$	12-Dec-2018	19:02:06.861	Anchor	SBE41 Message						   25.63  	25.629999	25.160857	-0.003484	0.033333	0.016667	-0.000177	3.000000																							

$	12-Dec-2018	19:02:07.803	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.629999	152.486876	NaN	138.500000																																

$	12-Dec-2018	19:02:08.680	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.629999	152.486436	NaN	138.500000																																

$	12-Dec-2018	19:02:09.679	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.629999	152.466618	NaN	138.500000																																

$	12-Dec-2018	19:02:09.861	Anchor	SBE41 Message						   24.80  	24.799999	25.048013	-0.007739	0.276667	0.155000	-0.000021	3.000000																							

$	12-Dec-2018	19:02:10.804	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.488649	NaN	138.500000																																

$	12-Dec-2018	19:02:11.680	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.474009	NaN	138.500000																																

$	12-Dec-2018	19:02:12.680	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.491346	NaN	138.500000																																

$	12-Dec-2018	19:02:12.876	Anchor	SBE41 Message						   24.80  	24.799999	25.152319	0.005182	0.000000	0.138333	-0.000125	3.000000																							

$	12-Dec-2018	19:02:13.804	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.467348	NaN	138.500000																																

$	12-Dec-2018	19:02:14.681	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.453929	NaN	138.500000																																

$	12-Dec-2018	19:02:15.680	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	152.485535	NaN	138.500000																																

$	12-Dec-2018	19:02:15.861	Anchor	SBE41 Message						   25.71  	25.709999	25.367184	0.022064	-0.303333	-0.151667	-0.000367	3.000000																							

$	12-Dec-2018	19:02:16.709	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.469032	NaN	138.500000																																

$	12-Dec-2018	19:02:17.681	Anchor	Normal State Exit

$	12-Dec-2018	19:02:17.796	Anchor	ERROR: Missed PH message 28

$	12-Dec-2018	19:02:17.800	Anchor	ERROR: Missed NO3 message 24

$	12-Dec-2018	19:02:18.002	Anchor	Sending Optode Do Sample

$	12-Dec-2018	19:02:18.007	Anchor	Sending Optode Wakeup

$	12-Dec-2018	19:02:17.685	Anchor	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.459182	NaN	138.500000																																

$	12-Dec-2018	19:02:18.478	Start CP Mode	State Entry

$	12-Dec-2018	19:02:18.483	Start CP Mode	Stopping SBE41 Timer

$	12-Dec-2018	19:02:18.707	Start CP Mode	Sending FLBB run command

$	12-Dec-2018	19:02:18.724	Start CP Mode	Sending radiometer pollSample

$	12-Dec-2018	19:02:18.933	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.709999	152.466305	NaN	138.500000																																

$	12-Dec-2018	19:02:19.479	Start CP Mode	Radiometer Message																																	SATFI40287,-0.0045,0.0001,-0.0066,-0.2024 			

$	12-Dec-2018	19:02:19.402	Start CP Mode	SBE41 Message						   25.77  	25.770000	25.469887	0.021491	-0.020000	-0.161667	-0.000462	3.000000																							

$	12-Dec-2018	19:02:19.922	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.474239	NaN	138.500000																																

$	12-Dec-2018	19:02:20.478	Start CP Mode	SBE41 Timer Stopped, Sending SBE41 CR/LF

$	12-Dec-2018	19:02:19.797	Start CP Mode	Optode Message																												4330,1532,265.404,22.357,29.183,29.113,36.152,7.039,538.4,467.2,115.0  								

$	12-Dec-2018	19:02:19.800	Start CP Mode	Optode Message																												#  								

$	12-Dec-2018	19:02:20.710	Start CP Mode	FLBB Message																											/99/99,99:99:99,695,2309,700,4130,534  									

$	12-Dec-2018	19:02:21.120	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.478725	NaN	138.500000																																

$	12-Dec-2018	19:02:21.481	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:21.779	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.455516	NaN	138.500000																																

$	12-Dec-2018	19:02:22.478	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:22.782	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.472079	NaN	138.500000																																

$	12-Dec-2018	19:02:23.479	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:23.687	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.471311	NaN	138.500000																																

$	12-Dec-2018	19:02:24.479	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:24.783	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.495943	NaN	138.500000																																

$	12-Dec-2018	19:02:25.479	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:25.783	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.456738	NaN	138.500000																																

$	12-Dec-2018	19:02:26.479	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:26.686	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.462117	NaN	138.500000																																

$	12-Dec-2018	19:02:27.480	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:27.686	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.466230	NaN	138.500000																																

$	12-Dec-2018	19:02:28.480	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:28.784	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.472764	NaN	138.500000																																

$	12-Dec-2018	19:02:29.480	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:29.784	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.483016	NaN	138.500000																																

$	12-Dec-2018	19:02:30.480	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:30.784	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.436696	NaN	138.500000																																

$	12-Dec-2018	19:02:31.480	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:31.784	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.458221	NaN	138.500000																																

$	12-Dec-2018	19:02:32.480	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:32.689	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.481481	NaN	138.500000																																

$	12-Dec-2018	19:02:33.481	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:33.784	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.463011	NaN	138.500000																																

$	12-Dec-2018	19:02:34.480	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:34.784	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.470395	NaN	138.500000																																

$	12-Dec-2018	19:02:35.480	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:35.687	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.462408	NaN	138.500000																																

$	12-Dec-2018	19:02:36.481	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:36.688	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.471147	NaN	138.500000																																

$	12-Dec-2018	19:02:37.481	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:37.785	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.481556	NaN	138.500000																																

$	12-Dec-2018	19:02:38.481	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:38.688	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.490698	NaN	138.500000																																

$	12-Dec-2018	19:02:39.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:39.786	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.474910	NaN	138.500000																																

$	12-Dec-2018	19:02:40.481	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:40.688	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.458824	NaN	138.500000																																

$	12-Dec-2018	19:02:41.481	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:41.690	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.487718	NaN	138.500000																																

$	12-Dec-2018	19:02:42.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:42.786	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.450301	NaN	138.500000																																

$	12-Dec-2018	19:02:43.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:43.689	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.476691	NaN	138.500000																																

$	12-Dec-2018	19:02:44.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:44.786	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.480513	NaN	138.500000																																

$	12-Dec-2018	19:02:45.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:45.786	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.469903	NaN	138.500000																																

$	12-Dec-2018	19:02:46.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:46.786	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.473859	NaN	138.500000																																

$	12-Dec-2018	19:02:47.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:47.785	Start CP Mode	PTH Message														795.757324	23.766667	53.909821	24.632776																			

$	12-Dec-2018	19:02:47.987	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.461603	30.450000	139.000000																																

$	12-Dec-2018	19:02:48.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:48.688	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.453437	30.450000	139.000000																																

$	12-Dec-2018	19:02:49.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:49.689	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.474254	30.450000	139.000000																																

$	12-Dec-2018	19:02:50.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:50.786	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.467273	30.450000	139.000000																																

$	12-Dec-2018	19:02:51.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:51.688	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.465343	30.450000	139.000000																																

$	12-Dec-2018	19:02:52.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:52.786	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.473204	30.450000	139.000000																																

$	12-Dec-2018	19:02:53.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:53.691	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.462058	30.450000	139.000000																																

$	12-Dec-2018	19:02:54.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:54.690	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.471215	30.450000	139.000000																																

$	12-Dec-2018	19:02:55.483	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:55.689	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.480066	30.450000	139.000000																																

$	12-Dec-2018	19:02:56.483	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:56.690	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.487479	30.450000	139.000000																																

$	12-Dec-2018	19:02:57.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:57.786	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.466223	30.450000	139.000000																																

$	12-Dec-2018	19:02:58.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:58.786	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.469002	30.450000	139.000000																																

$	12-Dec-2018	19:02:59.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:02:59.688	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.469113	30.450000	139.000000																																

$	12-Dec-2018	19:03:00.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:03:00.691	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.456999	30.450000	139.000000																																

$	12-Dec-2018	19:03:01.482	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:03:01.689	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.466439	30.450000	139.000000																																

$	12-Dec-2018	19:03:02.483	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:03:02.790	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.463935	30.450000	139.000000																																

$	12-Dec-2018	19:03:03.483	Start CP Mode	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:03:02.588	Start CP Mode	Unparsed SBE41 Message						S>  																														

$	12-Dec-2018	19:03:03.806	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.478129	30.450000	139.000000																																

$	12-Dec-2018	19:03:04.483	Start CP Mode	Got CR/LF response, sending startprofile

$	12-Dec-2018	19:03:04.787	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.466767	30.450000	139.000000																																

$	12-Dec-2018	19:03:05.483	Start CP Mode	Didn't get startprofile response, retrying

$	12-Dec-2018	19:03:05.506	Start CP Mode	Calculated new PID Constant, h = 3000

$	12-Dec-2018	19:03:05.604	Start CP Mode	Unparsed SBE41 Message						S>startprofile5  																														

$	12-Dec-2018	19:03:05.928	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.478978	30.450000	139.000000																																

$	12-Dec-2018	19:03:06.483	Start CP Mode	Didn't get startprofile response, retrying

$	12-Dec-2018	19:03:06.587	Start CP Mode	Calculated new PID Constant, h = 3000

$	12-Dec-2018	19:03:06.794	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.468637	30.450000	139.000000																																

$	12-Dec-2018	19:03:07.483	Start CP Mode	Didn't get startprofile response, retrying

$	12-Dec-2018	19:03:07.586	Start CP Mode	Calculated new PID Constant, h = 3000

$	12-Dec-2018	19:03:07.794	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.471974	30.450000	139.000000																																

$	12-Dec-2018	19:03:08.483	Start CP Mode	Didn't get startprofile response, retrying

$	12-Dec-2018	19:03:08.488	Start CP Mode	Calculated new PID Constant, h = 3000

$	12-Dec-2018	19:03:08.792	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.471647	30.450000	139.000000																																

$	12-Dec-2018	19:03:09.483	Start CP Mode	Didn't get startprofile response, retrying

$	12-Dec-2018	19:03:09.488	Start CP Mode	Calculated new PID Constant, h = 3000

$	12-Dec-2018	19:03:09.381	Start CP Mode	Unparsed SBE41 Message						profile started, pump delay = 5 seconds  																														

$	12-Dec-2018	19:03:10.007	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.475208	30.450000	139.000000																																

$	12-Dec-2018	19:03:10.498	Start CP Mode	Got startprofile response

$	12-Dec-2018	19:03:10.600	Start CP Mode	Calculated new PID Constant, h = 3000

$	12-Dec-2018	19:03:10.702	Start CP Mode	Starting valve open

$	12-Dec-2018	19:03:10.908	Start CP Mode	Normal State Exit

$	12-Dec-2018	19:03:10.912	Start CP Mode	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.456313	30.450000	139.000000																																

$	12-Dec-2018	19:03:11.497	DeAnchor	State Entry

$	12-Dec-2018	19:03:11.604	DeAnchor	Deanchor JV = 200000.0

$	12-Dec-2018	19:03:11.608	DeAnchor	Initializing SBE 41 LPF

$	12-Dec-2018	19:03:11.812	DeAnchor	Starting valve open

$	12-Dec-2018	19:03:12.122	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.454734	30.450000	139.000000																																

$	12-Dec-2018	19:03:12.701	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.454033	30.450000	139.000000																																

$	12-Dec-2018	19:03:13.700	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.485155	30.450000	139.000000																																

$	12-Dec-2018	19:03:14.700	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.453676	30.450000	139.000000																																

$	12-Dec-2018	19:03:15.137	DeAnchor	Unparsed SBE41 Message						S>   24.79, 21.9864, 34.1360  																														

$	12-Dec-2018	19:03:16.041	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.770000	152.454086	30.450000	139.000000																																

$	12-Dec-2018	19:03:16.152	DeAnchor	SBE41 Message						   24.79, 21.9693, 34.1539  	24.790001	25.769985	0.000002	0.980000	0.490000	-0.000000	1.000000																							

$	12-Dec-2018	19:03:16.831	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.223684	30.450000	139.000000																																

$	12-Dec-2018	19:03:17.134	DeAnchor	SBE41 Message						   24.79, 21.9630, 34.1582  	24.790001	25.767095	0.000359	0.000000	0.490000	0.000013	1.000000																							

$	12-Dec-2018	19:03:17.830	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	152.124174	30.450000	139.000000																																

$	12-Dec-2018	19:03:18.133	DeAnchor	SBE41 Message						   24.80, 21.9560, 34.1602  	24.799999	25.739479	0.003773	-0.009998	-0.004999	0.000118	1.000000																							

$	12-Dec-2018	19:03:18.734	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	151.981130	30.450000	139.000000																																

$	12-Dec-2018	19:03:19.133	DeAnchor	SBE41 Message						   24.80, 21.9535, 34.1587  	24.799999	25.634675	0.016729	0.000000	-0.004999	0.000478	1.000000																							

$	12-Dec-2018	19:03:19.833	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.799999	151.862621	30.450000	139.000000																																

$	12-Dec-2018	19:03:20.131	DeAnchor	SBE41 Message						   24.79, 21.9559, 34.1582  	24.790001	25.417585	0.043571	0.009998	0.004999	0.001236	1.000000																							

$	12-Dec-2018	19:03:20.830	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.790001	151.615374	30.450000	139.000000																																

$	12-Dec-2018	19:03:21.131	DeAnchor	SBE41 Message						   24.82, 21.9618, 34.1565  	24.820000	25.144079	0.077398	-0.029999	-0.010000	0.002340	1.000000																							

$	12-Dec-2018	19:03:21.734	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.820000	151.491798	30.450000	139.000000																																

$	12-Dec-2018	19:03:22.130	DeAnchor	SBE41 Message						   25.63, 21.9678, 34.1549  	25.629999	24.928745	0.104047	-0.809999	-0.419999	0.003558	1.000000																							

$	12-Dec-2018	19:03:22.735	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.629999	151.289932	30.450000	139.000000																																

$	12-Dec-2018	19:03:23.128	DeAnchor	SBE41 Message						   25.66, 21.9709, 34.1553  	25.660000	24.828877	0.116426	-0.030001	-0.420000	0.004668	1.000000																							

$	12-Dec-2018	19:03:23.833	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.660000	151.135631	30.450000	139.000000																																

$	12-Dec-2018	19:03:24.127	DeAnchor	SBE41 Message						   25.65, 21.9711, 34.1566  	25.650000	24.827384	0.116630	0.010000	-0.010000	0.005559	1.000000																							

$	12-Dec-2018	19:03:24.734	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.650000	150.976583	30.450000	139.000000																																

$	12-Dec-2018	19:03:25.126	DeAnchor	SBE41 Message						   25.66, 21.9713, 34.1569  	25.660000	24.916881	0.105209	-0.010000	0.000000	0.006167	1.000000																							

$	12-Dec-2018	19:03:25.832	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.660000	150.801152	30.450000	139.000000																																

$	12-Dec-2018	19:03:26.124	DeAnchor	SBE41 Message						   25.64, 21.9719, 34.1567  	25.639999	25.105954	0.078395	0.020000	0.005000	0.006437	1.000000																							

$	12-Dec-2018	19:03:26.737	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.639999	150.594220	30.450000	139.000000																																

$	12-Dec-2018	19:03:27.124	DeAnchor	SBE41 Message						   25.66, 21.9727, 34.1567  	25.660000	25.344297	0.035928	-0.020000	0.000000	0.006387	1.000000																							

$	12-Dec-2018	19:03:27.738	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	25.660000	150.422387	30.450000	139.000000																																

$	12-Dec-2018	19:03:28.122	DeAnchor	SBE41 Message						   24.54, 21.9737, 34.1564  	24.540001	25.533623	-0.014368	1.119999	0.549999	0.006134	1.000000																							

$	12-Dec-2018	19:03:28.833	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.540001	150.276750	30.450000	139.000000																																

$	12-Dec-2018	19:03:29.121	DeAnchor	SBE41 Message						   24.07, 21.9759, 34.1564  	24.070000	25.622272	-0.059183	0.470001	0.795000	0.005820	1.000000																							

$	12-Dec-2018	19:03:29.736	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.070000	150.023863	30.450000	139.000000																																

$	12-Dec-2018	19:03:30.121	DeAnchor	SBE41 Message						   24.08, 21.9760, 34.1571  	24.080000	25.613735	-0.084771	-0.010000	0.230000	0.005551	1.000000																							

$	12-Dec-2018	19:03:30.834	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.080000	149.879575	30.450000	139.000000																																

$	12-Dec-2018	19:03:31.119	DeAnchor	SBE41 Message						   24.08, 21.9772, 34.1567  	24.080000	25.483925	-0.081070	0.000000	-0.005000	0.005432	1.000000																							

$	12-Dec-2018	19:03:31.832	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.080000	149.729222	30.450000	139.000000																																

$	12-Dec-2018	19:03:32.119	DeAnchor	SBE41 Message						   24.08, 21.9769, 34.1574  	24.080000	25.187562	-0.044593	0.000000	0.000000	0.005590	1.000000																							

$	12-Dec-2018	19:03:32.832	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.080000	149.582006	30.450000	139.000000																																

$	12-Dec-2018	19:03:33.118	DeAnchor	SBE41 Message						   24.07, 21.9775, 34.1575  	24.070000	24.773163	0.017748	0.010000	0.005000	0.006090	1.000000																							

$	12-Dec-2018	19:03:33.831	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.070000	149.342902	30.450000	139.000000																																

$	12-Dec-2018	19:03:34.118	DeAnchor	SBE41 Message						   24.09, 21.9785, 34.1572  	24.090000	24.396863	0.087708	-0.020000	-0.005000	0.006859	1.000000																							

$	12-Dec-2018	19:03:34.735	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.090000	149.122611	30.450000	139.000000																																

$	12-Dec-2018	19:03:35.116	DeAnchor	SBE41 Message						   24.08, 21.9794, 34.1572  	24.080000	24.175886	0.144555	0.010000	-0.005000	0.007741	1.000000																							

$	12-Dec-2018	19:03:35.831	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.080000	148.920268	30.450000	139.000000																																

$	12-Dec-2018	19:03:36.115	DeAnchor	SBE41 Message						   24.08, 21.9804, 34.1571  	24.080000	24.095789	0.177905	0.000000	0.005000	0.008619	1.000000																							

$	12-Dec-2018	19:03:36.831	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.080000	148.790702	30.450000	139.000000																																

$	12-Dec-2018	19:03:37.114	DeAnchor	SBE41 Message						   24.07, 21.9810, 34.1573  	24.070000	24.079951	0.190832	0.010000	0.005000	0.009447	1.000000																							

$	12-Dec-2018	19:03:37.735	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.070000	148.636818	30.450000	139.000000																																

$	12-Dec-2018	19:03:38.112	DeAnchor	SBE41 Message						   24.09, 21.9817, 34.1574  	24.090000	24.079365	0.189824	-0.020000	-0.005000	0.010223	1.000000																							

$	12-Dec-2018	19:03:38.832	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.090000	148.433708	30.450000	139.000000																																

$	12-Dec-2018	19:03:39.112	DeAnchor	SBE41 Message						   24.09, 21.9819, 34.1576  	24.090000	24.080331	0.173610	0.000000	-0.010000	0.010955	1.000000																							

$	12-Dec-2018	19:03:39.830	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.090000	148.257695	30.450000	139.000000																																

$	12-Dec-2018	19:03:40.111	DeAnchor	SBE41 Message						   24.08, 21.9824, 34.1579  	24.080000	24.080390	0.136917	0.010000	0.005000	0.011652	1.000000																							

$	12-Dec-2018	19:03:40.830	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.080000	148.123138	30.450000	139.000000																																

$	12-Dec-2018	19:03:41.111	DeAnchor	SBE41 Message						   24.09, 21.9837, 34.1573  	24.090000	24.079921	0.085714	-0.010000	0.000000	0.012322	1.000000																							

$	12-Dec-2018	19:03:41.735	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.090000	147.909224	30.450000	139.000000																																

$	12-Dec-2018	19:03:42.109	DeAnchor	SBE41 Message						   24.08, 21.9849, 34.1570  	24.080000	24.080768	0.039078	0.010000	0.000000	0.012973	1.000000																							

$	12-Dec-2018	19:03:42.830	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.080000	147.825696	30.450000	139.000000																																

$	12-Dec-2018	19:03:43.108	DeAnchor	SBE41 Message						   24.07, 21.9849, 34.1576  	24.070000	24.083112	0.011468	0.010000	0.010000	0.013612	1.000000																							

$	12-Dec-2018	19:03:43.830	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.070000	147.613920	30.450000	139.000000																																

$	12-Dec-2018	19:03:44.108	DeAnchor	SBE41 Message						   24.08, 21.9861, 34.1571  	24.080000	24.084990	0.001334	-0.010000	0.000000	0.014249	1.000000																							

$	12-Dec-2018	19:03:44.829	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.080000	147.398725	30.450000	139.000000																																

$	12-Dec-2018	19:03:45.107	DeAnchor	SBE41 Message						   24.08, 21.9867, 34.1571  	24.080000	24.084942	-0.000618	0.000000	-0.005000	0.014895	1.000000																							

$	12-Dec-2018	19:03:45.734	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.080000	147.176482	30.450000	139.000000																																

$	12-Dec-2018	19:03:46.105	DeAnchor	SBE41 Message						   24.07, 21.9868, 34.1575  	24.070000	24.083054	-0.000457	0.010000	0.005000	0.015567	1.000000																							

$	12-Dec-2018	19:03:46.829	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.070000	147.092775	30.450000	139.000000																																

$	12-Dec-2018	19:03:47.104	DeAnchor	SBE41 Message						   24.09, 21.9872, 34.1575  	24.090000	24.080282	0.000006	-0.020000	-0.005000	0.016284	1.000000																							

$	12-Dec-2018	19:03:47.829	DeAnchor	PTH Message														796.613037	23.641667	53.795380	24.632776																			

$	12-Dec-2018	19:03:48.031	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	24.090000	146.908984	29.775000	189.000000																																

$	12-Dec-2018	19:03:48.144	DeAnchor	SBE41 Message						   23.24, 21.9876, 34.1578  	23.240000	24.078032	0.000291	0.850000	0.415000	0.017073	1.000000																							

$	12-Dec-2018	19:03:48.829	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	23.240000	146.659762	29.775000	189.000000																																

$	12-Dec-2018	19:03:49.103	DeAnchor	SBE41 Message						   22.67, 21.9892, 34.1575  	22.670000	24.074777	0.000636	0.570000	0.710000	0.017974	1.000000																							

$	12-Dec-2018	19:03:49.733	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	22.670000	146.492861	29.775000	189.000000																																

$	12-Dec-2018	19:03:50.101	DeAnchor	SBE41 Message						   22.67, 21.9895, 34.1576  	22.670000	24.049960	0.003809	0.000000	0.285000	0.019076	1.000000																							

$	12-Dec-2018	19:03:50.732	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	22.670000	146.375969	29.775000	189.000000																																

$	12-Dec-2018	19:03:51.100	DeAnchor	SBE41 Message						   22.68, 21.9896, 34.1580  	22.680000	23.945364	0.017029	-0.010000	-0.005000	0.020586	1.000000																							

$	12-Dec-2018	19:03:51.731	DeAnchor	RecMode true, PID Off, BE Ext'g, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	22.680000	146.132261	29.775000	189.000000																																

$	12-Dec-2018	19:03:52.497	DeAnchor	Deanchor velocity (slope) threshold exceeded

$	12-Dec-2018	19:03:52.099	DeAnchor	SBE41 Message						   22.67, 21.9896, 34.1584  	22.670000	23.699519	0.047657	0.010000	0.000000	0.022953	1.000000																							

$	12-Dec-2018	19:03:52.832	DeAnchor	Normal State Exit

$	12-Dec-2018	19:03:52.837	DeAnchor	RecMode true, PID Off, Motor Stopped, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	22.670000	146.021537	29.775000	189.000000																																

$	12-Dec-2018	19:03:53.497	Ascend	State Entry

$	12-Dec-2018	19:03:53.610	Ascend	Starting valve open

$	12-Dec-2018	19:03:53.712	Ascend	Skipping ascend table depth = 100

$	12-Dec-2018	19:03:53.814	Ascend	Skipping ascend table depth = 95

$	12-Dec-2018	19:03:53.916	Ascend	Skipping ascend table depth = 90

$	12-Dec-2018	19:03:54.018	Ascend	Skipping ascend table depth = 85

$	12-Dec-2018	19:03:54.121	Ascend	Skipping ascend table depth = 80

$	12-Dec-2018	19:03:54.223	Ascend	Skipping ascend table depth = 75

$	12-Dec-2018	19:03:54.327	Ascend	Skipping ascend table depth = 70

$	12-Dec-2018	19:03:54.429	Ascend	Skipping ascend table depth = 65

$	12-Dec-2018	19:03:54.531	Ascend	Skipping ascend table depth = 60

$	12-Dec-2018	19:03:54.633	Ascend	Skipping ascend table depth = 55

$	12-Dec-2018	19:03:54.735	Ascend	Skipping ascend table depth = 50

$	12-Dec-2018	19:03:54.837	Ascend	Skipping ascend table depth = 45

$	12-Dec-2018	19:03:54.939	Ascend	Skipping ascend table depth = 40

$	12-Dec-2018	19:03:55.043	Ascend	Skipping ascend table depth = 35

$	12-Dec-2018	19:03:55.048	Ascend	Skipping ascend table depth = 30

$	12-Dec-2018	19:03:55.151	Ascend	Skipping ascend table depth = 25

$	12-Dec-2018	19:03:55.351	Ascend	Starting ISUS sample

$	12-Dec-2018	19:03:55.454	Ascend	Starting MSC Sequence CTDValues (time, press, temp, salinity) 0, 0.00, 0.000, 0.00

$	12-Dec-2018	19:03:53.099	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.009967	0.287841	4.800014	4.800014	39321.000000	1.000000										

$	12-Dec-2018	19:03:53.099	Ascend	SBE41 Message						   22.68, 21.9907, 34.1579  	22.680000	23.338610	0.092283	-0.010000	0.000000	0.027116	1.000000																							

$	12-Dec-2018	19:03:56.104	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.680000	145.784371	29.775000	189.000000																																

$	12-Dec-2018	19:03:56.216	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.7188631

$	12-Dec-2018	19:03:56.322	Ascend	MSC: OUT: CTD,0,0.0000,0.0000,0.0000

$	12-Dec-2018	19:03:54.121	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019901	0.480159	3.487836	3.487836	28572.000000	1.000000										

$	12-Dec-2018	19:03:54.121	Ascend	SBE41 Message						   22.67, 21.9915, 34.1579  	22.670000	22.993063	0.134801	0.010000	0.000000	0.035565	1.000000																							

$	12-Dec-2018	19:03:55.146	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.009901	0.720000	4.480146	4.480146	36701.000000	1.000000										

$	12-Dec-2018	19:03:55.146	Ascend	SBE41 Message						   22.67, 21.9916, 34.1582  	22.670000	22.777641	0.161134	0.000000	0.005000	0.060599	1.000000																							

$	12-Dec-2018	19:03:56.096	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.010000	1.007841	5.520019	5.520019	45219.000000	1.000000										

$	12-Dec-2018	19:03:56.096	Ascend	SBE41 Message						   22.68, 21.9926, 34.1580  	22.680000	22.694143	0.171216	-0.010000	-0.005000	0.062726	1.000000																							

$	12-Dec-2018	19:03:57.466	Ascend	MSC: RX: 

$	12-Dec-2018	19:03:57.672	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.680000	145.696446	29.775000	189.000000																																

$	12-Dec-2018	19:03:57.784	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.5608934

$	12-Dec-2018	19:03:57.888	Ascend	MSC: RX: 

$	12-Dec-2018	19:03:57.121	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009934	1.248001	5.007850	5.007850	41024.000000	1.000000										

$	12-Dec-2018	19:03:57.121	Ascend	SBE41 Message						   22.68, 21.9927, 34.1583  	22.680000	22.676169	0.173048	0.000000	-0.005000	0.064351	1.000000																							

$	12-Dec-2018	19:03:58.536	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.680000	145.667396	29.775000	189.000000																																

$	12-Dec-2018	19:03:58.092	Ascend	Vel PID Values																		0.050000	0.010000	0.009967	0.010000	1.440160	4.447983	4.447983	36437.000000	1.000000										

$	12-Dec-2018	19:03:58.092	Ascend	SBE41 Message						   22.67, 21.9934, 34.1582  	22.670000	22.674192	0.170206	0.010000	0.005000	0.065514	1.000000																							

$	12-Dec-2018	19:03:59.337	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.670000	145.602353	29.775000	189.000000																																

$	12-Dec-2018	19:03:59.134	Ascend	Vel PID Values																		0.050000	-0.020000	-0.019901	-0.029868	1.775683	7.040179	7.040179	57673.000000	1.000000										

$	12-Dec-2018	19:03:59.134	Ascend	SBE41 Message						   22.69, 21.9941, 34.1579  	22.690001	22.673852	0.157277	-0.020000	-0.005000	0.066498	1.000000																							

$	12-Dec-2018	19:04:00.336	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.690001	145.565905	29.775000	189.000000																																

$	12-Dec-2018	19:04:00.134	Ascend	Vel PID Values																		0.050000	0.010000	0.009901	0.029801	1.968160	4.975691	4.975691	40760.000000	1.000000										

$	12-Dec-2018	19:04:00.134	Ascend	SBE41 Message						   22.68, 21.9942, 34.1581  	22.680000	22.674446	0.126768	0.010000	-0.005000	0.067758	1.000000																							

$	12-Dec-2018	19:04:01.132	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.009867	2.208001	5.968143	5.968143	48891.000000	1.000000										

$	12-Dec-2018	19:04:01.132	Ascend	SBE41 Message						   22.68, 21.9947, 34.1581  	22.680000	22.675895	0.081942	0.000000	0.005000	0.069550	1.000000																							

$	12-Dec-2018	19:04:01.679	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.680000	145.579539	29.775000	189.000000																																

$	12-Dec-2018	19:04:01.791	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0015443

$	12-Dec-2018	19:04:02.097	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.680000	145.464867	29.775000	189.000000																																

$	12-Dec-2018	19:04:02.089	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000033	2.448001	6.208006	6.208006	50855.000000	1.000000										

$	12-Dec-2018	19:04:02.089	Ascend	SBE41 Message						   22.68, 21.9945, 34.1587  	22.680000	22.677500	0.039013	0.000000	0.000000	0.071713	1.000000																							

$	12-Dec-2018	19:04:03.145	Ascend	Vel PID Values																		0.050000	0.010000	0.009967	0.009967	2.640160	5.647987	5.647987	46268.000000	1.000000										

$	12-Dec-2018	19:04:03.145	Ascend	SBE41 Message						   22.67, 21.9959, 34.1579  	22.670000	22.679105	0.012180	0.010000	0.005000	0.073814	1.000000																							

$	12-Dec-2018	19:04:03.690	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.670000	145.478040	29.775000	189.000000																																

$	12-Dec-2018	19:04:03.803	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0074824

$	12-Dec-2018	19:04:04.109	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.670000	145.389728	29.775000	189.000000																																

$	12-Dec-2018	19:04:04.087	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019901	2.927842	7.440165	7.440165	60949.000000	1.000000										

$	12-Dec-2018	19:04:04.087	Ascend	SBE41 Message						   22.68, 21.9963, 34.1581  	22.680000	22.680525	0.001683	-0.010000	0.000000	0.075531	1.000000																							

$	12-Dec-2018	19:04:05.155	Ascend	Vel PID Values																		0.050000	-0.010000	-0.010000	-0.000066	3.215842	7.727869	7.727869	63306.000000	1.000000										

$	12-Dec-2018	19:04:05.155	Ascend	SBE41 Message						   22.69, 21.9970, 34.1567  	22.690001	22.680867	-0.000581	-0.010000	-0.010000	0.076477	1.000000																							

$	12-Dec-2018	19:04:05.704	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.690001	145.378113	29.775000	189.000000																																

$	12-Dec-2018	19:04:05.815	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0059685

$	12-Dec-2018	19:04:06.120	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.690001	145.294562	29.775000	189.000000																																

$	12-Dec-2018	19:04:06.084	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009967	3.456002	7.215847	7.215847	59112.000000	1.000000										

$	12-Dec-2018	19:04:06.084	Ascend	SBE41 Message						   22.69, 21.9973, 34.1581  	22.690001	22.679748	-0.000688	0.000000	-0.005000	0.076306	1.000000																							

$	12-Dec-2018	19:04:07.166	Ascend	Vel PID Values																		0.050000	0.270000	0.269102	0.269135	2.404313	-14.143919	-14.143919	-115866.000000	1.000000										

$	12-Dec-2018	19:04:07.166	Ascend	SBE41 Message						   22.42, 21.9982, 34.1580  	22.420000	22.678342	-0.000556	0.270000	0.135000	0.074740	1.000000																							

$	12-Dec-2018	19:04:07.713	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.420000	145.260572	29.775000	189.000000																																

$	12-Dec-2018	19:04:07.824	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0040199

$	12-Dec-2018	19:04:08.130	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	22.420000	145.174146	29.775000	189.000000																																

$	12-Dec-2018	19:04:08.083	Ascend	Vel PID Values																		0.050000	0.980000	0.977635	0.708533	657.815318	-71.995455	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:08.083	Ascend	SBE41 Message						   21.44, 21.9989, 34.1583  	21.440001	22.677793	-0.000414	0.980000	0.625000	0.071978	1.000000																							

$	12-Dec-2018	19:04:09.172	Ascend	Vel PID Values																		0.050000	0.120001	0.122859	-0.854777	-17408.971100	652.214557	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:09.172	Ascend	SBE41 Message						   21.32, 21.9988, 34.1588  	21.320000	22.670066	0.000721	0.120001	0.550000	0.068339	1.000000																							

$	12-Dec-2018	19:04:09.719	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.320000	145.166561	29.775000	189.000000																																

$	12-Dec-2018	19:04:09.832	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0027344

$	12-Dec-2018	19:04:10.138	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.320000	145.162046	29.775000	189.000000																																

$	12-Dec-2018	19:04:10.081	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009563	-0.132422	503216.437614	-17404.170760	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:10.081	Ascend	SBE41 Message						   21.33, 21.9994, 34.1588  	21.330000	22.617163	0.007459	-0.010000	0.055000	0.064403	1.000000																							

$	12-Dec-2018	19:04:11.184	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009530	-14591896.452533	503220.437677	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:11.184	Ascend	SBE41 Message						   21.33, 22.0000, 34.1586  	21.330000	22.452095	0.028064	0.000000	-0.005000	0.060978	1.000000																							

$	12-Dec-2018	19:04:11.730	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.330000	145.157911	29.775000	189.000000																																

$	12-Dec-2018	19:04:11.842	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0050293

$	12-Dec-2018	19:04:12.148	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.330000	144.967265	29.775000	189.000000																																

$	12-Dec-2018	19:04:12.079	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	0.000033	423163377.363568	-14591892.452537	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:12.079	Ascend	SBE41 Message						   21.33, 22.0008, 34.1584  	21.330000	22.147850	0.065859	0.000000	0.000000	0.059265	1.000000																							

$	12-Dec-2018	19:04:13.181	Ascend	Vel PID Values																		0.050000	0.020000	0.019934	0.019934	-12271736515.398317	423163379.763540	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:13.181	Ascend	SBE41 Message						   21.31, 22.0013, 34.1585  	21.309999	21.788179	0.110389	0.020000	0.010000	0.059871	1.000000																							

$	12-Dec-2018	19:04:13.853	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.309999	144.806907	29.775000	189.000000																																

$	12-Dec-2018	19:04:13.964	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0178542

$	12-Dec-2018	19:04:14.269	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.309999	144.516744	29.775000	189.000000																																

$	12-Dec-2018	19:04:14.076	Ascend	Vel PID Values																		0.050000	-0.020000	-0.019867	-0.039801	355880357278.886290	-12271736509.798307	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:14.076	Ascend	SBE41 Message						   21.33, 22.0012, 34.1589  	21.330000	21.510495	0.144619	-0.020000	0.000000	0.062393	1.000000																							

$	12-Dec-2018	19:04:15.085	Ascend	Vel PID Values																		0.050000	0.000000	-0.000066	0.019801	-10320530359707.461000	355880357282.886290	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:15.085	Ascend	SBE41 Message						   21.33, 22.0020, 34.1587  	21.330000	21.373733	0.161394	0.000000	-0.010000	0.065722	1.000000																							

$	12-Dec-2018	19:04:15.961	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.330000	144.365110	29.775000	189.000000																																

$	12-Dec-2018	19:04:16.075	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0076166

$	12-Dec-2018	19:04:16.178	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	0.000066	299295380429896.560000	-10320530359703.461000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:16.178	Ascend	SBE41 Message						   21.33, 22.0023, 34.1587  	21.330000	21.333126	0.166372	0.000000	0.000000	0.068969	1.000000																							

$	12-Dec-2018	19:04:16.725	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.330000	144.024201	29.775000	189.000000																																

$	12-Dec-2018	19:04:17.073	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	0.000000	-8679566032465620.000000	299295380429900.560000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:17.073	Ascend	SBE41 Message						   21.33, 22.0035, 34.1583  	21.330000	21.325623	0.166341	0.000000	0.000000	0.071805	1.000000																							

$	12-Dec-2018	19:04:17.627	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.330000	143.949285	29.775000	189.000000																																

$	12-Dec-2018	19:04:18.073	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	0.000000	251707414941501340.000000	-8679566032465616.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:18.073	Ascend	SBE41 Message						   21.33, 22.0037, 34.1584  	21.330000	21.324369	0.159926	0.000000	0.000000	0.074134	1.000000																							

$	12-Dec-2018	19:04:18.627	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.330000	143.947035	29.775000	189.000000																																

$	12-Dec-2018	19:04:19.071	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.009967	-7299515033303536600.000000	251707414941501340.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:19.071	Ascend	SBE41 Message						   21.34, 22.0040, 34.1587  	21.340000	21.325564	0.139329	-0.010000	-0.005000	0.075940	1.000000																							

$	12-Dec-2018	19:04:19.627	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.340000	143.945254	29.775000	189.000000																																

$	12-Dec-2018	19:04:20.071	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019901	211685935965802560000.000000	-7299515033303536600.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:20.071	Ascend	SBE41 Message						   21.33, 22.0042, 34.1581  	21.330000	21.327890	0.101393	0.010000	0.000000	0.077221	1.000000																							

$	12-Dec-2018	19:04:20.627	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.330000	143.941998	29.775000	189.000000																																

$	12-Dec-2018	19:04:21.069	Ascend	Vel PID Values																		0.050000	0.010000	0.010000	0.000066	-6138892143008274200000.000000	211685935965802560000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:21.069	Ascend	SBE41 Message						   21.32, 22.0048, 34.1582  	21.320000	21.329718	0.056682	0.010000	0.010000	0.077983	1.000000																							

$	12-Dec-2018	19:04:21.627	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.320000	143.930942	29.775000	189.000000																																

$	12-Dec-2018	19:04:22.069	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019934	178027872147239950000000.000000	-6138892143008274200000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:22.069	Ascend	SBE41 Message						   21.33, 22.0059, 34.1585  	21.330000	21.330981	0.022192	-0.010000	0.000000	0.078221	1.000000																							

$	12-Dec-2018	19:04:22.626	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.330000	143.929072	29.775000	189.000000																																

$	12-Dec-2018	19:04:23.068	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009901	-5162808292269957800000000.000000	178027872147239950000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:23.068	Ascend	SBE41 Message						   21.33, 22.0060, 34.1588  	21.330000	21.331936	0.005166	0.000000	-0.005000	0.077931	1.000000																							

$	12-Dec-2018	19:04:23.626	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.330000	143.932298	29.775000	189.000000																																

$	12-Dec-2018	19:04:24.067	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.009934	149721440475828800000000000.000000	-5162808292269957800000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:24.067	Ascend	SBE41 Message						   21.34, 22.0070, 34.1583  	21.340000	21.331732	0.000171	-0.010000	-0.005000	0.077117	1.000000																							

$	12-Dec-2018	19:04:24.628	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.340000	143.924616	29.775000	189.000000																																

$	12-Dec-2018	19:04:25.065	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009934	-4341921773799035100000000000.000000	149721440475828800000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:25.065	Ascend	SBE41 Message						   21.34, 22.0072, 34.1586  	21.340000	21.330030	-0.000546	0.000000	-0.005000	0.075792	1.000000																							

$	12-Dec-2018	19:04:25.629	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	21.340000	143.924855	29.775000	189.000000																																

$	12-Dec-2018	19:04:26.064	Ascend	Vel PID Values																		0.050000	0.630001	0.627905	0.627938	125915731440172020000000000000.000000	-4341921773799035100000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:26.064	Ascend	SBE41 Message						   20.71, 22.0077, 34.1587  	20.709999	21.328570	-0.000520	0.630001	0.315001	0.073962	1.000000																							

$	12-Dec-2018	19:04:26.629	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	20.709999	143.917635	29.775000	189.000000																																

$	12-Dec-2018	19:04:27.063	Ascend	Vel PID Values																		0.050000	0.689999	0.689795	0.061891	-3651556211764988700000000000000.000000	125915731440172020000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:27.063	Ascend	SBE41 Message						   20.02, 22.0072, 34.1595  	20.020000	21.327566	-0.000248	0.689999	0.660000	0.071628	1.000000																							

$	12-Dec-2018	19:04:27.629	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	20.020000	143.923648	29.775000	189.000000																																

$	12-Dec-2018	19:04:28.062	Ascend	Vel PID Values																		0.050000	0.020000	0.022230	-0.667566	105895130141184680000000000000000.000000	-3651556211764988700000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:28.062	Ascend	SBE41 Message						   20.00, 22.0077, 34.1595  	20.000000	21.310892	0.002101	0.020000	0.355000	0.068837	1.000000																							

$	12-Dec-2018	19:04:28.628	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 16  Pres, BP, BatV, BatI: 	20.000000	143.946163	29.775000	189.000000																																

$	12-Dec-2018	19:04:29.087	Ascend	ERROR: Missed PH message 29

$	12-Dec-2018	19:04:29.092	Ascend	ERROR: Missed NO3 message 25

$	12-Dec-2018	19:04:29.293	Ascend	Sending Optode Do Sample

$	12-Dec-2018	19:04:29.395	Ascend	Sending Optode Wakeup

$	12-Dec-2018	19:04:29.900	Ascend	Sending FLBB run command

$	12-Dec-2018	19:04:29.917	Ascend	Sending radiometer pollSample

$	12-Dec-2018	19:04:29.062	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009897	-0.032126	-3070958774094356300000000000000000.000000	105895130141184680000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:29.062	Ascend	SBE41 Message						   20.01, 22.0085, 34.1587  	20.010000	21.227556	0.012629	-0.010000	0.005000	0.065848	1.000000																							

$	12-Dec-2018	19:04:30.120	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019831	89057804448736339000000000000000000.000000	-3070958774094356300000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:30.120	Ascend	SBE41 Message						   20.00, 22.0092, 34.1591  	20.000000	21.015942	0.038948	0.010000	0.000000	0.063179	1.000000																							

$	12-Dec-2018	19:04:30.715	Ascend	Radiometer Message																																	SATFI40287,-0.0048,0.0009,-0.0063,-0.2163 			

$	12-Dec-2018	19:04:31.131	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.000000	143.923134	29.775000	189.000000																																

$	12-Dec-2018	19:04:31.245	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.1637924

$	12-Dec-2018	19:04:30.991	Ascend	Optode Message																												%!4330,1532,265.166,22.400,29.175,29.106,36.143,7.037,538.1,467.4,113.6  								

$	12-Dec-2018	19:04:30.993	Ascend	Optode Message																												#  								

$	12-Dec-2018	19:04:31.059	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019868	-2582676329013353900000000000000000000.000000	89057804448736339000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:31.059	Ascend	SBE41 Message						   20.01, 22.0092, 34.1594  	20.010000	20.685612	0.079915	-0.010000	0.000000	0.061568	1.000000																							

$	12-Dec-2018	19:04:31.821	Ascend	FLBB Message																											/99/99,99:99:99,695,2313,700,4130,533  									

$	12-Dec-2018	19:04:32.233	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.010000	144.293010	29.775000	189.000000																																

$	12-Dec-2018	19:04:32.345	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0953620

$	12-Dec-2018	19:04:32.058	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019868	74897613541387263000000000000000000000.000000	-2582676329013353900000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:32.058	Ascend	SBE41 Message						   20.00, 22.0100, 34.1591  	20.000000	20.349233	0.121504	0.010000	0.000000	0.061524	1.000000																							

$	12-Dec-2018	19:04:32.998	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.000000	144.574612	29.775000	189.000000																																

$	12-Dec-2018	19:04:33.108	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019868	-2172030792700230600000000000000000000000.000000	74897613541387263000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:33.108	Ascend	SBE41 Message						   20.01, 22.0110, 34.1587  	20.010000	20.124621	0.149102	-0.010000	0.000000	0.063059	1.000000																							

$	12-Dec-2018	19:04:33.997	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.010000	144.572742	29.775000	189.000000																																

$	12-Dec-2018	19:04:34.108	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019868	62988892988306683000000000000000000000000.000000	-2172030792700230600000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:34.108	Ascend	SBE41 Message						   20.00, 22.0111, 34.1576  	20.000000	20.030020	0.160654	0.010000	0.000000	0.065663	1.000000																							

$	12-Dec-2018	19:04:35.005	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.000000	144.595474	29.775000	189.000000																																

$	12-Dec-2018	19:04:35.135	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019868	-1826677896660894000000000000000000000000000.000000	62988892988306683000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:35.135	Ascend	SBE41 Message						   20.01, 22.0114, 34.1592  	20.010000	20.007397	0.163343	-0.010000	0.000000	0.068610	1.000000																							

$	12-Dec-2018	19:04:35.995	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.010000	144.616224	29.775000	189.000000																																

$	12-Dec-2018	19:04:36.107	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019868	52973659003165933000000000000000000000000000.000000	-1826677896660894000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:36.107	Ascend	SBE41 Message						   20.00, 22.0120, 34.1592  	20.000000	20.005041	0.161560	0.010000	0.000000	0.071386	1.000000																							

$	12-Dec-2018	19:04:36.996	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.000000	144.647717	29.775000	189.000000																																

$	12-Dec-2018	19:04:37.107	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019868	-1536236111091812100000000000000000000000000000.000000	52973659003165933000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:37.107	Ascend	SBE41 Message						   20.01, 22.0121, 34.1594  	20.010000	20.005000	0.151226	-0.010000	0.000000	0.073715	1.000000																							

$	12-Dec-2018	19:04:37.996	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.010000	144.688189	29.775000	189.000000																																

$	12-Dec-2018	19:04:38.106	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019868	44550847221662558000000000000000000000000000000.000000	-1536236111091812100000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:38.106	Ascend	SBE41 Message						   20.00, 22.0129, 34.1592  	20.000000	20.005000	0.125024	0.010000	0.000000	0.075560	1.000000																							

$	12-Dec-2018	19:04:38.995	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.000000	144.718423	29.775000	189.000000																																

$	12-Dec-2018	19:04:39.106	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019868	-1291974569428214100000000000000000000000000000000.000000	44550847221662558000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:39.106	Ascend	SBE41 Message						   20.01, 22.0134, 34.1589  	20.010000	20.005000	0.084156	-0.010000	0.000000	0.076889	1.000000																							

$	12-Dec-2018	19:04:39.994	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.010000	144.732073	29.775000	189.000000																																

$	12-Dec-2018	19:04:40.106	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009901	37467262513418210000000000000000000000000000000000.000000	-1291974569428214100000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:40.106	Ascend	SBE41 Message						   20.01, 22.0138, 34.1592  	20.010000	20.005000	0.042558	0.000000	-0.005000	0.077713	1.000000																							

$	12-Dec-2018	19:04:40.994	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.010000	144.761808	29.775000	189.000000																																

$	12-Dec-2018	19:04:41.446	Ascend	MSC: Timeout for CommandSendCTD

$	12-Dec-2018	19:04:41.451	Ascend	MSC: ERROR Failed Sample on Sequence #2

$	12-Dec-2018	19:04:41.552	Ascend	ERROR: MSC failed sample sequence.

$	12-Dec-2018	19:04:41.109	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	0.000033	-1086550612889128100000000000000000000000000000000000.000000	37467262513418210000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:41.109	Ascend	SBE41 Message						   20.01, 22.0144, 34.1591  	20.010000	20.005030	0.014784	0.000000	0.000000	0.078030	1.000000																							

$	12-Dec-2018	19:04:42.104	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.009967	31509967773784714000000000000000000000000000000000000.000000	-1086550612889128100000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:42.104	Ascend	SBE41 Message						   20.02, 22.0143, 34.1585  	20.020000	20.005282	0.003058	-0.010000	-0.005000	0.077852	1.000000																							

$	12-Dec-2018	19:04:42.653	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.020000	144.771241	29.775000	189.000000																																

$	12-Dec-2018	19:04:42.765	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.3187889

$	12-Dec-2018	19:04:43.071	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	20.020000	145.008743	29.775000	189.000000																																

$	12-Dec-2018	19:04:43.048	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009934	-913789065439756700000000000000000000000000000000000000.000000	31509967773784714000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:43.048	Ascend	SBE41 Message						   20.02, 22.0152, 34.1589  	20.020000	20.006129	0.000157	0.000000	-0.005000	0.077166	1.000000																							

$	12-Dec-2018	19:04:44.118	Ascend	Vel PID Values																		0.050000	0.080000	0.079734	0.079767	26499882897752944000000000000000000000000000000000000000.000000	-913789065439756700000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:44.118	Ascend	SBE41 Message						   19.94, 22.0155, 34.1592  	19.940001	20.007810	-0.000342	0.080000	0.040000	0.075957	1.000000																							

$	12-Dec-2018	19:04:44.666	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	19.940001	145.245992	29.775000	189.000000																																

$	12-Dec-2018	19:04:44.777	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0066908

$	12-Dec-2018	19:04:44.986	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	19.940001	145.616241	29.775000	189.000000																																

$	12-Dec-2018	19:04:45.098	Ascend	Vel PID Values																		0.050000	0.860001	0.857404	0.777671	-768496604034835360000000000000000000000000000000000000000.000000	26499882897752944000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:45.098	Ascend	SBE41 Message						   19.08, 22.0160, 34.1596  	19.080000	20.010032	-0.000622	0.860001	0.470000	0.074226	1.000000																							

$	12-Dec-2018	19:04:46.133	Ascend	Vel PID Values																		0.050000	0.400000	0.401526	-0.455878	22286401517010226000000000000000000000000000000000000000000.000000	-768496604034835360000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:46.133	Ascend	SBE41 Message						   18.68, 22.0168, 34.1594  	18.680000	20.008270	-0.000405	0.400000	0.630000	0.072011	1.000000																							

$	12-Dec-2018	19:04:46.681	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.680000	145.753473	29.775000	189.000000																																

$	12-Dec-2018	19:04:46.794	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0097579

$	12-Dec-2018	19:04:47.099	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.680000	146.153361	29.775000	189.000000																																

$	12-Dec-2018	19:04:47.043	Ascend	Vel PID Values																		0.050000	0.010000	0.011301	-0.390225	-646305643993296540000000000000000000000000000000000000000000.000000	22286401517010226000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:47.043	Ascend	SBE41 Message						   18.67, 22.0165, 34.1600  	18.670000	19.977352	0.003417	0.010000	0.205000	0.069427	1.000000																							

$	12-Dec-2018	19:04:48.146	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009932	-0.021232	18742863675805600000000000000000000000000000000000000000000000.000000	-646305643993296540000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:48.146	Ascend	SBE41 Message						   18.68, 22.0175, 34.1596  	18.680000	19.858557	0.018102	-0.010000	0.000000	0.066755	1.000000																							

$	12-Dec-2018	19:04:48.696	Ascend	PTH Message														790.804932	23.741667	53.856415	24.632776																			

$	12-Dec-2018	19:04:48.898	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.680000	146.382354	28.400000	265.500000																																

$	12-Dec-2018	19:04:49.012	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.2125562

$	12-Dec-2018	19:04:49.318	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.680000	146.725118	28.400000	265.500000																																

$	12-Dec-2018	19:04:49.115	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019865	-543543046598362390000000000000000000000000000000000000000000000.000000	18742863675805600000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:49.115	Ascend	SBE41 Message						   18.67, 22.0178, 34.1597  	18.670000	19.603418	0.049652	0.010000	0.000000	0.064516	1.000000																							

$	12-Dec-2018	19:04:50.040	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019868	15762748351352508000000000000000000000000000000000000000000000000.000000	-543543046598362390000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:50.040	Ascend	SBE41 Message						   18.68, 22.0180, 34.1598  	18.680000	19.255486	0.092714	-0.010000	0.000000	0.063208	1.000000																							

$	12-Dec-2018	19:04:50.914	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.680000	146.997504	28.400000	265.500000																																

$	12-Dec-2018	19:04:51.027	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0104341

$	12-Dec-2018	19:04:51.130	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009901	-457119702189222750000000000000000000000000000000000000000000000000.000000	15762748351352508000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:51.130	Ascend	SBE41 Message						   18.68, 22.0183, 34.1599  	18.680000	18.942743	0.131516	0.000000	-0.005000	0.063018	1.000000																							

$	12-Dec-2018	19:04:51.680	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.680000	147.374101	28.400000	265.500000																																

$	12-Dec-2018	19:04:52.235	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.680000	147.505723	28.400000	265.500000																																

$	12-Dec-2018	19:04:52.038	Ascend	Vel PID Values																		0.050000	0.010000	0.009967	0.010000	13256471363487461000000000000000000000000000000000000000000000000000.000000	-457119702189222750000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:52.038	Ascend	SBE41 Message						   18.67, 22.0193, 34.1596  	18.670000	18.758397	0.154555	0.010000	0.005000	0.064040	1.000000																							

$	12-Dec-2018	19:04:53.049	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.009934	-384437669541136380000000000000000000000000000000000000000000000000000.000000	13256471363487461000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:53.049	Ascend	SBE41 Message						   18.67, 22.0197, 34.1591  	18.670000	18.690543	0.163251	0.000000	0.005000	0.066032	1.000000																							

$	12-Dec-2018	19:04:53.927	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.670000	147.528768	28.400000	265.500000																																

$	12-Dec-2018	19:04:54.046	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0142930

$	12-Dec-2018	19:04:54.045	Ascend	Vel PID Values																		0.050000	-0.020000	-0.019934	-0.019967	11148692416692954000000000000000000000000000000000000000000000000000000.000000	-384437669541136380000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:54.045	Ascend	SBE41 Message						   18.69, 22.0200, 34.1596  	18.690001	18.677017	0.164713	-0.020000	-0.010000	0.068587	1.000000																							

$	12-Dec-2018	19:04:54.698	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.690001	147.422656	28.400000	265.500000																																

$	12-Dec-2018	19:04:55.036	Ascend	Vel PID Values																		0.050000	0.020000	0.019867	0.039801	-323312080084095640000000000000000000000000000000000000000000000000000000.000000	11148692416692954000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:55.036	Ascend	SBE41 Message						   18.67, 22.0204, 34.1598  	18.670000	18.676184	0.160972	0.020000	0.000000	0.071142	1.000000																							

$	12-Dec-2018	19:04:55.601	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.670000	147.321522	28.400000	265.500000																																

$	12-Dec-2018	19:04:56.034	Ascend	Vel PID Values																		0.050000	-0.020000	-0.019868	-0.039735	9376050322438774400000000000000000000000000000000000000000000000000000000.000000	-323312080084095640000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:56.034	Ascend	SBE41 Message						   18.69, 22.0211, 34.1595  	18.690001	18.675866	0.146286	-0.020000	0.000000	0.073316	1.000000																							

$	12-Dec-2018	19:04:56.602	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.690001	147.299483	28.400000	265.500000																																

$	12-Dec-2018	19:04:57.032	Ascend	Vel PID Values																		0.050000	0.000000	-0.000066	0.019801	-271905459350724470000000000000000000000000000000000000000000000000000000000.000000	9376050322438774400000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:57.032	Ascend	SBE41 Message						   18.69, 22.0213, 34.1597  	18.690001	18.675516	0.114748	0.000000	-0.010000	0.075058	1.000000																							

$	12-Dec-2018	19:04:57.600	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.690001	147.325896	28.400000	265.500000																																

$	12-Dec-2018	19:04:58.032	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	0.000066	7885258321171009600000000000000000000000000000000000000000000000000000000000.000000	-271905459350724470000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:04:58.032	Ascend	SBE41 Message						   18.69, 22.0213, 34.1599  	18.690001	18.676411	0.071599	0.000000	0.000000	0.076306	1.000000																							

$	12-Dec-2018	19:04:58.601	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.690001	147.308335	28.400000	265.500000																																

$	12-Dec-2018	19:04:59.030	Ascend	Vel PID Values																		0.050000	0.010000	0.009967	0.009967	-228672491313959240000000000000000000000000000000000000000000000000000000000000.000000	7885258321171009600000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:04:59.030	Ascend	SBE41 Message						   18.68, 22.0214, 34.1599  	18.680000	18.678677	0.032646	0.010000	0.005000	0.077045	1.000000																							

$	12-Dec-2018	19:04:59.601	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.680000	147.300601	28.400000	265.500000																																

$	12-Dec-2018	19:05:00.030	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019901	6631502248104818700000000000000000000000000000000000000000000000000000000000000.000000	-228672491313959240000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:00.030	Ascend	SBE41 Message						   18.69, 22.0222, 34.1597  	18.690001	18.681635	0.009489	-0.010000	0.000000	0.077268	1.000000																							

$	12-Dec-2018	19:05:00.603	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.690001	147.317439	28.400000	265.500000																																

$	12-Dec-2018	19:05:01.029	Ascend	Vel PID Values																		0.050000	0.140001	0.139502	0.149436	-192313565195039730000000000000000000000000000000000000000000000000000000000000000.000000	6631502248104818700000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:01.029	Ascend	SBE41 Message						   18.55, 22.0226, 34.1598  	18.549999	18.684658	0.000727	0.140001	0.065001	0.076977	1.000000																							

$	12-Dec-2018	19:05:01.602	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	18.549999	147.304408	28.400000	265.500000																																

$	12-Dec-2018	19:05:02.027	Ascend	Vel PID Values																		0.050000	0.779999	0.777868	0.638366	5577093390656152200000000000000000000000000000000000000000000000000000000000000000.000000	-192313565195039730000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:02.027	Ascend	SBE41 Message						   17.77, 22.0223, 34.1605  	17.770000	18.686305	-0.001149	0.779999	0.460000	0.076177	1.000000																							

$	12-Dec-2018	19:05:02.605	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 17  Pres, BP, BatV, BatI: 	17.770000	147.306524	28.400000	265.500000																																

$	12-Dec-2018	19:05:03.059	Ascend	ERROR: Missed PH message 30

$	12-Dec-2018	19:05:03.063	Ascend	ERROR: Missed NO3 message 26

$	12-Dec-2018	19:05:03.265	Ascend	Sending Optode Do Sample

$	12-Dec-2018	19:05:03.269	Ascend	Sending Optode Wakeup

$	12-Dec-2018	19:05:03.873	Ascend	Sending FLBB run command

$	12-Dec-2018	19:05:03.890	Ascend	Sending radiometer pollSample

$	12-Dec-2018	19:05:03.026	Ascend	Vel PID Values																		0.050000	0.300001	0.301595	-0.476273	-161735708329028410000000000000000000000000000000000000000000000000000000000000000000.000000	5577093390656152200000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:03.026	Ascend	SBE41 Message						   17.47, 22.0235, 34.1601  	17.469999	18.680541	-0.000540	0.300001	0.540000	0.074897	1.000000																							

$	12-Dec-2018	19:05:04.113	Ascend	Vel PID Values																		0.050000	0.010000	0.010968	-0.290627	4690335541541824300000000000000000000000000000000000000000000000000000000000000000000.000000	-161735708329028410000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:04.113	Ascend	SBE41 Message						   17.46, 22.0236, 34.1604  	17.459999	18.642621	0.004108	0.010000	0.155001	0.073228	1.000000																							

$	12-Dec-2018	19:05:04.790	Ascend	Radiometer Message																																	SATFI40287,-0.0058,0.0009,-0.0067,-0.2070 			

$	12-Dec-2018	19:05:05.107	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.459999	147.314787	28.400000	265.500000																																

$	12-Dec-2018	19:05:05.220	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.1672090

$	12-Dec-2018	19:05:05.324	Ascend	Starting ISUS sample

$	12-Dec-2018	19:05:05.427	Ascend	Starting MSC Sequence CTDValues (time, press, temp, salinity) 0, 0.00, 0.000, 0.00

$	12-Dec-2018	19:05:05.432	Ascend	MSC: OUT: W

$	12-Dec-2018	19:05:05.882	Ascend	MSC: OUT: W

$	12-Dec-2018	19:05:04.964	Ascend	Optode Message																												4330,1532,265.175,22.406,29.172,29.103,36.143,7.040,538.1,467.3,113.4  								

$	12-Dec-2018	19:05:04.967	Ascend	Optode Message																												#  								

$	12-Dec-2018	19:05:05.025	Ascend	Vel PID Values																		0.050000	0.000000	0.000035	-0.010933	-136019730704712900000000000000000000000000000000000000000000000000000000000000000000000.000000	4690335541541824300000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:05.025	Ascend	SBE41 Message						   17.46, 22.0240, 34.1578  	17.459999	18.520288	0.019189	0.000000	0.005000	0.071537	1.000000																							

$	12-Dec-2018	19:05:06.461	Ascend	FLBB Message																											/99/99,99:99:99,695,2308,700,4130,533  									

$	12-Dec-2018	19:05:06.670	Ascend	MSC: RX: 

$	12-Dec-2018	19:05:06.771	Ascend	MSC: RX: 

$	12-Dec-2018	19:05:06.097	Ascend	Vel PID Values																		0.050000	0.009998	0.009965	0.009930	3944572190436674100000000000000000000000000000000000000000000000000000000000000000000000.000000	-136019730704712900000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:06.097	Ascend	SBE41 Message						   17.45, 22.0242, 34.1602  	17.450001	18.276831	0.049402	0.009998	0.004999	0.070194	1.000000																							

$	12-Dec-2018	19:05:07.221	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:05:07.225	Ascend	MSC: Got ACK

$	12-Dec-2018	19:05:07.626	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.450001	147.777595	28.400000	265.500000																																

$	12-Dec-2018	19:05:07.739	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.5127018

$	12-Dec-2018	19:05:07.106	Ascend	Vel PID Values																		0.050000	-0.019999	-0.019899	-0.029864	-114392593522663540000000000000000000000000000000000000000000000000000000000000000000000000.000000	3944572190436674100000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:07.106	Ascend	SBE41 Message						   17.47, 22.0242, 34.1596  	17.469999	17.960408	0.088818	-0.019999	-0.005000	0.069442	1.000000																							

$	12-Dec-2018	19:05:08.077	Ascend	Vel PID Values																		0.050000	0.010000	0.009901	0.029799	3317385212157242600000000000000000000000000000000000000000000000000000000000000000000000000.000000	-114392593522663540000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:08.077	Ascend	SBE41 Message						   17.46, 22.0249, 34.1605  	17.459999	17.686188	0.123117	0.010000	-0.004999	0.069133	1.000000																							

$	12-Dec-2018	19:05:08.741	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.459999	148.259610	28.400000	265.500000																																

$	12-Dec-2018	19:05:08.853	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.1082049

$	12-Dec-2018	19:05:08.959	Ascend	MSC: OUT: CTD,0,0.0000,0.0000,0.0000

$	12-Dec-2018	19:05:09.061	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.009867	-96204171152560029000000000000000000000000000000000000000000000000000000000000000000000000000.000000	3317385212157242600000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:09.061	Ascend	SBE41 Message						   17.46, 22.0251, 34.1606  	17.459999	17.528954	0.142966	0.000000	0.005000	0.069043	1.000000																							

$	12-Dec-2018	19:05:09.622	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.459999	148.438908	28.400000	265.500000																																

$	12-Dec-2018	19:05:09.861	Ascend	MSC: RX: NAK,CTD,EConvertErr: 0 

$	12-Dec-2018	19:05:09.963	Ascend	MSC: EConvertErr: 0

$	12-Dec-2018	19:05:10.169	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.459999	148.606107	28.400000	265.500000																																

$	12-Dec-2018	19:05:10.066	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000033	2789920963424240500000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-96204171152560029000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:10.066	Ascend	SBE41 Message						   17.46, 22.0253, 34.1606  	17.459999	17.471890	0.150251	0.000000	0.000000	0.069272	1.000000																							

$	12-Dec-2018	19:05:11.094	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.009967	-80907707939302983000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	2789920963424240500000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:11.094	Ascend	SBE41 Message						   17.47, 22.0254, 34.1610  	17.469999	17.460629	0.150935	-0.010000	-0.005000	0.070156	1.000000																							

$	12-Dec-2018	19:05:11.758	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.469999	148.596272	28.400000	265.500000																																

$	12-Dec-2018	19:05:11.869	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0098111

$	12-Dec-2018	19:05:11.973	Ascend	MSC: OUT: W

$	12-Dec-2018	19:05:12.326	Ascend	MSC: OUT: W

$	12-Dec-2018	19:05:12.075	Ascend	Vel PID Values																		0.050000	0.019999	0.019899	0.029866	2346323530239786700000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-80907707939302983000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:12.075	Ascend	SBE41 Message						   17.45, 22.0261, 34.1601  	17.450001	17.460647	0.146224	0.019999	0.004999	0.071736	1.000000																							

$	12-Dec-2018	19:05:12.680	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:05:12.782	Ascend	MSC: Got ACK

$	12-Dec-2018	19:05:13.185	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.450001	148.610957	28.400000	265.500000																																

$	12-Dec-2018	19:05:13.296	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.4212393

$	12-Dec-2018	19:05:13.399	Ascend	MSC: OUT: MODE,1

$	12-Dec-2018	19:05:13.504	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:05:13.512	Ascend	MSC: Got ACK

$	12-Dec-2018	19:05:13.017	Ascend	Vel PID Values																		0.050000	-0.019999	-0.019866	-0.039764	-68043382376953809000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	2346323530239786700000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:13.017	Ascend	SBE41 Message						   17.47, 22.0269, 34.1592  	17.469999	17.461400	0.130971	-0.019999	0.000000	0.073634	1.000000																							

$	12-Dec-2018	19:05:14.060	Ascend	MSC: RX: ACK,MODE,1 

$	12-Dec-2018	19:05:14.059	Ascend	Vel PID Values																		0.050000	0.010000	0.009901	0.029766	1973258088931660400000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-68043382376953809000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:14.059	Ascend	SBE41 Message						   17.46, 22.0270, 34.1603  	17.459999	17.461605	0.100813	0.010000	-0.004999	0.075384	1.000000																							

$	12-Dec-2018	19:05:14.714	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.459999	148.336776	28.400000	265.500000																																

$	12-Dec-2018	19:05:14.825	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.5239304

$	12-Dec-2018	19:05:15.130	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.459999	148.330532	28.400000	265.500000																																

$	12-Dec-2018	19:05:15.831	Ascend	MSC: OUT: ACK

$	12-Dec-2018	19:05:15.013	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019834	-57224484579018154000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	1973258088931660400000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:15.013	Ascend	SBE41 Message						   17.47, 22.0276, 34.1604  	17.469999	17.461682	0.061663	-0.010000	0.000000	0.076739	1.000000																							

$	12-Dec-2018	19:05:16.013	Ascend	Vel PID Values																		0.050000	0.019999	0.019899	0.029833	1659510052791526500000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-57224484579018154000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:16.013	Ascend	SBE41 Message						   17.45, 22.0285, 34.1599  	17.450001	17.461711	0.027751	0.019999	0.004999	0.077635	1.000000																							

$	12-Dec-2018	19:05:16.833	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.450001	148.312546	28.400000	265.500000																																

$	12-Dec-2018	19:05:16.944	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.1139936

$	12-Dec-2018	19:05:17.049	Ascend	Vel PID Values																		0.050000	-0.019999	-0.019866	-0.039764	-48125791530954271000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	1659510052791526500000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:17.049	Ascend	SBE41 Message						   17.47, 22.0285, 34.1604  	17.469999	17.461887	0.008290	-0.019999	0.000000	0.078038	1.000000																							

$	12-Dec-2018	19:05:17.600	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.469999	148.821495	28.400000	265.500000																																

$	12-Dec-2018	19:05:17.952	Ascend	MSC: OUT: TS

$	12-Dec-2018	19:05:18.055	Ascend	Vel PID Values																		0.050000	0.010000	0.009901	0.029766	1395647954397673800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-48125791530954271000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:18.055	Ascend	SBE41 Message						   17.46, 22.0287, 34.1604  	17.459999	17.462470	0.001164	0.010000	-0.004999	0.077960	1.000000																							

$	12-Dec-2018	19:05:18.403	Ascend	MSC: RX: ACK,TS,CMD 

$	12-Dec-2018	19:05:18.707	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.459999	148.970924	28.400000	265.500000																																

$	12-Dec-2018	19:05:19.155	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.459999	148.954339	28.400000	265.500000																																

$	12-Dec-2018	19:05:19.011	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.009868	-40473790677532541000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	1395647954397673800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:19.011	Ascend	SBE41 Message						   17.46, 22.0284, 34.1610  	17.459999	17.462860	-0.000276	0.000000	0.005000	0.077416	1.000000																							

$	12-Dec-2018	19:05:20.010	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000033	1173739929648443700000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-40473790677532541000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:20.010	Ascend	SBE41 Message						   17.46, 22.0297, 34.1603  	17.459999	17.462470	-0.000226	0.000000	0.000000	0.076396	1.000000																							

$	12-Dec-2018	19:05:20.850	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.459999	148.962080	28.400000	265.500000																																

$	12-Dec-2018	19:05:20.963	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0104704

$	12-Dec-2018	19:05:21.066	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000000	-34038457959804867000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	1173739929648443700000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:21.066	Ascend	SBE41 Message						   17.46, 22.0297, 34.1598  	17.459999	17.461887	-0.000060	0.000000	0.000000	0.074877	1.000000																							

$	12-Dec-2018	19:05:21.617	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.459999	149.411313	28.400000	265.500000																																

$	12-Dec-2018	19:05:22.170	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.459999	149.554968	28.400000	265.500000																																

$	12-Dec-2018	19:05:22.007	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.009967	987115280834341120000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-34038457959804867000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:22.007	Ascend	SBE41 Message						   17.47, 22.0297, 34.1609  	17.469999	17.461682	-0.000010	-0.010000	-0.005000	0.072903	1.000000																							

$	12-Dec-2018	19:05:23.006	Ascend	Vel PID Values																		0.050000	-0.010000	-0.010000	-0.000033	-28626343144195895000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	987115280834341120000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:23.006	Ascend	SBE41 Message						   17.48, 22.0308, 34.1601  	17.480000	17.461459	0.000028	-0.010000	-0.010000	0.070461	1.000000																							

$	12-Dec-2018	19:05:23.871	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.480000	149.558038	28.400000	265.500000																																

$	12-Dec-2018	19:05:23.983	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0152786

$	12-Dec-2018	19:05:24.291	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	17.480000	149.516270	28.400000	265.500000																																

$	12-Dec-2018	19:05:24.089	Ascend	Vel PID Values																		0.050000	0.350000	0.348802	0.358803	830163951181680880000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-28626343144195895000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:24.089	Ascend	SBE41 Message						   17.13, 22.0310, 34.1600  	17.129999	17.461152	0.000069	0.350000	0.170000	0.067491	1.000000																							

$	12-Dec-2018	19:05:25.009	Ascend	Vel PID Values																		0.050000	0.709999	0.708799	0.359997	-24074754584268746000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	830163951181680880000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:25.009	Ascend	SBE41 Message						   16.42, 22.0317, 34.1605  	16.420000	17.460895	0.000123	0.709999	0.530000	0.064068	1.000000																							

$	12-Dec-2018	19:05:25.590	Ascend	MSC: RX: ACK,TS,DAT 

$	12-Dec-2018	19:05:25.993	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.420000	149.310417	28.400000	265.500000																																

$	12-Dec-2018	19:05:26.106	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.1162188

$	12-Dec-2018	19:05:26.105	Ascend	Vel PID Values																		0.050000	0.160000	0.161828	-0.546971	698167882943793590000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-24074754584268746000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:26.105	Ascend	SBE41 Message						   16.26, 22.0315, 34.1609  	16.260000	17.452001	0.001294	0.160000	0.434999	0.060171	1.000000																							

$	12-Dec-2018	19:05:26.760	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.260000	149.008125	28.400000	265.500000																																

$	12-Dec-2018	19:05:27.112	Ascend	MSC: OUT: SL

$	12-Dec-2018	19:05:27.003	Ascend	Vel PID Values																		0.050000	0.000000	0.000535	-0.161293	-20246868605370014000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	698167882943793590000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:27.003	Ascend	SBE41 Message						   16.26, 22.0317, 34.1600  	16.260000	17.399071	0.007885	0.000000	0.080000	0.056151	1.000000																							

$	12-Dec-2018	19:05:27.981	Ascend	Error in I2CPortBase.Execute

$	12-Dec-2018	19:05:28.085	Ascend	2485 Sample out of range

$	12-Dec-2018	19:05:28.187	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.260000	250.000000	28.400000	265.500000																																

$	12-Dec-2018	19:05:28.299	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.1870269

$	12-Dec-2018	19:05:28.921	Ascend	Error in I2CPortBase.Execute

$	12-Dec-2018	19:05:28.926	Ascend	2485 Sample out of range

$	12-Dec-2018	19:05:29.126	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.260000	250.000000	28.400000	265.500000																																

$	12-Dec-2018	19:05:29.305	Ascend	MSC: RX: 0x85EA,A,01/31/2012 01:30:30,0,-1.00,-1.0000,-1.0000,224,145,1,65,22.86,56.23,11.576,5.252e-01,10555.00,2250.00,1176.89,13.19,-294515428540103E463D053C963CFF3E7A4083439F478F4CA952785910600466BD6D4C72A9769A78D479677825759571DC6DD869C265F1626FFE95E585D965D9D5EBD607162F0664E6A346EF7741679A57FBA85908B3290219446972698A0987096BE934D8E978900829A7C3875AF6F9369F264FE60435C51

$	12-Dec-2018	19:05:29.316	Ascend	MSC: Data: 0x85EA,A,01/31/2012 01:30:30,0,-1.00,-1.0000,-1.0000,224,145,1,65,22.86,56.23,11.576,5.252e-01,10555.00,2250.00,1176.89,13.19,-294515428540103E463D053C963CFF3E7A4083439F478F4CA952785910600466BD6D4C72A9769A78D479677825759571DC6DD869C265F1626FFE95E585D965D9D5EBD607162F0664E6A346EF7741679A57FBA85908B3290219446972698A0987096BE934D8E978900829A7C3875AF6F9369F264FE60435C51

$	12-Dec-2018	19:05:29.291	Ascend	Vel PID Values																		0.050000	-0.020000	-0.019933	-0.020469	587159189555730470000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-20246868605370014000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:29.291	Ascend	SBE41 Message						   16.28, 22.0318, 34.1611  	16.280001	17.245999	0.026761	-0.020000	-0.010000	0.052340	1.000000																							

$	12-Dec-2018	19:05:29.292	Ascend	Vel PID Values																		0.050000	0.010000	0.009901	0.029834	-17027616497116184000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	587159189555730470000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:29.292	Ascend	SBE41 Message						   16.27, 22.0319, 34.1613  	16.270000	16.976694	0.059990	0.010000	-0.005000	0.050296	1.000000																							

$	12-Dec-2018	19:05:30.010	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.009867	493800878416369380000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-17027616497116184000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:30.010	Ascend	SBE41 Message						   16.27, 22.0328, 34.1596  	16.270000	16.666552	0.098327	0.000000	0.005000	0.050273	1.000000																							

$	12-Dec-2018	19:05:30.676	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.270000	148.867309	28.400000	265.500000																																

$	12-Dec-2018	19:05:30.787	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.4829696

$	12-Dec-2018	19:05:30.794	Ascend	MSC: Done Sampling.

$	12-Dec-2018	19:05:30.896	Ascend	Starting pH sample

$	12-Dec-2018	19:05:31.096	Ascend	MSC: OUT: W

$	12-Dec-2018	19:05:31.448	Ascend	MSC: OUT: W

$	12-Dec-2018	19:05:31.095	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000033	-14320225474074711000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	493800878416369380000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:31.095	Ascend	SBE41 Message						   16.27, 22.0331, 34.1611  	16.270000	16.428973	0.127705	0.000000	0.000000	0.051795	1.000000																							

$	12-Dec-2018	19:05:31.802	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:05:32.002	Ascend	MSC: Got ACK

$	12-Dec-2018	19:05:32.104	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:05:32.112	Ascend	MSC: Got ACK

$	12-Dec-2018	19:05:32.001	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.009967	415286538748166620000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-14320225474074711000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:32.001	Ascend	SBE41 Message						   16.28, 22.0338, 34.1606  	16.280001	16.311067	0.142279	-0.010000	-0.005000	0.054010	1.000000																							

$	12-Dec-2018	19:05:32.765	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.280001	148.485735	28.400000	265.500000																																

$	12-Dec-2018	19:05:32.876	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.0820012

$	12-Dec-2018	19:05:32.980	Ascend	MSC: OUT: MODE,2

$	12-Dec-2018	19:05:33.088	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019901	-12043309623696833000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	415286538748166620000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:33.088	Ascend	SBE41 Message						   16.27, 22.0335, 34.1613  	16.270000	16.276326	0.146563	0.010000	0.000000	0.056144	1.000000																							

$	12-Dec-2018	19:05:33.652	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.270000	148.467861	28.400000	265.500000																																

$	12-Dec-2018	19:05:33.883	Ascend	MSC: RX: ACK,MODE,2 

$	12-Dec-2018	19:05:34.188	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.270000	148.470059	28.400000	265.500000																																

$	12-Dec-2018	19:05:33.995	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019868	349255979087208160000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-12043309623696833000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:33.995	Ascend	SBE41 Message						   16.28, 22.0344, 34.1610  	16.280001	16.271605	0.146043	-0.010000	0.000000	0.057914	1.000000																							

$	12-Dec-2018	19:05:34.993	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009901	-10128423393529036000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	349255979087208160000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:34.993	Ascend	SBE41 Message						   16.28, 22.0343, 34.1610  	16.280001	16.271861	0.139441	0.000000	-0.005000	0.059244	1.000000																							

$	12-Dec-2018	19:05:35.781	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.280001	148.475267	28.400000	265.500000																																

$	12-Dec-2018	19:05:35.892	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0106357

$	12-Dec-2018	19:05:35.996	Ascend	MSC: OUT: ACK

$	12-Dec-2018	19:05:35.995	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.009934	293724278412342020000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-10128423393529036000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:35.995	Ascend	SBE41 Message						   16.29, 22.0349, 34.1610  	16.290001	16.272782	0.120366	-0.010000	-0.005000	0.060126	1.000000																							

$	12-Dec-2018	19:05:36.651	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.290001	148.474105	28.400000	265.500000																																

$	12-Dec-2018	19:05:37.101	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.290001	148.317330	28.400000	265.500000																																

$	12-Dec-2018	19:05:37.899	Ascend	MSC: OUT: TS

$	12-Dec-2018	19:05:36.992	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019901	-8518004073957919400000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	293724278412342020000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:36.992	Ascend	SBE41 Message						   16.28, 22.0347, 34.1614  	16.280001	16.274242	0.086862	0.010000	0.000000	0.060557	1.000000																							

$	12-Dec-2018	19:05:38.001	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.009901	247022118144779680000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-8518004073957919400000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:38.001	Ascend	SBE41 Message						   16.28, 22.0356, 34.1610  	16.280001	16.276100	0.048274	0.000000	0.005000	0.060529	1.000000																							

$	12-Dec-2018	19:05:38.699	Ascend	MSC: RX: ACK,TS,CMD 

$	12-Dec-2018	19:05:39.004	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.280001	148.316063	28.400000	265.500000																																

$	12-Dec-2018	19:05:39.115	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.2166361

$	12-Dec-2018	19:05:38.989	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000033	-7163641426198610900000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	247022118144779680000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:38.989	Ascend	SBE41 Message						   16.28, 22.0362, 34.1610  	16.280001	16.278541	0.018596	0.000000	0.000000	0.060037	1.000000																							

$	12-Dec-2018	19:05:39.774	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.280001	148.567669	28.400000	265.500000																																

$	12-Dec-2018	19:05:39.989	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.009967	207745601359759720000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-7163641426198610900000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:39.989	Ascend	SBE41 Message						   16.29, 22.0366, 34.1610  	16.290001	16.281120	0.003701	-0.010000	-0.005000	0.059073	1.000000																							

$	12-Dec-2018	19:05:40.675	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.290001	148.631319	28.400000	265.500000																																

$	12-Dec-2018	19:05:40.988	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019901	-6024622439433031800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	207745601359759720000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:40.988	Ascend	SBE41 Message						   16.28, 22.0367, 34.1612  	16.280001	16.282550	-0.000770	0.010000	0.000000	0.057641	1.000000																							

$	12-Dec-2018	19:05:41.673	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.280001	148.634940	28.400000	265.500000																																

$	12-Dec-2018	19:05:41.987	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019868	174714050743557920000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-6024622439433031800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:41.987	Ascend	SBE41 Message						   16.29, 22.0371, 34.1612  	16.290001	16.282471	-0.001344	-0.010000	0.000000	0.055750	1.000000																							

$	12-Dec-2018	19:05:42.676	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.290001	148.717478	28.400000	265.500000																																

$	12-Dec-2018	19:05:42.984	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009901	-5066707471563179900000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	174714050743557920000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:42.984	Ascend	SBE41 Message						   16.29, 22.0379, 34.1608  	16.290001	16.282170	-0.001275	0.000000	-0.005000	0.053428	1.000000																							

$	12-Dec-2018	19:05:43.675	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.290001	148.713380	28.400000	265.500000																																

$	12-Dec-2018	19:05:43.985	Ascend	Vel PID Values																		0.050000	0.010000	0.009967	0.010000	146934516675332220000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-5066707471563179900000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:43.985	Ascend	SBE41 Message						   16.28, 22.0381, 34.1610  	16.280001	16.282812	-0.001241	0.010000	0.005000	0.050785	1.000000																							

$	12-Dec-2018	19:05:44.676	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.280001	148.724504	28.400000	265.500000																																

$	12-Dec-2018	19:05:44.983	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019901	-4261100983584634800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	146934516675332220000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:44.983	Ascend	SBE41 Message						   16.29, 22.0385, 34.1610  	16.290001	16.284183	-0.001229	-0.010000	0.000000	0.048171	1.000000																							

$	12-Dec-2018	19:05:45.676	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.290001	148.719013	28.400000	265.500000																																

$	12-Dec-2018	19:05:45.983	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009901	123571928523954410000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-4261100983584634800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:45.983	Ascend	SBE41 Message						   16.29, 22.0385, 34.1610  	16.290001	16.285536	-0.001167	0.000000	-0.005000	0.046060	1.000000																							

$	12-Dec-2018	19:05:46.676	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.290001	148.765460	28.400000	265.500000																																

$	12-Dec-2018	19:05:46.983	Ascend	Vel PID Values																		0.050000	0.020000	0.019934	0.019967	-3583585927194678000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	123571928523954410000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:46.983	Ascend	SBE41 Message						   16.27, 22.0391, 34.1612  	16.270000	16.286401	-0.000972	0.020000	0.010000	0.044788	1.000000																							

$	12-Dec-2018	19:05:47.677	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	16.270000	148.780316	28.400000	265.500000																																

$	12-Dec-2018	19:05:47.981	Ascend	Vel PID Values																		0.050000	0.480000	0.478470	0.458536	103923991888645650000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-3583585927194678000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:47.981	Ascend	SBE41 Message						   15.79, 22.0395, 34.1612  	15.790000	16.286617	-0.000680	0.480000	0.250000	0.044261	1.000000																							

$	12-Dec-2018	19:05:48.677	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 18  Pres, BP, BatV, BatI: 	15.790000	148.816898	28.400000	265.500000																																

$	12-Dec-2018	19:05:49.130	Ascend	ERROR: Missed PH message 31

$	12-Dec-2018	19:05:49.135	Ascend	ERROR: Missed NO3 message 27

$	12-Dec-2018	19:05:49.336	Ascend	Sending Optode Do Sample

$	12-Dec-2018	19:05:49.340	Ascend	Sending Optode Wakeup

$	12-Dec-2018	19:05:49.955	Ascend	Sending FLBB run command

$	12-Dec-2018	19:05:49.959	Ascend	Sending radiometer pollSample

$	12-Dec-2018	19:05:48.980	Ascend	Vel PID Values																		0.050000	0.670000	0.669365	0.190896	-3013795764770723700000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	103923991888645650000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:48.980	Ascend	SBE41 Message						   15.12, 22.0395, 34.1612  	15.120000	16.284861	-0.000286	0.670000	0.575000	0.044086	1.000000																							

$	12-Dec-2018	19:05:50.418	Ascend	MSC: RX: ACK,TS,DAT 

$	12-Dec-2018	19:05:50.061	Ascend	Vel PID Values																		0.050000	0.070000	0.071995	-0.597370	87400077178350987000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-3013795764770723700000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:50.061	Ascend	SBE41 Message						   15.05, 22.0401, 34.1615  	15.050000	16.268054	0.001782	0.070000	0.370000	0.043940	1.000000																							

$	12-Dec-2018	19:05:50.972	Ascend	Radiometer Message																																	SATFI40287,-0.0041,0.0010,-0.0068,-0.2177 			

$	12-Dec-2018	19:05:51.291	Ascend	PTH Message														793.220459	23.816667	53.963226	24.643501																			

$	12-Dec-2018	19:05:51.494	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.050000	148.800626	28.350000	275.500000																																

$	12-Dec-2018	19:05:51.605	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.4814938

$	12-Dec-2018	19:05:51.085	Ascend	Vel PID Values																		0.050000	0.000000	0.000236	-0.071759	-2534602238172178800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	87400077178350987000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:51.085	Ascend	SBE41 Message						   15.05, 22.0408, 34.1613  	15.050000	16.194362	0.010854	0.000000	0.035000	0.043870	1.000000																							

$	12-Dec-2018	19:05:51.086	Ascend	Optode Message																												4330,1532,265.029,22.412,29.176,29.107,36.146,7.039,538.1,467.0,113.2  								

$	12-Dec-2018	19:05:51.087	Ascend	Optode Message																												#  								

$	12-Dec-2018	19:05:52.189	Ascend	FLBB Message																											/99/99,99:99:99,695,2311,700,4130,533  									

$	12-Dec-2018	19:05:52.059	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.010203	73503464906993179000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-2534602238172178800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:52.059	Ascend	SBE41 Message						   15.06, 22.0409, 34.1614  	15.060000	16.009434	0.033797	-0.010000	-0.005000	0.044251	1.000000																							

$	12-Dec-2018	19:05:52.950	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.060000	149.444766	28.350000	275.500000																																

$	12-Dec-2018	19:05:53.064	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.4531292

$	12-Dec-2018	19:05:53.167	Ascend	MSC: OUT: SL

$	12-Dec-2018	19:05:53.062	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019900	-2131600482302802100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	73503464906993179000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:53.062	Ascend	SBE41 Message						   15.05, 22.0416, 34.1610  	15.050000	15.714709	0.070413	0.010000	0.000000	0.045449	1.000000																							

$	12-Dec-2018	19:05:53.823	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.050000	149.691224	28.350000	275.500000																																

$	12-Dec-2018	19:05:54.070	Ascend	MSC: RX: 0x1068,pH,01/31/2012 01:30:57,0,-1.00,-1.0000,-1.0000,252,145,0,22.89,56.21,11.744,4.640e-02,7.1234,0.000,-1.019563,0.000821,-1..892e-08,+2.124e-07 

$	12-Dec-2018	19:05:54.174	Ascend	MSC: Data: 0x1068,pH,01/31/2012 01:30:57,0,-1.00,-1.0000,-1.0000,252,145,0,22.89,56.21,11.744,4.640e-02,7.1234,0.000,-1.019563,0.000821,-1..892e-08,+2.124e-07 

$	12-Dec-2018	19:05:53.975	Ascend	Vel PID Values																		0.050000	-0.020000	-0.019900	-0.029834	61816413986781257000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-2131600482302802100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:53.975	Ascend	SBE41 Message						   15.07, 22.0420, 34.1612  	15.070000	15.403156	0.109122	-0.020000	-0.005000	0.047464	1.000000																							

$	12-Dec-2018	19:05:54.832	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.070000	149.708875	28.350000	275.500000																																

$	12-Dec-2018	19:05:55.069	Ascend	MSC: Done Sampling.

$	12-Dec-2018	19:05:55.171	Ascend	MSC: OUT: SLP

$	12-Dec-2018	19:05:54.973	Ascend	Vel PID Values																		0.050000	0.020000	0.019867	0.039766	-1792676005616656400000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	61816413986781257000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:54.973	Ascend	SBE41 Message						   15.05, 22.0415, 34.1620  	15.050000	15.183871	0.136375	0.020000	0.000000	0.049907	1.000000																							

$	12-Dec-2018	19:05:55.826	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.050000	149.711564	28.350000	275.500000																																

$	12-Dec-2018	19:05:56.069	Ascend	MSC: RX: ACK,SLP 

$	12-Dec-2018	19:05:55.972	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009901	-0.029767	51987604162883033000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-1792676005616656400000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:55.972	Ascend	SBE41 Message						   15.06, 22.0426, 34.1613  	15.060000	15.084740	0.148691	-0.010000	0.005000	0.052287	1.000000																							

$	12-Dec-2018	19:05:56.723	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.060000	149.732545	28.350000	275.500000																																

$	12-Dec-2018	19:05:56.971	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009868	-1507640520723607800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	51987604162883033000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:56.971	Ascend	SBE41 Message						   15.06, 22.0433, 34.1611  	15.060000	15.059145	0.151655	0.000000	-0.005000	0.054308	1.000000																							

$	12-Dec-2018	19:05:57.621	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.060000	149.752036	28.350000	275.500000																																

$	12-Dec-2018	19:05:57.970	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	0.000033	43721575100984626000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-1507640520723607800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:57.970	Ascend	SBE41 Message						   15.06, 22.0434, 34.1612  	15.060000	15.057184	0.149810	0.000000	0.000000	0.055880	1.000000																							

$	12-Dec-2018	19:05:58.621	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.060000	149.739869	28.350000	275.500000																																

$	12-Dec-2018	19:05:58.970	Ascend	Vel PID Values																		0.050000	-0.009999	-0.009966	-0.009966	-1267925677928554100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	43721575100984626000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:05:58.970	Ascend	SBE41 Message						   15.07, 22.0432, 34.1617  	15.070000	15.058055	0.140560	-0.009999	-0.005000	0.057005	1.000000																							

$	12-Dec-2018	19:05:59.635	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.070000	149.754770	28.350000	275.500000																																

$	12-Dec-2018	19:05:59.969	Ascend	Vel PID Values																		0.050000	0.020000	0.019900	0.029866	36769844659928071000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-1267925677928554100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:05:59.969	Ascend	SBE41 Message						   15.05, 22.0442, 34.1611  	15.050000	15.058347	0.117625	0.020000	0.005000	0.057678	1.000000																							

$	12-Dec-2018	19:06:00.620	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.050000	149.786249	28.350000	275.500000																																

$	12-Dec-2018	19:06:00.968	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009901	-0.029800	-1066325495137914100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	36769844659928071000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:00.968	Ascend	SBE41 Message						   15.06, 22.0442, 34.1615  	15.060000	15.058822	0.081101	-0.010000	0.005000	0.057884	1.000000																							

$	12-Dec-2018	19:06:01.621	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.060000	149.780892	28.350000	275.500000																																

$	12-Dec-2018	19:06:01.966	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009867	30923439358999509000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-1066325495137914100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:01.966	Ascend	SBE41 Message						   15.06, 22.0441, 34.1616  	15.060000	15.059971	0.042429	0.000000	-0.005000	0.057644	1.000000																							

$	12-Dec-2018	19:06:02.622	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.060000	149.877355	28.350000	275.500000																																

$	12-Dec-2018	19:06:02.965	Ascend	Vel PID Values																		0.050000	-0.009999	-0.009966	-0.009933	-896779741410985780000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	30923439358999509000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:02.965	Ascend	SBE41 Message						   15.07, 22.0448, 34.1617  	15.070000	15.060896	0.015202	-0.009999	-0.005000	0.056945	1.000000																							

$	12-Dec-2018	19:06:03.635	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.070000	149.892733	28.350000	275.500000																																

$	12-Dec-2018	19:06:03.964	Ascend	Vel PID Values																		0.050000	0.009999	0.009933	0.019899	26006612500918587000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-896779741410985780000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:03.964	Ascend	SBE41 Message						   15.06, 22.0446, 34.1619  	15.060000	15.060584	0.002986	0.009999	0.000000	0.055777	1.000000																							

$	12-Dec-2018	19:06:04.620	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.060000	149.910495	28.350000	275.500000																																

$	12-Dec-2018	19:06:04.963	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.009900	-754191762526639030000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	26006612500918587000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:04.963	Ascend	SBE41 Message						   15.06, 22.0457, 34.1615  	15.060000	15.059698	-0.000069	0.000000	0.005000	0.054169	1.000000																							

$	12-Dec-2018	19:06:05.621	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.060000	149.886474	28.350000	275.500000																																

$	12-Dec-2018	19:06:05.961	Ascend	Vel PID Values																		0.050000	-0.009999	-0.009966	-0.009999	21871561113272532000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-754191762526639030000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:05.961	Ascend	SBE41 Message						   15.07, 22.0458, 34.1617  	15.070000	15.059922	-0.000339	-0.009999	-0.005000	0.052125	1.000000																							

$	12-Dec-2018	19:06:06.631	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.070000	149.938874	28.350000	275.500000																																

$	12-Dec-2018	19:06:06.962	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009933	-634275272284903440000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	21871561113272532000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:06.962	Ascend	SBE41 Message						   15.07, 22.0467, 34.1611  	15.070000	15.061460	-0.000421	0.000000	-0.005000	0.049836	1.000000																							

$	12-Dec-2018	19:06:07.636	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.070000	149.930768	28.350000	275.500000																																

$	12-Dec-2018	19:06:07.959	Ascend	Vel PID Values																		0.050000	0.009999	0.009966	0.009999	18393982896262200000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-634275272284903440000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:07.959	Ascend	SBE41 Message						   15.06, 22.0470, 34.1614  	15.060000	15.062861	-0.000558	0.009999	0.005000	0.047573	1.000000																							

$	12-Dec-2018	19:06:08.621	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.060000	149.999097	28.350000	275.500000																																

$	12-Dec-2018	19:06:08.959	Ascend	Vel PID Values																		0.050000	-0.009999	-0.009933	-0.019899	-533425503991603760000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	18393982896262200000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:08.959	Ascend	SBE41 Message						   15.07, 22.0469, 34.1616  	15.070000	15.063541	-0.000583	-0.009999	0.000000	0.046027	1.000000																							

$	12-Dec-2018	19:06:09.626	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.070000	150.021374	28.350000	275.500000																																

$	12-Dec-2018	19:06:09.958	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009900	15469339615756510000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-533425503991603760000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:09.958	Ascend	SBE41 Message						   15.07, 22.0473, 34.1616  	15.070000	15.064387	-0.000546	0.000000	-0.005000	0.045350	1.000000																							

$	12-Dec-2018	19:06:10.620	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.070000	150.078386	28.350000	275.500000																																

$	12-Dec-2018	19:06:10.957	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	0.000033	-448610848856938800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	15469339615756510000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:10.957	Ascend	SBE41 Message						   15.07, 22.0475, 34.1618  	15.070000	15.065613	-0.000584	0.000000	0.000000	0.045266	1.000000																							

$	12-Dec-2018	19:06:11.621	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	15.070000	150.053866	28.350000	275.500000																																

$	12-Dec-2018	19:06:11.956	Ascend	Vel PID Values																		0.050000	0.110000	0.109634	0.109634	13009714616851226000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-448610848856938800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:11.956	Ascend	SBE41 Message						   14.96, 22.0473, 34.1621  	14.960000	15.066428	-0.000723	0.110000	0.055000	0.045396	1.000000																							

$	12-Dec-2018	19:06:12.622	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	14.960000	150.171213	28.350000	275.500000																																

$	12-Dec-2018	19:06:12.955	Ascend	Vel PID Values																		0.050000	0.790000	0.787736	0.678103	-377281723888685590000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	13009714616851226000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:12.955	Ascend	SBE41 Message						   14.17, 22.0480, 34.1619  	14.170000	15.066549	-0.000847	0.790000	0.450000	0.045404	1.000000																							

$	12-Dec-2018	19:06:13.623	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	14.170000	150.161929	28.350000	275.500000																																

$	12-Dec-2018	19:06:13.955	Ascend	Vel PID Values																		0.050000	0.360000	0.361427	-0.426310	10941169992771882000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-377281723888685590000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:13.955	Ascend	SBE41 Message						   13.81, 22.0483, 34.1623  	13.810000	15.061979	-0.000255	0.360000	0.575000	0.045241	1.000000																							

$	12-Dec-2018	19:06:14.624	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 19  Pres, BP, BatV, BatI: 	13.810000	150.159292	28.350000	275.500000																																

$	12-Dec-2018	19:06:15.081	Ascend	ERROR: Missed NO3 message 28

$	12-Dec-2018	19:06:15.185	Ascend	Sending Optode Do Sample

$	12-Dec-2018	19:06:15.189	Ascend	Sending Optode Wakeup

$	12-Dec-2018	19:06:15.791	Ascend	Sending FLBB run command

$	12-Dec-2018	19:06:15.808	Ascend	Sending radiometer pollSample

$	12-Dec-2018	19:06:14.952	Ascend	Vel PID Values																		0.050000	0.010000	0.011167	-0.350259	-317293929790384540000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	10941169992771882000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:14.952	Ascend	SBE41 Message						   13.80, 22.0485, 34.1624  	13.800000	15.028066	0.004127	0.010000	0.185000	0.044974	1.000000																							

$	12-Dec-2018	19:06:16.014	Ascend	Vel PID Values																		0.050000	-0.020000	-0.019898	-0.031065	9201523963921152500000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-317293929790384540000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:16.014	Ascend	SBE41 Message						   13.82, 22.0492, 34.1617  	13.820000	14.909772	0.018925	-0.020000	-0.005000	0.044889	1.000000																							

$	12-Dec-2018	19:06:16.618	Ascend	Radiometer Message																																	SATFI40287,-0.0050,0.0010,-0.0072,-0.2143 			

$	12-Dec-2018	19:06:17.036	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.820000	150.148906	28.350000	275.500000																																

$	12-Dec-2018	19:06:17.148	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.0795004

$	12-Dec-2018	19:06:17.251	Ascend	Starting ISUS sample

$	12-Dec-2018	19:06:17.354	Ascend	Starting MSC Sequence CTDValues (time, press, temp, salinity) 0, 0.00, 0.000, 0.00

$	12-Dec-2018	19:06:17.361	Ascend	MSC: OUT: W

$	12-Dec-2018	19:06:17.811	Ascend	MSC: OUT: W

$	12-Dec-2018	19:06:16.882	Ascend	Optode Message																												4330,1532,265.089,22.424,29.168,29.099,36.140,7.041,538.0,466.7,112.8  								

$	12-Dec-2018	19:06:16.885	Ascend	Optode Message																												#  								

$	12-Dec-2018	19:06:16.952	Ascend	Vel PID Values																		0.050000	0.000000	-0.000066	0.019831	-266844194953713430000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	9201523963921152500000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:16.952	Ascend	SBE41 Message						   13.82, 22.0498, 34.1619  	13.820000	14.665247	0.049243	0.000000	-0.010000	0.045452	1.000000																							

$	12-Dec-2018	19:06:18.393	Ascend	FLBB Message																											/99/99,99:99:99,695,2310,700,4130,532  									

$	12-Dec-2018	19:06:18.601	Ascend	MSC: RX: 

$	12-Dec-2018	19:06:18.703	Ascend	MSC: RX: 

$	12-Dec-2018	19:06:18.026	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	0.000066	7738481653657690200000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-266844194953713430000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:18.026	Ascend	SBE41 Message						   13.82, 22.0505, 34.1617  	13.820000	14.339202	0.089672	0.000000	0.000000	0.046888	1.000000																							

$	12-Dec-2018	19:06:19.156	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:06:19.259	Ascend	MSC: Got ACK

$	12-Dec-2018	19:06:19.466	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.820000	150.789015	28.350000	275.500000																																

$	12-Dec-2018	19:06:19.577	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.4236130

$	12-Dec-2018	19:06:19.042	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	0.000000	-224415967956073020000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	7738481653657690200000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:19.042	Ascend	SBE41 Message						   13.82, 22.0496, 34.1626  	13.820000	14.051275	0.125451	0.000000	0.000000	0.049007	1.000000																							

$	12-Dec-2018	19:06:20.034	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	0.000000	6508063070726117800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-224415967956073020000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:20.034	Ascend	SBE41 Message						   13.82, 22.0508, 34.1619  	13.820000	13.885450	0.146091	0.000000	0.000000	0.051305	1.000000																							

$	12-Dec-2018	19:06:20.588	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.820000	151.229903	28.350000	275.500000																																

$	12-Dec-2018	19:06:20.699	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.1165142

$	12-Dec-2018	19:06:20.804	Ascend	MSC: OUT: CTD,0,0.0000,0.0000,0.0000

$	12-Dec-2018	19:06:21.119	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.820000	151.435062	28.350000	275.500000																																

$	12-Dec-2018	19:06:20.948	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000000	-188733829051057400000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	6508063070726117800000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:20.948	Ascend	SBE41 Message						   13.82, 22.0503, 34.1626  	13.820000	13.827738	0.153270	0.000000	0.000000	0.053369	1.000000																							

$	12-Dec-2018	19:06:21.959	Ascend	MSC: RX: NAK,CTD,EConvertErr: 0 

$	12-Dec-2018	19:06:22.160	Ascend	MSC: EConvertErr: 0

$	12-Dec-2018	19:06:22.057	Ascend	Vel PID Values																		0.050000	0.020000	0.019933	0.019933	5473281042480664700000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-188733829051057400000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:22.057	Ascend	SBE41 Message						   13.80, 22.0508, 34.1626  	13.800000	13.818628	0.153836	0.020000	0.010000	0.054991	1.000000																							

$	12-Dec-2018	19:06:22.814	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.800000	151.653454	28.350000	275.500000																																

$	12-Dec-2018	19:06:22.928	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.2234075

$	12-Dec-2018	19:06:23.031	Ascend	Vel PID Values																		0.050000	0.020000	0.020000	0.000067	-158725150231939280000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	5473281042480664700000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:23.031	Ascend	SBE41 Message						   13.78, 22.0510, 34.1624  	13.780000	13.819353	0.149533	0.020000	0.020000	0.056152	1.000000																							

$	12-Dec-2018	19:06:23.586	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.780000	151.888341	28.350000	275.500000																																

$	12-Dec-2018	19:06:23.934	Ascend	MSC: OUT: W

$	12-Dec-2018	19:06:24.189	Ascend	MSC: OUT: W

$	12-Dec-2018	19:06:24.036	Ascend	Vel PID Values																		0.050000	0.020000	0.020000	-0.000001	4603029356726238900000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-158725150231939280000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:24.036	Ascend	SBE41 Message						   13.76, 22.0518, 34.1616  	13.760000	13.819258	0.134883	0.020000	0.020000	0.056847	1.000000																							

$	12-Dec-2018	19:06:24.641	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:06:24.649	Ascend	MSC: Got ACK

$	12-Dec-2018	19:06:24.849	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:06:24.853	Ascend	MSC: Got ACK

$	12-Dec-2018	19:06:24.952	Ascend	Vel PID Values																		0.050000	0.020000	0.020000	0.000001	-133487851345060930000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	4603029356726238900000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:24.952	Ascend	SBE41 Message						   13.74, 22.0518, 34.1611  	13.740000	13.816555	0.104951	0.020000	0.020000	0.057100	1.000000																							

$	12-Dec-2018	19:06:25.605	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.740000	152.021743	28.350000	275.500000																																

$	12-Dec-2018	19:06:25.718	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.7851264

$	12-Dec-2018	19:06:25.822	Ascend	MSC: OUT: MODE,1

$	12-Dec-2018	19:06:26.129	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.740000	152.015835	28.350000	275.500000																																

$	12-Dec-2018	19:06:25.942	Ascend	Vel PID Values																		0.050000	0.010000	0.010034	-0.009967	3871147689006767000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-133487851345060930000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:25.942	Ascend	SBE41 Message						   13.73, 22.0519, 34.1625  	13.730000	13.809357	0.065510	0.010000	0.015000	0.056913	1.000000																							

$	12-Dec-2018	19:06:26.981	Ascend	MSC: RX: ACK,MODE,1 

$	12-Dec-2018	19:06:26.963	Ascend	Vel PID Values																		0.050000	0.009999	0.009999	-0.000034	-112263282981196250000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	3871147689006767000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:26.963	Ascend	SBE41 Message						   13.72, 22.0525, 34.1624  	13.720000	13.796585	0.031486	0.009999	0.010000	0.056300	1.000000																							

$	12-Dec-2018	19:06:27.734	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.720000	151.993856	28.350000	275.500000																																

$	12-Dec-2018	19:06:27.846	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.1226075

$	12-Dec-2018	19:06:27.852	Ascend	MSC: OUT: ACK

$	12-Dec-2018	19:06:27.951	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.009966	3255635206454691400000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-112263282981196250000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:27.951	Ascend	SBE41 Message						   13.72, 22.0534, 34.1620  	13.720000	13.779659	0.013079	0.000000	0.005000	0.055283	1.000000																							

$	12-Dec-2018	19:06:28.609	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.720000	151.844315	28.350000	275.500000																																

$	12-Dec-2018	19:06:29.055	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.720000	151.779458	28.350000	275.500000																																

$	12-Dec-2018	19:06:29.853	Ascend	MSC: OUT: TS

$	12-Dec-2018	19:06:28.939	Ascend	Vel PID Values																		0.050000	0.010000	0.009967	0.009934	-94413420987186053000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	3255635206454691400000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:28.939	Ascend	SBE41 Message						   13.71, 22.0535, 34.1622  	13.710000	13.761692	0.008168	0.010000	0.005000	0.053869	1.000000																							

$	12-Dec-2018	19:06:29.955	Ascend	Vel PID Values																		0.050000	-0.020000	-0.019900	-0.029867	2737989208628395500000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-94413420987186053000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:29.955	Ascend	SBE41 Message						   13.73, 22.0535, 34.1624  	13.730000	13.745633	0.009029	-0.020000	-0.005000	0.052091	1.000000																							

$	12-Dec-2018	19:06:30.662	Ascend	MSC: RX: ACK,TS,CMD 

$	12-Dec-2018	19:06:30.967	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.730000	151.778519	28.350000	275.500000																																

$	12-Dec-2018	19:06:31.080	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.2269364

$	12-Dec-2018	19:06:30.937	Ascend	Vel PID Values																		0.050000	0.009999	0.009900	0.029799	-79401687050223470000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	2737989208628395500000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:30.937	Ascend	SBE41 Message						   13.72, 22.0540, 34.1621  	13.720000	13.733192	0.010658	0.009999	-0.005000	0.050170	1.000000																							

$	12-Dec-2018	19:06:31.739	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.720000	152.047738	28.350000	275.500000																																

$	12-Dec-2018	19:06:31.936	Ascend	Vel PID Values																		0.050000	-0.009999	-0.009933	-0.019833	2302648924456480600000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-79401687050223470000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:31.936	Ascend	SBE41 Message						   13.73, 22.0543, 34.1621  	13.730000	13.724875	0.011676	-0.009999	0.000000	0.048531	1.000000																							

$	12-Dec-2018	19:06:32.738	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.730000	152.144030	28.350000	275.500000																																

$	12-Dec-2018	19:06:32.936	Ascend	Vel PID Values																		0.050000	0.020000	0.019900	0.029833	-66776818809237942000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	2302648924456480600000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:32.936	Ascend	SBE41 Message						   13.71, 22.0543, 34.1624  	13.710000	13.720633	0.011866	0.020000	0.005000	0.047609	1.000000																							

$	12-Dec-2018	19:06:33.740	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.710000	152.153090	28.350000	275.500000																																

$	12-Dec-2018	19:06:33.934	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009901	-0.029800	1936527745467900100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-66776818809237942000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:33.934	Ascend	SBE41 Message						   13.72, 22.0549, 34.1623  	13.720000	13.720010	0.011052	-0.010000	0.005000	0.047527	1.000000																							

$	12-Dec-2018	19:06:34.741	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.720000	152.158327	28.350000	275.500000																																

$	12-Dec-2018	19:06:34.933	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009867	-56159304618569110000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	1936527745467900100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:34.933	Ascend	SBE41 Message						   13.72, 22.0554, 34.1624  	13.720000	13.721401	0.009300	0.000000	-0.005000	0.047967	1.000000																							

$	12-Dec-2018	19:06:35.740	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.720000	152.158201	28.350000	275.500000																																

$	12-Dec-2018	19:06:35.932	Ascend	Vel PID Values																		0.050000	-0.009999	-0.009966	-0.009933	1628619833938504000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-56159304618569110000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:35.932	Ascend	SBE41 Message						   13.73, 22.0558, 34.1623  	13.730000	13.722297	0.007095	-0.009999	-0.005000	0.048512	1.000000																							

$	12-Dec-2018	19:06:36.742	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.730000	152.175076	28.350000	275.500000																																

$	12-Dec-2018	19:06:37.186	Ascend	MSC: RX: ACK,TS,DAT 

$	12-Dec-2018	19:06:36.929	Ascend	Vel PID Values																		0.050000	0.009999	0.009933	0.019899	-47229975184216620000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	1628619833938504000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:36.929	Ascend	SBE41 Message						   13.72, 22.0567, 34.1617  	13.720000	13.721401	0.004983	0.009999	0.000000	0.048903	1.000000																							

$	12-Dec-2018	19:06:37.844	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.720000	152.221009	28.350000	275.500000																																

$	12-Dec-2018	19:06:37.954	Ascend	Vel PID Values																		0.050000	-0.009999	-0.009933	-0.019866	1369669280342282000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-47229975184216620000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:37.954	Ascend	SBE41 Message						   13.73, 22.0563, 34.1625  	13.730000	13.719951	0.003175	-0.009999	0.000000	0.049051	1.000000																							

$	12-Dec-2018	19:06:38.740	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.730000	152.288765	28.350000	275.500000																																

$	12-Dec-2018	19:06:39.186	Ascend	MSC: OUT: SL

$	12-Dec-2018	19:06:38.929	Ascend	Vel PID Values																		0.050000	0.549999	0.548136	0.558069	-39720409129926175000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	1369669280342282000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:38.929	Ascend	SBE41 Message						   13.18, 22.0566, 34.1626  	13.180000	13.719972	0.001634	0.549999	0.270000	0.048949	1.000000																							

$	12-Dec-2018	19:06:39.900	Ascend	Error in I2CPortBase.Execute

$	12-Dec-2018	19:06:39.904	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.180000	152.288765	28.350000	275.500000																																

$	12-Dec-2018	19:06:40.726	Ascend	Error in I2CPortBase.Execute

$	12-Dec-2018	19:06:40.830	Ascend	2485 Sample out of range

$	12-Dec-2018	19:06:40.932	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.180000	250.000000	28.350000	275.500000																																

$	12-Dec-2018	19:06:41.187	Ascend	MSC: RX: 0x802B,A,01/31/2012 01:31:42,0,-1.00,-1.0000,-1.0000,225,145,1,65,22.95,56.33,11.576,5.248e-01,10572.00,2247.00,1183.35,13.05,-2FF73E193CE83C863CD93E3C4051436247724CA2526158DD5FD366916D22728976657888792177F0756D71C76DB5697765A4622E5FB85E245D625D6F5E9E604762DE662A6A096ED273F2798B7F9A855B8AE08FFB940596E098709833966E93038E7C88E182727BE7757A6F8A69D164CF60285C2F58AF55EA538E51D050114EF74E354DEF4DD3,6652.2 

$	12-Dec-2018	19:06:41.296	Ascend	MSC: Data: 0x802B,A,01/31/2012 01:31:42,0,-1.00,-1.0000,-1.0000,225,145,1,65,22.95,56.33,11.576,5.248e-01,10572.00,2247.00,1183.35,13.05,-2FF73E193CE83C863CD93E3C4051436247724CA2526158DD5FD366916D22728976657888792177F0756D71C76DB5697765A4622E5FB85E245D625D6F5E9E604762DE662A6A096ED273F2798B7F9A855B8AE08FFB940596E098709833966E93038E7C88E182727BE7757A6F8A69D164CF60285C2F58AF55EA538E51D050114EF74E354DEF4DD3,6652.2 

$	12-Dec-2018	19:06:41.614	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.180000	152.453199	28.350000	275.500000																																

$	12-Dec-2018	19:06:42.187	Ascend	MSC: Done Sampling.

$	12-Dec-2018	19:06:42.191	Ascend	Starting pH sample

$	12-Dec-2018	19:06:42.390	Ascend	MSC: OUT: W

$	12-Dec-2018	19:06:42.744	Ascend	MSC: OUT: W

$	12-Dec-2018	19:06:42.846	Ascend	MSC: RX: 99999999999999999999999999999999999999999999999999999999

$	12-Dec-2018	19:06:41.443	Ascend	Vel PID Values																		0.050000	0.740001	0.739365	0.191230	1151891864767859000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-39720409129926175000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:41.443	Ascend	SBE41 Message						   12.44, 22.0567, 34.1626  	12.440000	13.720100	0.000590	0.740001	0.645000	0.048213	1.000000																							

$	12-Dec-2018	19:06:41.443	Ascend	Vel PID Values																		0.050000	0.040000	0.042328	-0.697037	-33404864078267915000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	1151891864767859000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:41.443	Ascend	SBE41 Message						   12.40, 22.0575, 34.1627  	12.400000	13.704607	0.001981	0.040000	0.390000	0.047562	1.000000																							

$	12-Dec-2018	19:06:41.926	Ascend	Vel PID Values																		0.050000	0.000000	0.000137	-0.042191	968741058269769550000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-33404864078267915000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:41.926	Ascend	SBE41 Message						   12.40, 22.0575, 34.1630  	12.400000	13.626658	0.011540	0.000000	0.020000	0.047067	1.000000																							

$	12-Dec-2018	19:06:44.011	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:06:44.017	Ascend	MSC: Got ACK

$	12-Dec-2018	19:06:42.950	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000137	-28093490689823321000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	968741058269769550000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:42.950	Ascend	SBE41 Message						   12.40, 22.0578, 34.1631  	12.400000	13.426059	0.036513	0.000000	0.000000	0.047019	1.000000																							

$	12-Dec-2018	19:06:44.568	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:06:44.573	Ascend	MSC: Got ACK

$	12-Dec-2018	19:06:44.973	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	12.400000	152.260661	28.350000	275.500000																																

$	12-Dec-2018	19:06:45.086	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.9002378

$	12-Dec-2018	19:06:45.190	Ascend	MSC: OUT: MODE,2

$	12-Dec-2018	19:06:44.011	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	-0.000000	814711230004876250000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-28093490689823321000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:44.011	Ascend	SBE41 Message						   12.40, 22.0586, 34.1627  	12.400000	13.104990	0.076327	0.000000	0.000000	0.047820	1.000000																							

$	12-Dec-2018	19:06:45.853	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	12.400000	151.811838	28.350000	275.500000																																

$	12-Dec-2018	19:06:44.923	Ascend	Vel PID Values																		0.050000	0.009999	0.009966	0.009966	-23626625670141410000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	814711230004876250000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:44.923	Ascend	SBE41 Message						   12.39, 22.0596, 34.1622  	12.390000	12.767630	0.117950	0.009999	0.005000	0.049492	1.000000																							

$	12-Dec-2018	19:06:46.446	Ascend	MSC: RX: ACK,MODE,2 

$	12-Dec-2018	19:06:45.963	Ascend	Vel PID Values																		0.050000	0.080000	0.079767	0.069801	685172144434100890000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-23626625670141410000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:45.963	Ascend	SBE41 Message						   12.31, 22.0585, 34.1634  	12.310000	12.533673	0.146734	0.080000	0.045000	0.051583	1.000000																							

$	12-Dec-2018	19:06:47.107	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	12.310000	151.664138	28.350000	275.500000																																

$	12-Dec-2018	19:06:47.221	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.1293833

$	12-Dec-2018	19:06:46.920	Ascend	Vel PID Values																		0.050000	0.100000	0.099933	0.020167	-19869992188588926000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	685172144434100890000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:46.920	Ascend	SBE41 Message						   12.21, 22.0600, 34.1625  	12.210000	12.429651	0.159634	0.100000	0.090000	0.053581	1.000000																							

$	12-Dec-2018	19:06:47.879	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	12.210000	151.656613	28.350000	275.500000																																

$	12-Dec-2018	19:06:48.226	Ascend	MSC: OUT: ACK

$	12-Dec-2018	19:06:47.991	Ascend	Vel PID Values																		0.050000	-0.840000	-0.836872	-0.936806	576229773469078870000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-19869992188588926000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:47.991	Ascend	SBE41 Message						   13.05, 22.0605, 34.1622  	13.050000	12.399292	0.163420	-0.840000	-0.370000	0.055189	1.000000																							

$	12-Dec-2018	19:06:48.887	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.050000	151.666939	28.350000	275.500000																																

$	12-Dec-2018	19:06:48.998	Ascend	Vel PID Values																		0.050000	-0.110000	-0.112424	0.724448	-16710663430603287000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	576229773469078870000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:48.998	Ascend	SBE41 Message						   13.16, 22.0603, 34.1624  	13.160000	12.384859	0.163278	-0.110000	-0.475000	0.056352	1.000000																							

$	12-Dec-2018	19:06:49.782	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.160000	151.647545	28.350000	275.500000																																

$	12-Dec-2018	19:06:50.227	Ascend	MSC: OUT: TS

$	12-Dec-2018	19:06:49.917	Ascend	Vel PID Values																		0.050000	0.020000	0.019563	0.131987	484609239487495260000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-16710663430603287000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:49.917	Ascend	SBE41 Message						   13.14, 22.0600, 34.1628  	13.140000	12.377481	0.154521	0.020000	-0.045000	0.057053	1.000000																							

$	12-Dec-2018	19:06:50.887	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.140000	151.645906	28.350000	275.500000																																

$	12-Dec-2018	19:06:51.227	Ascend	MSC: RX: ACK,TS,CMD 

$	12-Dec-2018	19:06:50.998	Ascend	Vel PID Values																		0.050000	0.000000	0.000066	-0.019497	-14053667945137362000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	484609239487495260000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:50.998	Ascend	SBE41 Message						   13.14, 22.0609, 34.1627  	13.140000	12.423568	0.123985	0.000000	0.010000	0.057067	1.000000																							

$	12-Dec-2018	19:06:51.886	Ascend	PTH Message														798.303711	23.716667	54.077667	24.643501																			

$	12-Dec-2018	19:06:52.090	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.140000	151.637383	28.325000	227.500000																																

$	12-Dec-2018	19:06:51.980	Ascend	Vel PID Values																		0.050000	-0.009999	-0.009966	-0.010032	407556370408983500000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-14053667945137362000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:51.980	Ascend	SBE41 Message						   13.15, 22.0609, 34.1628  	13.150000	12.574261	0.065633	-0.009999	-0.005000	0.056040	1.000000																							

$	12-Dec-2018	19:06:52.783	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.150000	151.645243	28.325000	227.500000																																

$	12-Dec-2018	19:06:52.915	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009933	-11819134741860521000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	407556370408983500000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:52.915	Ascend	SBE41 Message						   13.15, 22.0614, 34.1625  	13.150000	12.800855	-0.004109	0.000000	-0.005000	0.053737	1.000000																							

$	12-Dec-2018	19:06:53.783	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	13.150000	151.636317	28.325000	227.500000																																

$	12-Dec-2018	19:06:53.914	Ascend	Vel PID Values																		0.050000	0.570000	0.568103	0.568136	342754907513955120000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-11819134741860521000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:53.914	Ascend	SBE41 Message						   12.58, 22.0618, 34.1626  	12.580000	13.000845	-0.057780	0.570000	0.285000	0.050295	1.000000																							

$	12-Dec-2018	19:06:54.784	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	12.580000	151.648983	28.325000	227.500000																																

$	12-Dec-2018	19:06:54.911	Ascend	Vel PID Values																		0.050000	0.900000	0.898898	0.330795	-9939892317904699100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	342754907513955120000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:54.911	Ascend	SBE41 Message						   11.68, 22.0618, 34.1631  	11.680000	13.105819	-0.083646	0.900000	0.735000	0.046208	1.000000																							

$	12-Dec-2018	19:06:55.787	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 20  Pres, BP, BatV, BatI: 	11.680000	151.660055	28.325000	227.500000																																

$	12-Dec-2018	19:06:56.234	Ascend	ERROR: Missed PH message 32

$	12-Dec-2018	19:06:56.337	Ascend	Sending Optode Do Sample

$	12-Dec-2018	19:06:56.440	Ascend	Sending Optode Wakeup

$	12-Dec-2018	19:06:56.946	Ascend	Sending FLBB run command

$	12-Dec-2018	19:06:56.963	Ascend	Sending radiometer pollSample

$	12-Dec-2018	19:06:55.913	Ascend	Vel PID Values																		0.050000	0.060000	0.062794	-0.836104	288256877219236250000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.00000	-9939892317904699100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:55.913	Ascend	SBE41 Message						   11.62, 22.0618, 34.1635  	11.620000	13.120026	-0.089178	0.060000	0.480000	0.042152	1.000000																							

$	12-Dec-2018	19:06:56.910	Ascend	Vel PID Values																		0.050000	-0.020000	-0.019730	-0.082523	-8359449439357851500000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000	288256877219236250000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.00000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:56.910	Ascend	SBE41 Message						   11.64, 22.0623, 34.1634  	11.640000	13.039750	-0.081046	-0.020000	0.020000	0.038992	1.000000																							

$	12-Dec-2018	19:06:57.872	Ascend	Radiometer Message																																	SATFI40287,-0.0045,0.0010,-0.0070,-0.2037 			

$	12-Dec-2018	19:06:57.985	Ascend	Vel PID Values																		0.050000	0.010000	0.009901	0.029630	242424033741377730000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.00	-8359449439357851500000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:57.985	Ascend	SBE41 Message						   11.63, 22.0624, 34.1635  	11.630000	12.817404	-0.054456	0.010000	-0.005000	0.037571	1.000000																							

$	12-Dec-2018	19:06:58.453	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.630000	151.641943	28.325000	227.500000																																

$	12-Dec-2018	19:06:58.564	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.3368849

$	12-Dec-2018	19:06:58.036	Ascend	Optode Message																												4330,1532,264.870,22.432,29.175,29.105,36.145,7.039,537.9,466.6,112.6  								

$	12-Dec-2018	19:06:58.039	Ascend	Optode Message																												#  								

$	12-Dec-2018	19:06:58.798	Ascend	FLBB Message																											/99/99,99:99:99,695,2303,700,4130,532  									

$	12-Dec-2018	19:06:59.005	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019835	-7030296978499954200000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.	242424033741377730000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.00	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:06:59.005	Ascend	SBE41 Message						   11.64, 22.0631, 34.1632  	11.640000	12.453220	-0.003698	-0.010000	0.000000	0.038299	1.000000																							

$	12-Dec-2018	19:06:59.564	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.640000	151.637338	28.325000	227.500000																																

$	12-Dec-2018	19:06:59.675	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.1044959

$	12-Dec-2018	19:06:59.982	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.640000	151.620805	28.325000	227.500000																																

$	12-Dec-2018	19:06:59.908	Ascend	Vel PID Values																		0.050000	0.070001	0.069735	0.079668	203878612376498680000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-7030296978499954200000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:06:59.908	Ascend	SBE41 Message						   11.57, 22.0627, 34.1637  	11.570000	12.063514	0.063165	0.070001	0.030000	0.040798	1.000000																							

$	12-Dec-2018	19:07:00.923	Ascend	Vel PID Values																		0.050000	0.120000	0.119833	0.050099	-59124797589184615000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	203878612376498680000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:00.923	Ascend	SBE41 Message						   11.45, 22.0638, 34.1632  	11.450000	11.789573	0.125106	0.120000	0.095000	0.044199	1.000000																							

$	12-Dec-2018	19:07:01.613	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.450000	151.599340	28.325000	227.500000																																

$	12-Dec-2018	19:07:01.725	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0419268

$	12-Dec-2018	19:07:02.030	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.450000	151.434913	28.325000	227.500000																																

$	12-Dec-2018	19:07:02.731	Ascend	MSC: RX: ACK,TS,DAT 

$	12-Dec-2018	19:07:01.906	Ascend	Vel PID Values																		0.050000	-0.730000	-0.727172	-0.847005	171461913008635380000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-59124797589184615000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:01.906	Ascend	SBE41 Message						   12.18, 22.0639, 34.1631  	12.180000	11.666066	0.165146	-0.730000	-0.305000	0.047683	1.000000																							

$	12-Dec-2018	19:07:02.905	Ascend	Vel PID Values																		0.050000	-0.259999	-0.261559	0.465614	-49723954772504260000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	171461913008635380000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:02.905	Ascend	SBE41 Message						   12.44, 22.0641, 34.1630  	12.440000	11.627745	0.182882	-0.259999	-0.495000	0.050909	1.000000																							

$	12-Dec-2018	19:07:03.743	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	12.440000	151.250608	28.325000	227.500000																																

$	12-Dec-2018	19:07:03.854	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.1245952

$	12-Dec-2018	19:07:04.161	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	12.440000	150.962345	28.325000	227.500000																																

$	12-Dec-2018	19:07:04.860	Ascend	MSC: OUT: SL

$	12-Dec-2018	19:07:03.958	Ascend	Vel PID Values																		0.050000	0.009999	0.009098	0.270657	144199468840262370000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-49723954772504260000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:03.958	Ascend	SBE41 Message						   12.43, 22.0647, 34.1628  	12.430000	11.619088	0.185698	0.009999	-0.125000	0.053749	1.000000																							

$	12-Dec-2018	19:07:04.963	Ascend	Vel PID Values																		0.050000	0.000000	0.000032	-0.009066	-41817845963676086000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	144199468840262370000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:04.963	Ascend	SBE41 Message						   12.43, 22.0650, 34.1630  	12.430000	11.659769	0.170708	0.000000	0.005000	0.056059	1.000000																							

$	12-Dec-2018	19:07:05.668	Ascend	MSC: RX: 0x49A9,pH,01/31/2012 01:32:09,0,-1.00,-1.0000,-1.0000,253,145,0,22.97,56.29,11.744,4.640e-02,7.1234,0.000,-1.015421,0.003736,-1.

$	12-Dec-2018	19:07:05.674	Ascend	MSC: Data: 0x49A9,pH,01/31/2012 01:32:09,0,-1.00,-1.0000,-1.0000,253,145,0,22.97,56.29,11.744,4.640e-02,7.1234,0.000,-1.015421,0.003736,-1.

$	12-Dec-2018	19:07:06.080	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	12.430000	150.868498	28.325000	227.500000																																

$	12-Dec-2018	19:07:06.192	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.3319009

$	12-Dec-2018	19:07:06.295	Ascend	MSC: Done Sampling.

$	12-Dec-2018	19:07:06.299	Ascend	MSC: OUT: SLP

$	12-Dec-2018	19:07:05.903	Ascend	Vel PID Values																		0.050000	-0.009999	-0.009966	-0.009998	121271753294660650000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-41817845963676086000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:05.903	Ascend	SBE41 Message						   12.44, 22.0651, 34.1630  	12.440000	11.800922	0.125723	-0.009999	-0.005000	0.057490	1.000000																							

$	12-Dec-2018	19:07:07.056	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	12.440000	150.425732	28.325000	227.500000																																

$	12-Dec-2018	19:07:07.198	Ascend	MSC: RX: ACK,SLP 

$	12-Dec-2018	19:07:06.900	Ascend	Vel PID Values																		0.050000	0.009999	0.009933	0.019899	-35168808455451586000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	121271753294660650000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:06.900	Ascend	SBE41 Message						   12.43, 22.0656, 34.1630  	12.430000	12.028208	0.052559	0.009999	0.000000	0.057757	1.000000																							

$	12-Dec-2018	19:07:07.854	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	12.430000	150.265791	28.325000	227.500000																																

$	12-Dec-2018	19:07:07.966	Ascend	Vel PID Values																		0.050000	0.440001	0.438569	0.428637	101989544520809600000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-35168808455451586000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:07.966	Ascend	SBE41 Message						   11.99, 22.0659, 34.1606  	11.990000	12.246214	-0.022607	0.440001	0.225000	0.056940	1.000000																							

$	12-Dec-2018	19:07:08.754	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.990000	150.235675	28.325000	227.500000																																

$	12-Dec-2018	19:07:08.899	Ascend	Vel PID Values																		0.050000	1.080000	1.077868	0.639298	-29576967911034782000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	101989544520809600000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:08.899	Ascend	SBE41 Message						   10.91, 22.0663, 34.1634  	10.910000	12.375059	-0.072434	1.080000	0.760000	0.055453	1.000000																							

$	12-Dec-2018	19:07:09.752	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.910000	150.212020	28.325000	227.500000																																

$	12-Dec-2018	19:07:09.898	Ascend	Vel PID Values																		0.050000	0.059999	0.063390	-1.014478	857732069420008780000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-29576967911034782000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:09.898	Ascend	SBE41 Message						   10.85, 22.0661, 34.1638  	10.850000	12.406433	-0.091607	0.059999	0.570000	0.053643	1.000000																							

$	12-Dec-2018	19:07:10.752	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.850000	150.212236	28.325000	227.500000																																

$	12-Dec-2018	19:07:10.897	Ascend	Vel PID Values																		0.050000	0.000000	0.000205	-0.063185	-24874230013180256000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	857732069420008780000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:10.897	Ascend	SBE41 Message						   10.85, 22.0666, 34.1637  	10.850000	12.337745	-0.087863	0.000000	0.030000	0.051882	1.000000																							

$	12-Dec-2018	19:07:11.751	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.850000	150.227629	28.325000	227.500000																																

$	12-Dec-2018	19:07:11.896	Ascend	Vel PID Values																		0.050000	-0.009999	-0.009966	-0.010171	721352670382227440000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-24874230013180256000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:11.896	Ascend	SBE41 Message						   10.86, 22.0672, 34.1636  	10.860000	12.122841	-0.062354	-0.009999	-0.005000	0.050675	1.000000																							

$	12-Dec-2018	19:07:12.752	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.860000	150.205128	28.325000	227.500000																																

$	12-Dec-2018	19:07:12.894	Ascend	Vel PID Values																		0.050000	0.049999	0.049800	0.059765	-20919227441084599000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	721352670382227440000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:12.894	Ascend	SBE41 Message						   10.81, 22.0674, 34.1636  	10.810000	11.752460	-0.011500	0.049999	0.020000	0.050543	1.000000																							

$	12-Dec-2018	19:07:13.751	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.810000	150.210060	28.325000	227.500000																																

$	12-Dec-2018	19:07:13.893	Ascend	Vel PID Values																		0.050000	0.090000	0.089867	0.040067	606657595791453320000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-20919227441084599000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:13.893	Ascend	SBE41 Message						   10.72, 22.0672, 34.1640  	10.720000	11.338339	0.057203	0.090000	0.070000	0.051566	1.000000																							

$	12-Dec-2018	19:07:14.752	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.720000	150.197014	28.325000	227.500000																																

$	12-Dec-2018	19:07:14.892	Ascend	Vel PID Values																		0.050000	-0.690000	-0.687404	-0.777271	-17593070277952144000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	606657595791453320000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:14.892	Ascend	SBE41 Message						   11.41, 22.0683, 34.1632  	11.410000	11.034064	0.122983	-0.690000	-0.300000	0.053258	1.000000																							

$	12-Dec-2018	19:07:15.751	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.410000	150.216602	28.325000	227.500000																																

$	12-Dec-2018	19:07:15.891	Ascend	Vel PID Values																		0.050000	-0.390000	-0.390994	0.296410	510199038060612250000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-17593070277952144000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:15.891	Ascend	SBE41 Message						   11.80, 22.0681, 34.1634  	11.800000	10.889170	0.167901	-0.390000	-0.540000	0.054973	1.000000																							

$	12-Dec-2018	19:07:16.754	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.800000	150.210910	28.325000	227.500000																																

$	12-Dec-2018	19:07:16.890	Ascend	Vel PID Values																		0.050000	0.000000	-0.001300	0.389695	-14795772103757757000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	510199038060612250000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:16.890	Ascend	SBE41 Message						   11.80, 22.0683, 34.1635  	11.800000	10.853697	0.188241	0.000000	-0.195000	0.056288	1.000000																							

$	12-Dec-2018	19:07:17.751	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.800000	150.205754	28.325000	227.500000																																

$	12-Dec-2018	19:07:17.890	Ascend	Vel PID Values																		0.050000	0.020000	0.019932	0.021231	429077391008974940000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-14795772103757757000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:17.890	Ascend	SBE41 Message						   11.78, 22.0687, 34.1635  	11.780000	10.901864	0.186158	0.020000	0.010000	0.056882	1.000000																							

$	12-Dec-2018	19:07:18.750	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.780000	150.192000	28.325000	227.500000																																

$	12-Dec-2018	19:07:18.888	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009901	-0.029832	-12443244339260274000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	429077391008974940000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:18.888	Ascend	SBE41 Message						   11.79, 22.0688, 34.1636  	11.790000	11.061144	0.157937	-0.010000	0.005000	0.056352	1.000000																							

$	12-Dec-2018	19:07:19.749	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.790000	150.180124	28.325000	227.500000																																

$	12-Dec-2018	19:07:19.887	Ascend	Vel PID Values																		0.050000	-0.010000	-0.010000	-0.000100	360854085838547990000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-12443244339260274000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:19.887	Ascend	SBE41 Message						   11.80, 22.0693, 34.1634  	11.800000	11.315388	0.099880	-0.010000	-0.010000	0.054183	1.000000																							

$	12-Dec-2018	19:07:20.750	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.800000	150.197178	28.325000	227.500000																																

$	12-Dec-2018	19:07:20.886	Ascend	Vel PID Values																		0.050000	0.500000	0.498303	0.508303	-10464768489317891000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	360854085838547990000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:20.886	Ascend	SBE41 Message						   11.30, 22.0699, 34.1633  	11.300000	11.564850	0.023189	0.500000	0.245000	0.050842	1.000000																							

$	12-Dec-2018	19:07:21.751	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.300000	150.213309	28.325000	227.500000																																

$	12-Dec-2018	19:07:21.886	Ascend	Vel PID Values																		0.050000	1.010000	1.008300	0.509997	303478286190218870000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-10464768489317891000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:21.886	Ascend	SBE41 Message						   10.29, 22.0701, 34.1638  	10.290000	11.717908	-0.046981	1.010000	0.755000	0.046896	1.000000																							

$	12-Dec-2018	19:07:22.751	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.290000	150.202349	28.325000	227.500000																																

$	12-Dec-2018	19:07:22.884	Ascend	Vel PID Values																		0.050000	0.050000	0.053192	-0.955108	-88008702995163466000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	303478286190218870000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:22.884	Ascend	SBE41 Message						   10.24, 22.0698, 34.1641  	10.240000	11.759421	-0.089763	0.050000	0.530000	0.043208	1.000000																							

$	12-Dec-2018	19:07:23.750	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.240000	150.207236	28.325000	227.500000																																

$	12-Dec-2018	19:07:23.883	Ascend	Vel PID Values																		0.050000	0.000000	0.000172	-0.053020	255225238685974040000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-88008702995163466000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:23.883	Ascend	SBE41 Message						   10.24, 22.0706, 34.1639  	10.240000	11.690932	-0.099220	0.000000	0.025000	0.040690	1.000000																							

$	12-Dec-2018	19:07:24.751	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.240000	150.205292	28.325000	227.500000																																

$	12-Dec-2018	19:07:24.883	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.010138	-74015319218932470000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	255225238685974040000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:24.883	Ascend	SBE41 Message						   10.25, 22.0709, 34.1636  	10.250000	11.474618	-0.076851	-0.010000	-0.005000	0.040011	1.000000																							

$	12-Dec-2018	19:07:25.759	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.250000	150.234967	28.325000	227.500000																																

$	12-Dec-2018	19:07:25.890	Ascend	Vel PID Values																		0.050000	0.050000	0.049801	0.059767	214644425734904180000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-74015319218932470000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:25.890	Ascend	SBE41 Message						   10.20, 22.0712, 34.1639  	10.200000	11.107928	-0.025524	0.050000	0.020000	0.041327	1.000000																							

$	12-Dec-2018	19:07:26.750	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.200000	150.242224	28.325000	227.500000																																

$	12-Dec-2018	19:07:26.878	Ascend	Vel PID Values																		0.050000	0.080000	0.079900	0.030099	-62246883463122205000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	214644425734904180000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:26.878	Ascend	SBE41 Message						   10.12, 22.0719, 34.1637  	10.120000	10.704408	0.044113	0.080000	0.065000	0.044268	1.000000																							

$	12-Dec-2018	19:07:27.751	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.120000	150.256313	28.325000	227.500000																																

$	12-Dec-2018	19:07:27.878	Ascend	Vel PID Values																		0.050000	0.040000	0.040133	-0.039767	180515962043054410000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-62246883463122205000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:27.878	Ascend	SBE41 Message						   10.08, 22.0721, 34.1636  	10.080000	10.412624	0.111684	0.040000	0.060000	0.048054	1.000000																							

$	12-Dec-2018	19:07:28.750	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.080000	150.323696	28.325000	227.500000																																

$	12-Dec-2018	19:07:28.878	Ascend	Vel PID Values																		0.050000	-0.970000	-0.966639	-1.006772	-52349628992485776000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	180515962043054410000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:28.878	Ascend	SBE41 Message						   11.05, 22.0713, 34.1639  	11.050000	10.273942	0.159723	-0.970000	-0.465000	0.052025	1.000000																							

$	12-Dec-2018	19:07:29.750	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.050000	150.337406	28.325000	227.500000																																

$	12-Dec-2018	19:07:29.876	Ascend	Vel PID Values																		0.050000	-0.010000	-0.013189	0.953450	151813924078208740000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-52349628992485776000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:29.876	Ascend	SBE41 Message						   11.06, 22.0727, 34.1631  	11.060000	10.223544	0.184900	-0.010000	-0.490000	0.055834	1.000000																							

$	12-Dec-2018	19:07:30.750	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.060000	150.403254	28.325000	227.500000																																

$	12-Dec-2018	19:07:30.876	Ascend	Vel PID Values																		0.050000	0.010000	0.009928	0.023118	-44026037982680540000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	151813924078208740000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:30.876	Ascend	SBE41 Message						   11.05, 22.0726, 34.1636  	11.050000	10.212700	0.191364	0.010000	0.000000	0.059075	1.000000																							

$	12-Dec-2018	19:07:31.750	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.050000	150.407307	28.325000	227.500000																																

$	12-Dec-2018	19:07:31.876	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019862	127675510149773560000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-44026037982680540000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:31.876	Ascend	SBE41 Message						   11.06, 22.0728, 34.1636  	11.060000	10.275248	0.175129	-0.010000	0.000000	0.061093	1.000000																							

$	12-Dec-2018	19:07:32.752	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.060000	150.408603	28.325000	227.500000																																

$	12-Dec-2018	19:07:32.874	Ascend	Vel PID Values																		0.050000	-0.009999	-0.009999	-0.000065	-37025897943434328000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	127675510149773560000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:32.874	Ascend	SBE41 Message						   11.07, 22.0725, 34.1642  	11.070000	10.457201	0.125845	-0.009999	-0.010000	0.061157	1.000000																							

$	12-Dec-2018	19:07:33.755	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.070000	150.433518	28.325000	227.500000																																

$	12-Dec-2018	19:07:33.873	Ascend	Vel PID Values																		0.050000	0.009999	0.009933	0.019932	107375104035959550000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-37025897943434328000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:33.873	Ascend	SBE41 Message						   11.06, 22.0735, 34.1632  	11.060000	10.711596	0.049014	0.009999	0.000000	0.059075	1.000000																							

$	12-Dec-2018	19:07:34.753	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	11.060000	150.521815	28.325000	227.500000																																

$	12-Dec-2018	19:07:34.870	Ascend	Vel PID Values																		0.050000	0.900001	0.897039	0.887106	-31138780170428270000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	107375104035959550000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:34.870	Ascend	SBE41 Message						   10.16, 22.0737, 34.1640  	10.160000	10.921960	-0.026922	0.900001	0.455000	0.055385	1.000000																							

$	12-Dec-2018	19:07:35.753	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	10.160000	150.547929	28.325000	227.500000																																

$	12-Dec-2018	19:07:35.869	Ascend	Vel PID Values																		0.050000	0.610000	0.610960	-0.286079	903024624942419810000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-31138780170428270000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:35.869	Ascend	SBE41 Message						    9.55, 22.0734, 34.1646  	9.550000	11.024119	-0.075660	0.610000	0.755000	0.050855	1.000000																							

$	12-Dec-2018	19:07:36.752	Ascend	RecMode true, PID On, BE Ret'g, Valve Open, depthNum = 21  Pres, BP, BatV, BatI: 	9.550000	150.569111	28.325000	227.500000																																

$	12-Dec-2018	19:07:37.203	Ascend	ERROR: Missed NO3 message 29

$	12-Dec-2018	19:07:37.307	Ascend	Sending Optode Do Sample

$	12-Dec-2018	19:07:37.311	Ascend	Sending Optode Wakeup

$	12-Dec-2018	19:07:37.913	Ascend	Sending FLBB run command

$	12-Dec-2018	19:07:37.930	Ascend	Sending radiometer pollSample

$	12-Dec-2018	19:07:36.868	Ascend	Vel PID Values																		0.050000	0.010000	0.011999	-0.598961	-Infinity	903024624942419810000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000	50.000000	409600.000000	1.000000										

$	12-Dec-2018	19:07:36.868	Ascend	SBE41 Message						    9.54, 22.0746, 34.1639  	9.540000	11.026975	-0.093182	0.010000	0.310000	0.046227	1.000000																							

$	12-Dec-2018	19:07:37.867	Ascend	Vel PID Values																		0.050000	0.000000	0.000037	-0.011962	NaN	-Infinity	-50.000000	-409600.000000	1.000000										

$	12-Dec-2018	19:07:37.867	Ascend	SBE41 Message						    9.54, 22.0748, 34.1640  	9.540000	10.918353	-0.085990	0.000000	0.005000	0.042472	1.000000																							

$	12-Dec-2018	19:07:38.729	Ascend	Radiometer Message																																	SATFI40287,-0.0047,0.0012,-0.0076,-0.2037 			

$	12-Dec-2018	19:07:38.941	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000037	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:38.941	Ascend	SBE41 Message						    9.54, 22.0753, 34.1639  	9.540000	10.653440	-0.054560	0.000000	0.000000	0.040719	1.000000																							

$	12-Dec-2018	19:07:39.490	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	9.540000	150.574498	28.325000	227.500000																																

$	12-Dec-2018	19:07:39.602	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.4058938

$	12-Dec-2018	19:07:39.705	Ascend	Starting ISUS sample

$	12-Dec-2018	19:07:39.808	Ascend	Starting MSC Sequence CTDValues (time, press, temp, salinity) 0, 0.00, 0.000, 0.00

$	12-Dec-2018	19:07:39.813	Ascend	MSC: OUT: W

$	12-Dec-2018	19:07:40.264	Ascend	MSC: OUT: W

$	12-Dec-2018	19:07:39.003	Ascend	Optode Message																												4330,1532,264.900,22.450,29.166,29.096,36.138,7.041,537.8,466.5,112.0  								

$	12-Dec-2018	19:07:39.006	Ascend	Optode Message																												#  								

$	12-Dec-2018	19:07:40.497	Ascend	FLBB Message																											/99/99,99:99:99,695,2314,700,4130,532  									

$	12-Dec-2018	19:07:39.912	Ascend	Vel PID Values																		0.050000	0.050000	0.049834	0.049834	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:39.912	Ascend	SBE41 Message						    9.49, 22.0755, 34.1640  	9.490000	10.264037	0.001358	0.050000	0.025000	0.041574	1.000000																							

$	12-Dec-2018	19:07:40.952	Ascend	MSC: RX: 

$	12-Dec-2018	19:07:41.151	Ascend	MSC: RX: 

$	12-Dec-2018	19:07:41.155	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:07:41.256	Ascend	MSC: Got ACK

$	12-Dec-2018	19:07:41.657	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	9.490000	151.101537	28.325000	227.500000																																

$	12-Dec-2018	19:07:41.768	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.1607930

$	12-Dec-2018	19:07:40.933	Ascend	Vel PID Values																		0.050000	0.090000	0.089867	0.040033	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:40.933	Ascend	SBE41 Message						    9.40, 22.0757, 34.1641  	9.400000	9.890393	0.070107	0.090000	0.070000	0.044512	1.000000																							

$	12-Dec-2018	19:07:41.873	Ascend	Vel PID Values																		0.050000	0.059999	0.060099	-0.029768	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:41.873	Ascend	SBE41 Message						    9.34, 22.0757, 34.1643  	9.340000	9.654085	0.130834	0.059999	0.075000	0.048442	1.000000																							

$	12-Dec-2018	19:07:42.761	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	9.340000	151.106238	28.325000	227.500000																																

$	12-Dec-2018	19:07:42.873	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0977774

$	12-Dec-2018	19:07:42.982	Ascend	MSC: OUT: CTD,0,0.0000,0.0000,0.0000

$	12-Dec-2018	19:07:42.871	Ascend	Vel PID Values																		0.050000	-0.930000	-0.926706	-0.986805	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:42.871	Ascend	SBE41 Message						   10.27, 22.0759, 34.1638  	10.270000	9.553175	0.169359	-0.930000	-0.435000	0.052511	1.000000																							

$	12-Dec-2018	19:07:43.638	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	10.270000	151.108362	28.325000	227.500000																																

$	12-Dec-2018	19:07:43.884	Ascend	MSC: RX: NAK,CTD,EConvertErr: 0 

$	12-Dec-2018	19:07:43.986	Ascend	MSC: EConvertErr: 0

$	12-Dec-2018	19:07:43.863	Ascend	Vel PID Values																		0.050000	-0.020000	-0.023022	0.903683	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:43.863	Ascend	SBE41 Message						   10.29, 22.0761, 34.1638  	10.290000	9.512596	0.187019	-0.020000	-0.475000	0.056357	1.000000																							

$	12-Dec-2018	19:07:44.633	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	10.290000	151.102118	28.325000	227.500000																																

$	12-Dec-2018	19:07:44.862	Ascend	Vel PID Values																		0.050000	0.000000	-0.000072	0.022951	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:44.862	Ascend	SBE41 Message						   10.29, 22.0767, 34.1639  	10.290000	9.496379	0.189359	0.000000	-0.010000	0.059706	1.000000																							

$	12-Dec-2018	19:07:45.432	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	10.290000	151.134506	28.325000	227.500000																																

$	12-Dec-2018	19:07:45.883	Ascend	MSC: OUT: W

$	12-Dec-2018	19:07:46.138	Ascend	MSC: OUT: W

$	12-Dec-2018	19:07:45.861	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009967	-0.009895	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:45.861	Ascend	SBE41 Message						   10.30, 22.0769, 34.1638  	10.300000	9.546776	0.169661	-0.010000	-0.005000	0.061939	1.000000																							

$	12-Dec-2018	19:07:46.583	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:07:46.587	Ascend	MSC: Got ACK

$	12-Dec-2018	19:07:46.989	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	10.300000	151.115701	28.325000	227.500000																																

$	12-Dec-2018	19:07:47.100	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.2180600

$	12-Dec-2018	19:07:47.206	Ascend	MSC: OUT: MODE,1

$	12-Dec-2018	19:07:47.310	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:07:47.411	Ascend	MSC: Got ACK

$	12-Dec-2018	19:07:46.860	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019901	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:46.860	Ascend	SBE41 Message						   10.29, 22.0770, 34.1641  	10.290000	9.713106	0.116303	0.010000	0.000000	0.062267	1.000000																							

$	12-Dec-2018	19:07:47.855	Ascend	MSC: RX: ACK,MODE,1 

$	12-Dec-2018	19:07:47.959	Ascend	Vel PID Values																		0.050000	0.010000	0.010000	0.000066	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:47.959	Ascend	SBE41 Message						   10.28, 22.0776, 34.1637  	10.280000	9.953988	0.038342	0.010000	0.010000	0.060398	1.000000																							

$	12-Dec-2018	19:07:48.502	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	10.280000	151.116394	28.325000	227.500000																																

$	12-Dec-2018	19:07:48.615	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.5073685

$	12-Dec-2018	19:07:48.921	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	10.280000	151.104517	28.325000	227.500000																																

$	12-Dec-2018	19:07:49.621	Ascend	MSC: OUT: ACK

$	12-Dec-2018	19:07:48.856	Ascend	Vel PID Values																		0.050000	0.860000	0.857171	0.847171	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:48.856	Ascend	SBE41 Message						    9.42, 22.0777, 34.1642  	9.420000	10.157515	-0.033047	0.860000	0.435000	0.056903	1.000000																							

$	12-Dec-2018	19:07:49.953	Ascend	Vel PID Values																		0.050000	0.650000	0.650694	-0.206477	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:49.953	Ascend	SBE41 Message						    8.77, 22.0780, 34.1644  	8.770000	10.257800	-0.074692	0.650000	0.755000	0.052476	1.000000																							

$	12-Dec-2018	19:07:50.610	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	8.770000	151.102915	28.325000	227.500000																																

$	12-Dec-2018	19:07:50.722	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.1021925

$	12-Dec-2018	19:07:51.028	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	8.770000	151.124030	28.325000	227.500000																																

$	12-Dec-2018	19:07:51.728	Ascend	MSC: OUT: TS

$	12-Dec-2018	19:07:50.855	Ascend	Vel PID Values																		0.050000	-0.009999	-0.007802	-0.658496	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:50.855	Ascend	SBE41 Message						    8.78, 22.0777, 34.1647  	8.780000	10.259330	-0.087381	-0.009999	0.320000	0.047936	1.000000																							

$	12-Dec-2018	19:07:51.854	Ascend	Vel PID Values																		0.050000	0.000000	-0.000030	0.007772	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:51.854	Ascend	SBE41 Message						    8.78, 22.0780, 34.1648  	8.780000	10.149550	-0.078832	0.000000	-0.005000	0.044232	1.000000																							

$	12-Dec-2018	19:07:52.518	Ascend	MSC: RX: ACK,TS,CMD 

$	12-Dec-2018	19:07:52.822	Ascend	PTH Message														796.776367	23.675000	54.222626	24.654226																			

$	12-Dec-2018	19:07:53.026	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	8.780000	151.129879	29.500000	195.000000																																

$	12-Dec-2018	19:07:53.139	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.4109777

$	12-Dec-2018	19:07:52.917	Ascend	Vel PID Values																		0.050000	0.009999	0.009966	0.009996	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:52.917	Ascend	SBE41 Message						    8.77, 22.0787, 34.1644  	8.770000	9.885668	-0.048197	0.009999	0.005000	0.042534	1.000000																							

$	12-Dec-2018	19:07:53.785	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	8.770000	151.108935	29.500000	195.000000																																

$	12-Dec-2018	19:07:53.896	Ascend	Vel PID Values																		0.050000	0.050000	0.049867	0.039901	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:53.896	Ascend	SBE41 Message						    8.72, 22.0785, 34.1646  	8.720000	9.498598	0.005930	0.050000	0.030000	0.043500	1.000000																							

$	12-Dec-2018	19:07:54.688	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	8.720000	151.120402	29.500000	195.000000																																

$	12-Dec-2018	19:07:54.850	Ascend	Vel PID Values																		0.050000	0.090000	0.089867	0.040000	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:54.850	Ascend	SBE41 Message						    8.63, 22.0790, 34.1647  	8.630000	9.126212	0.072591	0.090000	0.070000	0.046656	1.000000																							

$	12-Dec-2018	19:07:55.687	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	8.630000	151.120566	29.500000	195.000000																																

$	12-Dec-2018	19:07:55.849	Ascend	Vel PID Values																		0.050000	-0.220000	-0.218969	-0.308836	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:55.849	Ascend	SBE41 Message						    8.85, 22.0796, 34.1640  	8.850000	8.889789	0.131660	-0.220000	-0.065000	0.050877	1.000000																							

$	12-Dec-2018	19:07:56.688	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	8.850000	151.144311	29.500000	195.000000																																

$	12-Dec-2018	19:07:56.847	Ascend	Vel PID Values																		0.050000	-0.740000	-0.738268	-0.519299	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:56.847	Ascend	SBE41 Message						    9.59, 22.0795, 34.1643  	9.590000	8.788768	0.169354	-0.740000	-0.480000	0.055230	1.000000																							

$	12-Dec-2018	19:07:57.688	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	9.590000	151.117019	29.500000	195.000000																																

$	12-Dec-2018	19:07:57.847	Ascend	Vel PID Values																		0.050000	-0.010000	-0.012427	0.725841	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:57.847	Ascend	SBE41 Message						    9.60, 22.0796, 34.1643  	9.600000	8.753847	0.186083	-0.010000	-0.375000	0.059140	1.000000																							

$	12-Dec-2018	19:07:58.688	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	9.600000	151.121669	29.500000	195.000000																																

$	12-Dec-2018	19:07:59.159	Ascend	MSC: RX: ACK,TS,DAT 

$	12-Dec-2018	19:07:58.846	Ascend	Vel PID Values																		0.050000	0.000000	-0.000037	0.012389	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:58.846	Ascend	SBE41 Message						    9.60, 22.0801, 34.1641  	9.600000	8.760427	0.185441	0.000000	-0.005000	0.062011	1.000000																							

$	12-Dec-2018	19:07:59.805	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	9.600000	151.123628	29.500000	195.000000																																

$	12-Dec-2018	19:07:59.916	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	0.000037	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:07:59.916	Ascend	SBE41 Message						    9.60, 22.0800, 34.1645  	9.600000	8.851313	0.160593	0.000000	0.000000	0.063127	1.000000																							

$	12-Dec-2018	19:08:00.702	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	9.600000	151.130274	29.500000	195.000000																																

$	12-Dec-2018	19:08:01.159	Ascend	MSC: OUT: SL

$	12-Dec-2018	19:08:00.844	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	0.000000	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:00.844	Ascend	SBE41 Message						    9.60, 22.0808, 34.1640  	9.600000	9.052560	0.103043	0.000000	0.000000	0.062101	1.000000																							

$	12-Dec-2018	19:08:01.833	Ascend	Error in I2CPortBase.Execute

$	12-Dec-2018	19:08:01.838	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	9.600000	151.130274	29.500000	195.000000																																

$	12-Dec-2018	19:08:02.694	Ascend	Error in I2CPortBase.Execute

$	12-Dec-2018	19:08:02.796	Ascend	2485 Sample out of range

$	12-Dec-2018	19:08:02.800	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	9.600000	250.000000	29.500000	195.000000																																

$	12-Dec-2018	19:08:03.160	Ascend	MSC: RX: 0x50E3,A,01/31/2012 01:33:04,0,-1.00,-1.0000,-1.0000,226,145,1,65,23.07,56.18,11.576,5.248e-01,10549.00,2227.00,1184.45,11.95,-2B3E444051436147584C73522C58925FAA666B6CF77261763D784778E677D67537717D6D846960659262095F9A5E005D4C5D415E76602A62D6661969F16EA573BE79497F5C852C8A9B8FBC93E696B898199802961892CA8E3A88BD82487BCD753F6F3E698E64AE60005C0958AA55EB5373516C4FE54EDE4E0F4DB44DBA,6657.6 

$	12-Dec-2018	19:08:03.171	Ascend	MSC: Data: 0x50E3,A,01/31/2012 01:33:04,0,-1.00,-1.0000,-1.0000,226,145,1,65,23.07,56.18,11.576,5.248e-01,10549.00,2227.00,1184.45,11.95,-2B3E444051436147584C73522C58925FAA666B6CF77261763D784778E677D67537717D6D846960659262095F9A5E005D4C5D415E76602A62D6661969F16EA573BE79497F5C852C8A9B8FBC93E696B898199802961892CA8E3A88BD82487BCD753F6F3E698E64AE60005C0958AA55EB5373516C4FE54EDE4E0F4DB44DBA,6657.6 

$	12-Dec-2018	19:08:02.362	Ascend	Vel PID Values																		0.050000	0.040000	0.039867	0.039867	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:02.362	Ascend	SBE41 Message						    9.56, 22.0806, 34.1644  	9.560000	9.299820	0.024577	0.040000	0.020000	0.058960	1.000000																							

$	12-Dec-2018	19:08:03.723	Ascend	MSC: RX: 99999999999999999999999999999999999999999999999999999999

$	12-Dec-2018	19:08:03.427	Ascend	Vel PID Values																		0.050000	1.090000	1.086506	1.046639	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:03.427	Ascend	SBE41 Message						    8.47, 22.0808, 34.1648  	8.470000	9.487474	-0.044698	1.090000	0.565000	0.054513	1.000000																							

$	12-Dec-2018	19:08:03.843	Ascend	Vel PID Values																		0.050000	0.380000	0.382357	-0.704149	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:03.843	Ascend	SBE41 Message						    8.09, 22.0808, 34.1651  	8.090000	9.570439	-0.084216	0.380000	0.735000	0.049800	1.000000																							

$	12-Dec-2018	19:08:04.712	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	8.090000	151.122354	29.500000	195.000000																																

$	12-Dec-2018	19:08:04.825	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.6667312

$	12-Dec-2018	19:08:04.929	Ascend	MSC: Done Sampling.

$	12-Dec-2018	19:08:04.933	Ascend	Starting pH sample

$	12-Dec-2018	19:08:05.035	Ascend	MSC: OUT: W

$	12-Dec-2018	19:08:05.484	Ascend	MSC: OUT: W

$	12-Dec-2018	19:08:04.928	Ascend	Vel PID Values																		0.050000	-0.010000	-0.008698	-0.391055	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:04.928	Ascend	SBE41 Message						    8.10, 22.0817, 34.1646  	8.100000	9.556893	-0.095050	-0.010000	0.185000	0.045540	1.000000																							

$	12-Dec-2018	19:08:05.845	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:08:05.947	Ascend	MSC: Got ACK

$	12-Dec-2018	19:08:06.146	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:08:06.150	Ascend	MSC: Got ACK

$	12-Dec-2018	19:08:06.551	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	8.100000	151.126370	29.500000	195.000000																																

$	12-Dec-2018	19:08:06.664	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.8329323

$	12-Dec-2018	19:08:06.767	Ascend	MSC: OUT: MODE,2

$	12-Dec-2018	19:08:05.943	Ascend	Vel PID Values																		0.050000	0.000000	-0.000031	0.008667	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:05.943	Ascend	SBE41 Message						    8.10, 22.0812, 34.1652  	8.100000	9.423209	-0.082841	0.000000	-0.005000	0.042809	1.000000																							

$	12-Dec-2018	19:08:06.869	Ascend	Vel PID Values																		0.050000	0.020000	0.019934	0.019965	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:06.869	Ascend	SBE41 Message						    8.08, 22.0815, 34.1651  	8.080000	9.129905	-0.045743	0.020000	0.010000	0.042391	1.000000																							

$	12-Dec-2018	19:08:07.557	Ascend	MSC: RX: ACK,MODE,2 

$	12-Dec-2018	19:08:07.864	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	8.080000	151.131563	29.500000	195.000000																																

$	12-Dec-2018	19:08:07.975	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.3055684

$	12-Dec-2018	19:08:07.837	Ascend	Vel PID Values																		0.050000	0.050000	0.049900	0.029966	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:07.837	Ascend	SBE41 Message						    8.03, 22.0822, 34.1649  	8.030000	8.732393	0.014685	0.050000	0.035000	0.044244	1.000000																							

$	12-Dec-2018	19:08:08.623	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	8.030000	151.124224	29.500000	195.000000																																

$	12-Dec-2018	19:08:08.981	Ascend	MSC: OUT: ACK

$	12-Dec-2018	19:08:08.837	Ascend	Vel PID Values																		0.050000	0.080000	0.079900	0.030000	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:08.837	Ascend	SBE41 Message						    7.95, 22.0826, 34.1647  	7.950000	8.381393	0.083021	0.080000	0.065000	0.047591	1.000000																							

$	12-Dec-2018	19:08:09.629	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 22  Pres, BP, BatV, BatI: 	7.950000	151.133642	29.500000	195.000000																																

$	12-Dec-2018	19:08:09.986	Ascend	ERROR: Missed PH message 33

$	12-Dec-2018	19:08:10.090	Ascend	Sending Optode Do Sample

$	12-Dec-2018	19:08:10.191	Ascend	Sending Optode Wakeup

$	12-Dec-2018	19:08:10.698	Ascend	Sending FLBB run command

$	12-Dec-2018	19:08:10.715	Ascend	Sending radiometer pollSample

$	12-Dec-2018	19:08:09.835	Ascend	Vel PID Values																		0.050000	-0.320001	-0.318670	-0.398570	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:09.835	Ascend	SBE41 Message						    8.27, 22.0827, 34.1648  	8.270000	8.178400	0.138744	-0.320001	-0.120000	0.051521	1.000000																							

$	12-Dec-2018	19:08:10.919	Ascend	Vel PID Values																		0.050000	-0.570000	-0.569166	-0.250496	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:10.919	Ascend	SBE41 Message						    8.84, 22.0832, 34.1644  	8.840000	8.098047	0.171915	-0.570000	-0.445000	0.055397	1.000000																							

$	12-Dec-2018	19:08:11.607	Ascend	Radiometer Message																																	SATFI40287,-0.0041,0.0007,-0.0077,-0.1938 			

$	12-Dec-2018	19:08:11.924	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.840000	151.137196	29.500000	195.000000																																

$	12-Dec-2018	19:08:12.037	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.0568589

$	12-Dec-2018	19:08:12.140	Ascend	MSC: OUT: TS

$	12-Dec-2018	19:08:11.789	Ascend	Optode Message																												4330,1532,264.848,22.460,29.164,29.095,36.137,7.042,537.7,466.3,111.6  								

$	12-Dec-2018	19:08:11.792	Ascend	Optode Message																												#  								

$	12-Dec-2018	19:08:11.833	Ascend	Vel PID Values																		0.050000	0.010000	0.008072	0.577237	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:11.833	Ascend	SBE41 Message						    8.83, 22.0828, 34.1649  	8.830000	8.070641	0.185567	0.010000	-0.280000	0.058752	1.000000																							

$	12-Dec-2018	19:08:12.811	Ascend	FLBB Message																											/99/99,99:99:99,695,2304,700,4130,532  									

$	12-Dec-2018	19:08:12.924	Ascend	MSC: RX: ACK,TS,CMD 

$	12-Dec-2018	19:08:12.921	Ascend	Vel PID Values																		0.050000	0.000000	0.000030	-0.008041	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:12.921	Ascend	SBE41 Message						    8.83, 22.0837, 34.1644  	8.830000	8.083861	0.182235	0.000000	0.005000	0.060937	1.000000																							

$	12-Dec-2018	19:08:13.570	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.830000	151.120782	29.500000	195.000000																																

$	12-Dec-2018	19:08:13.681	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.6383475

$	12-Dec-2018	19:08:13.988	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.830000	151.129186	29.500000	195.000000																																

$	12-Dec-2018	19:08:13.831	Ascend	Vel PID Values																		0.050000	-0.020000	-0.019934	-0.019964	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:13.831	Ascend	SBE41 Message						    8.85, 22.0834, 34.1649  	8.850000	8.179915	0.153790	-0.020000	-0.010000	0.061369	1.000000																							

$	12-Dec-2018	19:08:14.916	Ascend	Vel PID Values																		0.050000	0.010000	0.009901	0.029834	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:14.916	Ascend	SBE41 Message						    8.84, 22.0838, 34.1649  	8.840000	8.371325	0.093820	0.010000	-0.005000	0.059801	1.000000																							

$	12-Dec-2018	19:08:15.575	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.840000	151.145406	29.500000	195.000000																																

$	12-Dec-2018	19:08:15.688	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0002112

$	12-Dec-2018	19:08:15.993	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.840000	151.120111	29.500000	195.000000																																

$	12-Dec-2018	19:08:15.829	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.009867	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:15.829	Ascend	SBE41 Message						    8.84, 22.0839, 34.1650  	8.840000	8.590279	0.017567	0.000000	0.005000	0.056586	1.000000																							

$	12-Dec-2018	19:08:16.923	Ascend	Vel PID Values																		0.050000	0.810000	0.807305	0.807272	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:16.923	Ascend	SBE41 Message						    8.03, 22.0844, 34.1650  	8.030000	8.747582	-0.045308	0.810000	0.405000	0.052298	1.000000																							

$	12-Dec-2018	19:08:17.582	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.030000	151.118591	29.500000	195.000000																																

$	12-Dec-2018	19:08:17.693	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0000038

$	12-Dec-2018	19:08:17.998	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.030000	151.118971	29.500000	195.000000																																

$	12-Dec-2018	19:08:17.827	Ascend	Vel PID Values																		0.050000	0.640000	0.640561	-0.166744	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:17.827	Ascend	SBE41 Message						    7.39, 22.0848, 34.1652  	7.390000	8.816088	-0.078901	0.640000	0.725000	0.047776	1.000000																							

$	12-Dec-2018	19:08:18.826	Ascend	Vel PID Values																		0.050000	0.010000	0.012097	-0.628464	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:18.826	Ascend	SBE41 Message						    7.38, 22.0854, 34.1647  	7.380000	8.810695	-0.088184	0.010000	0.325000	0.043887	1.000000																							

$	12-Dec-2018	19:08:19.685	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.380000	151.127227	29.500000	195.000000																																

$	12-Dec-2018	19:08:19.796	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0001273

$	12-Dec-2018	19:08:19.899	Ascend	Vel PID Values																		0.050000	0.000000	0.000037	-0.012060	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:19.899	Ascend	SBE41 Message						    7.38, 22.0855, 34.1651  	7.380000	8.709020	-0.079000	0.000000	0.005000	0.041504	1.000000																							

$	12-Dec-2018	19:08:20.445	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.380000	151.136547	29.500000	195.000000																																

$	12-Dec-2018	19:08:21.004	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.380000	151.127070	29.500000	195.000000																																

$	12-Dec-2018	19:08:20.824	Ascend	Vel PID Values																		0.050000	0.010000	0.009967	0.009930	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:20.824	Ascend	SBE41 Message						    7.37, 22.0856, 34.1651  	7.370000	8.460512	-0.046622	0.010000	0.005000	0.041185	1.000000																							

$	12-Dec-2018	19:08:21.823	Ascend	Vel PID Values																		0.050000	0.060000	0.059833	0.049866	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:21.823	Ascend	SBE41 Message						    7.31, 22.0859, 34.1652  	7.310000	8.090028	0.011096	0.060000	0.035000	0.042870	1.000000																							

$	12-Dec-2018	19:08:22.689	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.310000	151.128091	29.500000	195.000000																																

$	12-Dec-2018	19:08:22.801	Ascend	MSC: RX: ACK,TS,DAT 

$	12-Dec-2018	19:08:23.008	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.310000	151.139922	29.500000	195.000000																																

$	12-Dec-2018	19:08:22.904	Ascend	Vel PID Values																		0.050000	0.050000	0.050033	-0.009801	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:22.904	Ascend	SBE41 Message						    7.26, 22.0864, 34.1649  	7.260000	7.728167	0.079549	0.050000	0.055000	0.045967	1.000000																							

$	12-Dec-2018	19:08:23.821	Ascend	Vel PID Values																		0.050000	0.030000	0.030067	-0.019966	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:23.821	Ascend	SBE41 Message						    7.23, 22.0867, 34.1649  	7.230000	7.493823	0.135644	0.030000	0.040000	0.049746	1.000000																							

$	12-Dec-2018	19:08:24.691	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.230000	151.141979	29.500000	195.000000																																

$	12-Dec-2018	19:08:24.802	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0017649

$	12-Dec-2018	19:08:24.908	Ascend	MSC: OUT: SL

$	12-Dec-2018	19:08:24.908	Ascend	Vel PID Values																		0.050000	-0.860000	-0.857038	-0.887105	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:24.908	Ascend	SBE41 Message						    8.09, 22.0863, 34.1652  	8.090000	7.390196	0.167943	-0.860000	-0.415000	0.053585	1.000000																							

$	12-Dec-2018	19:08:25.554	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.090000	151.112035	29.500000	195.000000																																

$	12-Dec-2018	19:08:25.812	Ascend	MSC: RX: 0x5124,pH,01/31/2012 01:33:30,0,-1.00,-1.0000,-1.0000,254,145,0,23.09,56.10,11.744,4.640e-02,7.1234,0.000,-0.980707,0.001117,-0.969550,0.001474,+2.185e-07,+3.131e-08 

$	12-Dec-2018	19:08:25.819	Ascend	MSC: Data: 0x5124,pH,01/31/2012 01:33:30,0,-1.00,-1.0000,-1.0000,254,145,0,23.09,56.10,11.744,4.640e-02,7.1234,0.000,-0.980707,0.001117,-0.969550,0.001474,+2.185e-07,+3.131e-08 

$	12-Dec-2018	19:08:25.916	Ascend	Vel PID Values																		0.050000	0.000000	-0.002857	0.854181	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:25.916	Ascend	SBE41 Message						    8.09, 22.0870, 34.1648  	8.090000	7.348934	0.181528	0.000000	-0.430000	0.056985	1.000000																							

$	12-Dec-2018	19:08:26.567	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.090000	151.138015	29.500000	195.000000																																

$	12-Dec-2018	19:08:26.811	Ascend	MSC: Done Sampling.

$	12-Dec-2018	19:08:26.913	Ascend	MSC: OUT: SLP

$	12-Dec-2018	19:08:26.912	Ascend	Vel PID Values																		0.050000	0.010000	0.009962	0.012819	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:26.912	Ascend	SBE41 Message						    8.08, 22.0865, 34.1653  	8.080000	7.337661	0.182239	0.010000	0.005000	0.059372	1.000000																							

$	12-Dec-2018	19:08:27.559	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.080000	151.128776	29.500000	195.000000																																

$	12-Dec-2018	19:08:27.811	Ascend	MSC: RX: ACK,SLP 

$	12-Dec-2018	19:08:28.115	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.080000	151.131757	29.500000	195.000000																																

$	12-Dec-2018	19:08:27.913	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019896	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:27.913	Ascend	SBE41 Message						    8.09, 22.0875, 34.1647  	8.090000	7.394549	0.162599	-0.010000	0.000000	0.060150	1.000000																							

$	12-Dec-2018	19:08:28.828	Ascend	Vel PID Values																		0.050000	0.000000	-0.000033	0.009901	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:28.828	Ascend	SBE41 Message						    8.09, 22.0869, 34.1656  	8.090000	7.558856	0.111520	0.000000	-0.005000	0.059099	1.000000																							

$	12-Dec-2018	19:08:29.698	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.090000	151.136793	29.500000	195.000000																																

$	12-Dec-2018	19:08:29.811	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0007097

$	12-Dec-2018	19:08:29.914	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	0.000033	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:29.914	Ascend	SBE41 Message						    8.09, 22.0884, 34.1645  	8.090000	7.785274	0.037687	0.000000	0.000000	0.056306	1.000000																							

$	12-Dec-2018	19:08:30.459	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.090000	151.137121	29.500000	195.000000																																

$	12-Dec-2018	19:08:30.814	Ascend	Vel PID Values																		0.050000	0.050000	0.049834	0.049834	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:30.814	Ascend	SBE41 Message						    8.04, 22.0886, 34.1646  	8.040000	7.970084	-0.029932	0.050000	0.025000	0.052403	1.000000																							

$	12-Dec-2018	19:08:31.362	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	8.040000	151.151024	29.500000	195.000000																																

$	12-Dec-2018	19:08:31.812	Ascend	Vel PID Values																		0.050000	1.090000	1.086539	1.036705	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:31.812	Ascend	SBE41 Message						    6.95, 22.0886, 34.1649  	6.950000	8.060339	-0.070093	1.090000	0.570000	0.048157	1.000000																							

$	12-Dec-2018	19:08:32.362	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	6.950000	151.150532	29.500000	195.000000																																

$	12-Dec-2018	19:08:32.813	Ascend	Vel PID Values																		0.050000	0.320000	0.322556	-0.763983	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:32.813	Ascend	SBE41 Message						    6.63, 22.0883, 34.1658  	6.630000	8.080439	-0.085410	0.320000	0.705000	0.044453	1.000000																							

$	12-Dec-2018	19:08:33.362	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	6.630000	151.127763	29.500000	195.000000																																

$	12-Dec-2018	19:08:33.810	Ascend	Vel PID Values																		0.050000	0.000000	0.001069	-0.321487	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:33.810	Ascend	SBE41 Message						    6.63, 22.0884, 34.1657  	6.630000	8.047115	-0.086395	0.000000	0.160000	0.042025	1.000000																							

$	12-Dec-2018	19:08:34.362	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	6.630000	151.134260	29.500000	195.000000																																

$	12-Dec-2018	19:08:34.810	Ascend	Vel PID Values																		0.050000	-0.020000	-0.019932	-0.021001	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:34.810	Ascend	SBE41 Message						    6.65, 22.0889, 34.1657  	6.650000	7.911208	-0.070973	-0.020000	-0.010000	0.041266	1.000000																							

$	12-Dec-2018	19:08:35.364	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	6.650000	151.137263	29.500000	195.000000																																

$	12-Dec-2018	19:08:35.809	Ascend	Vel PID Values																		0.050000	0.050000	0.049767	0.069699	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:35.809	Ascend	SBE41 Message						    6.60, 22.0889, 34.1658  	6.600000	7.621509	-0.028095	0.050000	0.015000	0.042258	1.000000																							

$	12-Dec-2018	19:08:36.363	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	6.600000	151.126862	29.500000	195.000000																																

$	12-Dec-2018	19:08:36.808	Ascend	Vel PID Values																		0.050000	0.050000	0.049999	0.000232	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:36.808	Ascend	SBE41 Message						    6.55, 22.0898, 34.1653  	6.550000	7.234522	0.040103	0.050000	0.050000	0.044826	1.000000																							

$	12-Dec-2018	19:08:37.364	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	6.550000	151.133455	29.500000	195.000000																																

$	12-Dec-2018	19:08:37.807	Ascend	Vel PID Values																		0.050000	0.040000	0.040033	-0.009966	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:37.807	Ascend	SBE41 Message						    6.51, 22.0894, 34.1659  	6.510000	6.898338	0.109707	0.040000	0.045000	0.048475	1.000000																							

$	12-Dec-2018	19:08:38.368	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	6.510000	151.129842	29.500000	195.000000																																

$	12-Dec-2018	19:08:38.806	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009833	-0.049867	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:38.806	Ascend	SBE41 Message						    6.52, 22.0901, 34.1655  	6.520000	6.706910	0.156269	-0.010000	0.015000	0.052459	1.000000																							

$	12-Dec-2018	19:08:39.364	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	6.520000	151.146948	29.500000	195.000000																																

$	12-Dec-2018	19:08:39.805	Ascend	Vel PID Values																		0.050000	-0.800000	-0.797371	-0.787538	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:39.805	Ascend	SBE41 Message						    7.32, 22.0905, 34.1653  	7.320000	6.629894	0.176982	-0.800000	-0.405000	0.055969	1.000000																							

$	12-Dec-2018	19:08:40.370	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.320000	151.122048	29.500000	195.000000																																

$	12-Dec-2018	19:08:40.803	Ascend	Vel PID Values																		0.050000	0.010000	0.007309	0.804680	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:40.803	Ascend	SBE41 Message						    7.31, 22.0903, 34.1657  	7.310000	6.595362	0.183742	0.010000	-0.395000	0.058427	1.000000																							

$	12-Dec-2018	19:08:41.365	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.310000	151.145257	29.500000	195.000000																																

$	12-Dec-2018	19:08:41.802	Ascend	Vel PID Values																		0.050000	0.040000	0.039896	0.032587	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:41.802	Ascend	SBE41 Message						    7.27, 22.0910, 34.1651  	7.270000	6.586154	0.180738	0.040000	0.025000	0.059380	1.000000																							

$	12-Dec-2018	19:08:42.366	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.270000	151.141748	29.500000	195.000000																																

$	12-Dec-2018	19:08:42.802	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009834	-0.049730	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:42.802	Ascend	SBE41 Message						    7.28, 22.0911, 34.1654  	7.280000	6.646464	0.156439	-0.010000	0.015000	0.058878	1.000000																							

$	12-Dec-2018	19:08:43.368	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.280000	151.139840	29.500000	195.000000																																

$	12-Dec-2018	19:08:43.801	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019768	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:43.801	Ascend	SBE41 Message						    7.27, 22.0912, 34.1655  	7.270000	6.808973	0.100492	0.010000	0.000000	0.057017	1.000000																							

$	12-Dec-2018	19:08:44.367	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.270000	151.138544	29.500000	195.000000																																

$	12-Dec-2018	19:08:44.800	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019868	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:44.800	Ascend	SBE41 Message						    7.28, 22.0915, 34.1653  	7.280000	7.023232	0.026127	-0.010000	0.000000	0.053892	1.000000																							

$	12-Dec-2018	19:08:45.369	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.280000	151.143119	29.500000	195.000000																																

$	12-Dec-2018	19:08:45.798	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019868	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:45.798	Ascend	SBE41 Message						    7.27, 22.0920, 34.1652  	7.270000	7.189615	-0.036036	0.010000	0.000000	0.050196	1.000000																							

$	12-Dec-2018	19:08:46.369	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	7.270000	151.145279	29.500000	195.000000																																

$	12-Dec-2018	19:08:46.798	Ascend	Vel PID Values																		0.050000	0.640000	0.637903	0.627969	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:46.798	Ascend	SBE41 Message						    6.63, 22.0923, 34.1655  	6.630000	7.263324	-0.068842	0.640000	0.325000	0.046946	1.000000																							

$	12-Dec-2018	19:08:47.370	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	6.630000	151.164904	29.500000	195.000000																																

$	12-Dec-2018	19:08:47.797	Ascend	Vel PID Values																		0.050000	0.720000	0.719731	0.081827	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:47.797	Ascend	SBE41 Message						    5.91, 22.0925, 34.1657  	5.910000	7.275427	-0.079879	0.720000	0.680000	0.044838	1.000000																							

$	12-Dec-2018	19:08:48.371	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 23  Pres, BP, BatV, BatI: 	5.910000	151.155159	29.500000	195.000000																																

$	12-Dec-2018	19:08:48.833	Ascend	ERROR: Missed NO3 message 30

$	12-Dec-2018	19:08:48.937	Ascend	Sending Optode Do Sample

$	12-Dec-2018	19:08:48.941	Ascend	Sending Optode Wakeup

$	12-Dec-2018	19:08:49.544	Ascend	Sending FLBB run command

$	12-Dec-2018	19:08:49.561	Ascend	Sending radiometer pollSample

$	12-Dec-2018	19:08:48.796	Ascend	Vel PID Values																		0.050000	0.050000	0.052229	-0.667502	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:48.796	Ascend	SBE41 Message						    5.86, 22.0928, 34.1658  	5.860000	7.253631	-0.081455	0.050000	0.385000	0.043890	1.000000																							

$	12-Dec-2018	19:08:50.112	Ascend	Radiometer Message																																	SATFI40287,-0.0044,0.0015,-0.0080,-0.2243 			

$	12-Dec-2018	19:08:49.824	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009796	-0.062025	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:49.824	Ascend	SBE41 Message						    5.87, 22.0934, 34.1654  	5.870000	7.163023	-0.071376	-0.010000	0.020000	0.043841	1.000000																							

$	12-Dec-2018	19:08:50.774	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.870000	151.137486	29.500000	195.000000																																

$	12-Dec-2018	19:08:50.887	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.0621724

$	12-Dec-2018	19:08:50.894	Ascend	Starting ISUS sample

$	12-Dec-2018	19:08:51.094	Ascend	Starting MSC Sequence CTDValues (time, press, temp, salinity) 0, 0.00, 0.000, 0.00

$	12-Dec-2018	19:08:51.100	Ascend	MSC: OUT: W

$	12-Dec-2018	19:08:51.549	Ascend	MSC: OUT: W

$	12-Dec-2018	19:08:50.634	Ascend	Optode Message																												4330,1532,264.798,22.475,29.160,29.091,36.136,7.045,537.7,466.3,111.2  								

$	12-Dec-2018	19:08:50.638	Ascend	Optode Message																												#  								

$	12-Dec-2018	19:08:51.880	Ascend	FLBB Message																											/99/99,99:99:99,695,2311,700,4130,532  									

$	12-Dec-2018	19:08:50.884	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019730	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:50.884	Ascend	SBE41 Message						    5.86, 22.0930, 34.1659  	5.860000	6.941176	-0.036467	0.010000	0.000000	0.044692	1.000000																							

$	12-Dec-2018	19:08:52.236	Ascend	MSC: RX: 

$	12-Dec-2018	19:08:52.435	Ascend	MSC: RX: 

$	12-Dec-2018	19:08:51.880	Ascend	Vel PID Values																		0.050000	0.040000	0.039900	0.029966	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:51.880	Ascend	SBE41 Message						    5.82, 22.0932, 34.1661  	5.820000	6.595681	0.026371	0.040000	0.025000	0.046724	1.000000																							

$	12-Dec-2018	19:08:52.780	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:08:52.884	Ascend	MSC: Got ACK

$	12-Dec-2018	19:08:53.285	Ascend	PTH Message														796.776367	23.733333	54.276031	24.664951																			

$	12-Dec-2018	19:08:53.487	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.820000	151.152432	29.725000	170.500000																																

$	12-Dec-2018	19:08:53.598	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.7044507

$	12-Dec-2018	19:08:52.878	Ascend	Vel PID Values																		0.050000	0.040000	0.040000	0.000100	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:52.878	Ascend	SBE41 Message						    5.78, 22.0940, 34.1657  	5.780000	6.240778	0.096775	0.040000	0.040000	0.049707	1.000000																							

$	12-Dec-2018	19:08:53.791	Ascend	Vel PID Values																		0.050000	0.050000	0.049967	0.009967	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:53.791	Ascend	SBE41 Message						    5.73, 22.0938, 34.1660  	5.730000	5.998256	0.147371	0.050000	0.045000	0.052896	1.000000																							

$	12-Dec-2018	19:08:54.493	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.730000	151.135415	29.725000	170.500000																																

$	12-Dec-2018	19:08:54.604	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0000089

$	12-Dec-2018	19:08:54.708	Ascend	MSC: OUT: CTD,0,0.0000,0.0000,0.0000

$	12-Dec-2018	19:08:54.811	Ascend	Vel PID Values																		0.050000	-0.580000	-0.577903	-0.627870	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:54.811	Ascend	SBE41 Message						    6.31, 22.0942, 34.1658  	6.310000	5.887119	0.170261	-0.580000	-0.265000	0.055320	1.000000																							

$	12-Dec-2018	19:08:55.367	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	6.310000	151.158825	29.725000	170.500000																																

$	12-Dec-2018	19:08:55.610	Ascend	MSC: RX: NAK,CTD,EConvertErr: 0 

$	12-Dec-2018	19:08:55.714	Ascend	MSC: EConvertErr: 0

$	12-Dec-2018	19:08:56.019	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	6.310000	151.143014	29.725000	170.500000																																

$	12-Dec-2018	19:08:55.816	Ascend	Vel PID Values																		0.050000	-0.190000	-0.191294	0.386609	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:55.816	Ascend	SBE41 Message						    6.50, 22.0945, 34.1656  	6.500000	5.846163	0.176837	-0.190000	-0.385000	0.056602	1.000000																							

$	12-Dec-2018	19:08:56.857	Ascend	Vel PID Values																		0.050000	0.030000	0.029266	0.220560	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:56.857	Ascend	SBE41 Message						    6.47, 22.0945, 34.1646  	6.470000	5.832950	0.175764	0.030000	-0.080000	0.056787	1.000000																							

$	12-Dec-2018	19:08:57.516	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	6.470000	151.140094	29.725000	170.500000																																

$	12-Dec-2018	19:08:57.629	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0191755

$	12-Dec-2018	19:08:57.635	Ascend	MSC: OUT: W

$	12-Dec-2018	19:08:58.086	Ascend	MSC: OUT: W

$	12-Dec-2018	19:08:57.834	Ascend	Vel PID Values																		0.050000	0.030000	0.029999	0.000733	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:57.834	Ascend	SBE41 Message						    6.44, 22.0944, 34.1663  	6.440000	5.868144	0.160177	0.030000	0.030000	0.056050	1.000000																							

$	12-Dec-2018	19:08:58.436	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:08:58.537	Ascend	MSC: Got ACK

$	12-Dec-2018	19:08:58.737	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:08:58.741	Ascend	MSC: Got ACK

$	12-Dec-2018	19:08:58.840	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009867	-0.039865	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:58.840	Ascend	SBE41 Message						    6.45, 22.0950, 34.1660  	6.450000	5.988772	0.117796	-0.010000	0.010000	0.054335	1.000000																							

$	12-Dec-2018	19:08:59.484	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	6.450000	151.143670	29.725000	170.500000																																

$	12-Dec-2018	19:08:59.595	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.9612938

$	12-Dec-2018	19:08:59.699	Ascend	MSC: OUT: MODE,1

$	12-Dec-2018	19:08:59.801	Ascend	Vel PID Values																		0.050000	0.010000	0.009933	0.019800	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:08:59.801	Ascend	SBE41 Message						    6.44, 22.0953, 34.1658  	6.440000	6.172851	0.052289	0.010000	0.000000	0.051807	1.000000																							

$	12-Dec-2018	19:09:00.350	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	6.440000	151.132509	29.725000	170.500000																																

$	12-Dec-2018	19:09:00.602	Ascend	MSC: RX: ACK,MODE,1 

$	12-Dec-2018	19:09:00.908	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	6.440000	151.143745	29.725000	170.500000																																

$	12-Dec-2018	19:09:00.784	Ascend	Vel PID Values																		0.050000	0.010000	0.010000	0.000067	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:00.784	Ascend	SBE41 Message						    6.43, 22.0959, 34.1655  	6.430000	6.338523	-0.012100	0.010000	0.010000	0.049165	1.000000																							

$	12-Dec-2018	19:09:01.832	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009934	-0.019934	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:01.832	Ascend	SBE41 Message						    6.44, 22.0959, 34.1658  	6.440000	6.426883	-0.053033	-0.010000	0.000000	0.047216	1.000000																							

$	12-Dec-2018	19:09:02.490	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	6.440000	151.134327	29.725000	170.500000																																

$	12-Dec-2018	19:09:02.601	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.0001791

$	12-Dec-2018	19:09:02.706	Ascend	MSC: OUT: ACK

$	12-Dec-2018	19:09:02.808	Ascend	Vel PID Values																		0.050000	0.570000	0.568070	0.578004	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:02.808	Ascend	SBE41 Message						    5.87, 22.0958, 34.1663  	5.870000	6.449232	-0.069554	0.570000	0.280000	0.046257	1.000000																							

$	12-Dec-2018	19:09:03.353	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.870000	151.162311	29.725000	170.500000																																

$	12-Dec-2018	19:09:03.908	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.870000	151.157312	29.725000	170.500000																																

$	12-Dec-2018	19:09:04.706	Ascend	MSC: OUT: TS

$	12-Dec-2018	19:09:03.781	Ascend	Vel PID Values																		0.050000	0.750000	0.749398	0.181328	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:03.781	Ascend	SBE41 Message						    5.12, 22.0963, 34.1662  	5.120000	6.444536	-0.074043	0.750000	0.660000	0.046001	1.000000																							

$	12-Dec-2018	19:09:04.809	Ascend	Vel PID Values																		0.050000	0.060000	0.062295	-0.687103	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:04.809	Ascend	SBE41 Message						    5.06, 22.0970, 34.1658  	5.060000	6.421044	-0.072762	0.060000	0.405000	0.046004	1.000000																							

$	12-Dec-2018	19:09:05.496	Ascend	MSC: RX: ACK,TS,CMD 

$	12-Dec-2018	19:09:05.800	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.060000	151.157178	29.725000	170.500000																																

$	12-Dec-2018	19:09:05.911	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.2052524

$	12-Dec-2018	19:09:05.780	Ascend	Vel PID Values																		0.050000	0.000000	0.000203	-0.062092	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:05.780	Ascend	SBE41 Message						    5.06, 22.0973, 34.1661  	5.060000	6.336094	-0.057888	0.000000	0.030000	0.046298	1.000000																							

$	12-Dec-2018	19:09:06.560	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.060000	151.143521	29.725000	170.500000																																

$	12-Dec-2018	19:09:06.778	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000203	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:06.778	Ascend	SBE41 Message						    5.06, 22.0972, 34.1663  	5.060000	6.127450	-0.017157	0.000000	0.000000	0.047209	1.000000																							

$	12-Dec-2018	19:09:07.561	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.060000	151.160099	29.725000	170.500000																																

$	12-Dec-2018	19:09:07.777	Ascend	Vel PID Values																		0.050000	0.050000	0.049833	0.049833	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:07.777	Ascend	SBE41 Message						    5.01, 22.0977, 34.1660  	5.010000	5.795885	0.046614	0.050000	0.025000	0.048884	1.000000																							

$	12-Dec-2018	19:09:08.561	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.010000	151.155151	29.725000	170.500000																																

$	12-Dec-2018	19:09:08.776	Ascend	Vel PID Values																		0.050000	0.030000	0.030066	-0.019767	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:08.776	Ascend	SBE41 Message						    4.98, 22.0974, 34.1664  	4.980000	5.446645	0.110303	0.030000	0.040000	0.050924	1.000000																							

$	12-Dec-2018	19:09:09.561	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.980000	151.162006	29.725000	170.500000																																

$	12-Dec-2018	19:09:09.775	Ascend	Vel PID Values																		0.050000	0.040000	0.039967	0.009900	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:09.775	Ascend	SBE41 Message						    4.94, 22.0984, 34.1659  	4.940000	5.201260	0.151603	0.040000	0.035000	0.052578	1.000000																							

$	12-Dec-2018	19:09:10.560	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.940000	151.159883	29.725000	170.500000																																

$	12-Dec-2018	19:09:10.774	Ascend	Vel PID Values																		0.050000	-0.310000	-0.308835	-0.348802	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:10.774	Ascend	SBE41 Message						    5.25, 22.0981, 34.1662  	5.250000	5.085381	0.168729	-0.310000	-0.135000	0.053478	1.000000																							

$	12-Dec-2018	19:09:11.560	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.250000	151.147366	29.725000	170.500000																																

$	12-Dec-2018	19:09:12.015	Ascend	MSC: RX: ACK,TS,DAT 

$	12-Dec-2018	19:09:11.773	Ascend	Vel PID Values																		0.050000	-0.450000	-0.449532	-0.140697	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:11.773	Ascend	SBE41 Message						    5.70, 22.0981, 34.1662  	5.700000	5.041642	0.173573	-0.450000	-0.380000	0.053649	1.000000																							

$	12-Dec-2018	19:09:12.663	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.700000	151.174404	29.725000	170.500000																																

$	12-Dec-2018	19:09:12.773	Ascend	Vel PID Values																		0.050000	0.030000	0.028403	0.477935	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:12.773	Ascend	SBE41 Message						    5.67, 22.0983, 34.1660  	5.670000	5.022852	0.172982	0.030000	-0.210000	0.053244	1.000000																							

$	12-Dec-2018	19:09:13.560	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.670000	151.152872	29.725000	170.500000																																

$	12-Dec-2018	19:09:14.015	Ascend	MSC: OUT: SL

$	12-Dec-2018	19:09:13.770	Ascend	Vel PID Values																		0.050000	0.030000	0.029998	0.001594	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:13.770	Ascend	SBE41 Message						    5.64, 22.0987, 34.1661  	5.640000	5.040050	0.160322	0.030000	0.030000	0.052331	1.000000																							

$	12-Dec-2018	19:09:14.662	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.640000	151.161551	29.725000	170.500000																																

$	12-Dec-2018	19:09:15.015	Ascend	MSC: RX: 0x6534,A,01/31/2012 01:34:17,0,-1.00,-1.0000,-1.0000,227,145,1,65,23.17,56.08,11.576,5.246e-01,10531.00,2219.00,1182.83,13.17,-20.49,-191.97,3.406e-02,17,87,4875470444E7421B3FC73DF43CCC3C6C3CBB3E0C4032434847564C5D523058825F79663AC79377F4085008A988FB893D79695981797E4960392B68E3C889982137BBA75336F3E69B264A160055C1D58A155BF536A51784FF94EE54E2B4DC74DBD,6660.2 

$	12-Dec-2018	19:09:15.124	Ascend	MSC: Data: 0x6534,A,01/31/2012 01:34:17,0,-1.00,-1.0000,-1.0000,227,145,1,65,23.17,56.08,11.576,5.246e-01,10531.00,2219.00,1182.83,13.17,-20.49,-191.97,3.406e-02,17,87,4875470444E7421B3FC73DF43CCC3C6C3CBB3E0C4032434847564C5D523058825F79663AC79377F4085008A988FB893D79695981797E4960392B68E3C889982137BBA75336F3E69B264A160055C1D58A155BF536A51784FF94EE54E2B4DC74DBD,6660.2 

$	12-Dec-2018	19:09:14.772	Ascend	Vel PID Values																		0.050000	0.010000	0.010066	-0.019931	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:14.772	Ascend	SBE41 Message						    5.63, 22.0993, 34.1658  	5.630000	5.132804	0.123021	0.010000	0.020000	0.051037	1.000000																							

$	12-Dec-2018	19:09:15.779	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.630000	151.157044	29.725000	170.500000																																

$	12-Dec-2018	19:09:16.014	Ascend	MSC: Done Sampling.

$	12-Dec-2018	19:09:16.115	Ascend	Starting pH sample

$	12-Dec-2018	19:09:16.120	Ascend	MSC: OUT: W

$	12-Dec-2018	19:09:16.569	Ascend	MSC: OUT: W

$	12-Dec-2018	19:09:15.768	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.010033	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:15.768	Ascend	SBE41 Message						    5.63, 22.0990, 34.1662  	5.630000	5.301889	0.061091	0.000000	0.005000	0.049719	1.000000																							

$	12-Dec-2018	19:09:17.014	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:09:17.018	Ascend	MSC: Got ACK

$	12-Dec-2018	19:09:17.218	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:09:17.222	Ascend	MSC: Got ACK

$	12-Dec-2018	19:09:16.768	Ascend	Vel PID Values																		0.050000	0.010000	0.009967	0.009934	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:16.768	Ascend	SBE41 Message						    5.62, 22.0993, 34.1662  	5.620000	5.481300	-0.004293	0.010000	0.005000	0.048798	1.000000																							

$	12-Dec-2018	19:09:17.968	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.620000	151.150301	29.725000	170.500000																																

$	12-Dec-2018	19:09:18.079	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.0647610

$	12-Dec-2018	19:09:18.184	Ascend	MSC: OUT: MODE,2

$	12-Dec-2018	19:09:17.766	Ascend	Vel PID Values																		0.050000	0.010000	0.010000	0.000033	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:17.766	Ascend	SBE41 Message						    5.61, 22.0994, 34.1664  	5.610000	5.596900	-0.048947	0.010000	0.010000	0.048421	1.000000																							

$	12-Dec-2018	19:09:18.831	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.610000	151.173197	29.725000	170.500000																																

$	12-Dec-2018	19:09:19.086	Ascend	MSC: RX: ACK,MODE,2 

$	12-Dec-2018	19:09:18.765	Ascend	Vel PID Values																		0.050000	0.190000	0.189401	0.179401	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:18.765	Ascend	SBE41 Message						    5.42, 22.1001, 34.1660  	5.420000	5.636065	-0.068136	0.190000	0.100000	0.048393	1.000000																							

$	12-Dec-2018	19:09:19.732	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	5.420000	151.170693	29.725000	170.500000																																

$	12-Dec-2018	19:09:19.847	Ascend	Vel PID Values																		0.050000	0.910000	0.907603	0.718202	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:19.847	Ascend	SBE41 Message						    4.51, 22.1003, 34.1663  	4.510000	5.635068	-0.073431	0.910000	0.550000	0.048444	1.000000																							

$	12-Dec-2018	19:09:20.631	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.510000	151.159376	29.725000	170.500000																																

$	12-Dec-2018	19:09:21.087	Ascend	MSC: OUT: ACK

$	12-Dec-2018	19:09:20.763	Ascend	Vel PID Values																		0.050000	0.260000	0.262159	-0.645444	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:20.763	Ascend	SBE41 Message						    4.25, 22.1001, 34.1668  	4.250000	5.618798	-0.073737	0.260000	0.585000	0.048523	1.000000																							

$	12-Dec-2018	19:09:21.744	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.250000	151.156269	29.725000	170.500000																																

$	12-Dec-2018	19:09:21.855	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009098	-0.271257	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:21.855	Ascend	SBE41 Message						    4.26, 22.1007, 34.1664  	4.260000	5.565047	-0.064947	-0.010000	0.125000	0.048627	1.000000																							

$	12-Dec-2018	19:09:22.646	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.260000	151.168078	29.725000	170.500000																																

$	12-Dec-2018	19:09:23.087	Ascend	MSC: OUT: TS

$	12-Dec-2018	19:09:22.761	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009998	-0.000900	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:22.761	Ascend	SBE41 Message						    4.27, 22.1014, 34.1662  	4.270000	5.412870	-0.034644	-0.010000	-0.010000	0.048888	1.000000																							

$	12-Dec-2018	19:09:23.735	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.270000	151.158199	29.725000	170.500000																																

$	12-Dec-2018	19:09:24.087	Ascend	MSC: RX: ACK,TS,CMD 

$	12-Dec-2018	19:09:23.844	Ascend	Vel PID Values																		0.050000	0.030000	0.029867	0.039865	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:23.844	Ascend	SBE41 Message						    4.24, 22.1012, 34.1666  	4.240000	5.126977	0.021627	0.030000	0.010000	0.049321	1.000000																							

$	12-Dec-2018	19:09:24.733	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.240000	151.173189	29.725000	170.500000																																

$	12-Dec-2018	19:09:24.843	Ascend	Vel PID Values																		0.050000	0.040000	0.039966	0.010099	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:24.843	Ascend	SBE41 Message						    4.20, 22.1011, 34.1667  	4.200000	4.772138	0.087701	0.040000	0.035000	0.049916	1.000000																							

$	12-Dec-2018	19:09:25.632	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.200000	151.169971	29.725000	170.500000																																

$	12-Dec-2018	19:09:25.760	Ascend	Vel PID Values																		0.050000	0.040000	0.040000	0.000033	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:25.760	Ascend	SBE41 Message						    4.16, 22.1017, 34.1653  	4.160000	4.479142	0.138251	0.040000	0.040000	0.050531	1.000000																							

$	12-Dec-2018	19:09:26.631	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.160000	151.157685	29.725000	170.500000																																

$	12-Dec-2018	19:09:26.757	Ascend	Vel PID Values																		0.050000	0.030000	0.030033	-0.009967	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:26.757	Ascend	SBE41 Message						    4.13, 22.1013, 34.1670  	4.130000	4.319307	0.162892	0.030000	0.035000	0.050939	1.000000																							

$	12-Dec-2018	19:09:27.631	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.130000	151.167400	29.725000	170.500000																																

$	12-Dec-2018	19:09:27.756	Ascend	Vel PID Values																		0.050000	-0.350000	-0.348735	-0.378768	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:27.756	Ascend	SBE41 Message						    4.48, 22.1019, 34.1668  	4.480000	4.258166	0.170355	-0.350000	-0.160000	0.050970	1.000000																							

$	12-Dec-2018	19:09:28.632	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.480000	151.156284	29.725000	170.500000																																

$	12-Dec-2018	19:09:28.755	Ascend	Vel PID Values																		0.050000	-0.320000	-0.320098	0.028637	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:28.755	Ascend	SBE41 Message						    4.80, 22.1025, 34.1664  	4.800000	4.230077	0.171817	-0.320000	-0.335000	0.050620	1.000000																							

$	12-Dec-2018	19:09:29.633	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.800000	151.159115	29.725000	170.500000																																

$	12-Dec-2018	19:09:29.754	Ascend	Vel PID Values																		0.050000	0.030000	0.028835	0.348933	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:29.754	Ascend	SBE41 Message						    4.77, 22.1027, 34.1662  	4.770000	4.211512	0.167442	0.030000	-0.145000	0.050117	1.000000																							

$	12-Dec-2018	19:09:30.632	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.770000	151.175082	29.725000	170.500000																																

$	12-Dec-2018	19:09:30.753	Ascend	Vel PID Values																		0.050000	0.020000	0.020031	-0.008804	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:30.753	Ascend	SBE41 Message						    4.75, 22.1029, 34.1666  	4.750000	4.228247	0.146521	0.020000	0.025000	0.049846	1.000000																							

$	12-Dec-2018	19:09:31.632	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.750000	151.171103	29.725000	170.500000																																

$	12-Dec-2018	19:09:31.753	Ascend	Vel PID Values																		0.050000	0.000000	0.000067	-0.019965	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:31.753	Ascend	SBE41 Message						    4.75, 22.1035, 34.1661  	4.750000	4.316835	0.100190	0.000000	0.010000	0.050068	1.000000																							

$	12-Dec-2018	19:09:32.633	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.750000	151.170060	29.725000	170.500000																																

$	12-Dec-2018	19:09:32.752	Ascend	Vel PID Values																		0.050000	0.020000	0.019934	0.019867	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:32.752	Ascend	SBE41 Message						    4.73, 22.1032, 34.1666  	4.730000	4.469944	0.037369	0.020000	0.010000	0.050664	1.000000																							

$	12-Dec-2018	19:09:33.634	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.730000	151.180983	29.725000	170.500000																																

$	12-Dec-2018	19:09:33.750	Ascend	Vel PID Values																		0.050000	0.000000	0.000066	-0.019867	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:33.750	Ascend	SBE41 Message						    4.73, 22.1038, 34.1663  	4.730000	4.623917	-0.017908	0.000000	0.010000	0.051318	1.000000																							

$	12-Dec-2018	19:09:34.635	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.730000	151.173070	29.725000	170.500000																																

$	12-Dec-2018	19:09:35.090	Ascend	MSC: RX: ACK,TS,DAT 

$	12-Dec-2018	19:09:34.749	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000066	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:34.749	Ascend	SBE41 Message						    4.73, 22.1033, 34.1668  	4.730000	4.717968	-0.049317	0.000000	0.000000	0.051792	1.000000																							

$	12-Dec-2018	19:09:35.740	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.730000	151.160218	29.725000	170.500000																																

$	12-Dec-2018	19:09:35.850	Ascend	Vel PID Values																		0.050000	0.040000	0.039867	0.039867	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:35.850	Ascend	SBE41 Message						    4.69, 22.1042, 34.1660  	4.690000	4.747576	-0.060555	0.040000	0.020000	0.052066	1.000000																							

$	12-Dec-2018	19:09:36.635	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	4.690000	151.179284	29.725000	170.500000																																

$	12-Dec-2018	19:09:37.090	Ascend	MSC: OUT: SL

$	12-Dec-2018	19:09:36.748	Ascend	Vel PID Values																		0.050000	0.710000	0.707770	0.667903	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:36.748	Ascend	SBE41 Message						    3.98, 22.1043, 34.1668  	3.980000	4.745414	-0.063769	0.710000	0.375000	0.052208	1.000000																							

$	12-Dec-2018	19:09:37.742	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 24  Pres, BP, BatV, BatI: 	3.980000	151.175410	29.725000	170.500000																																

$	12-Dec-2018	19:09:38.095	Ascend	ERROR: Missed PH message 34

$	12-Dec-2018	19:09:38.199	Ascend	Sending Optode Do Sample

$	12-Dec-2018	19:09:38.205	Ascend	Sending Optode Wakeup

$	12-Dec-2018	19:09:38.807	Ascend	Sending FLBB run command

$	12-Dec-2018	19:09:38.823	Ascend	Sending radiometer pollSample

$	12-Dec-2018	19:09:39.027	Ascend	MSC: RX: 0x6818,pH,01/31/2012 01:34:42,0,-1.00,-1.0000,-1.0000,255,145,0,23.19,56.03,11.744,4.660e-02,7.1234,0.000,-0.976962,0.000608,-0.,-1.445e-09,+3.126e-08 

$	12-Dec-2018	19:09:39.131	Ascend	MSC: Data: 0x6818,pH,01/31/2012 01:34:42,0,-1.00,-1.0000,-1.0000,255,145,0,23.19,56.03,11.744,4.660e-02,7.1234,0.000,-0.976962,0.000608,-0.,-1.445e-09,+3.126e-08 

$	12-Dec-2018	19:09:37.852	Ascend	Vel PID Values																		0.050000	0.520000	0.520629	-0.187142	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:37.852	Ascend	SBE41 Message						    3.46, 22.1047, 34.1667  	3.460000	4.734547	-0.064714	0.520000	0.615000	0.052027	1.000000																							

$	12-Dec-2018	19:09:38.745	Ascend	Vel PID Values																		0.050000	0.020000	0.021665	-0.498964	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:38.745	Ascend	SBE41 Message						    3.44, 22.1046, 34.1671  	3.440000	4.703643	-0.058807	0.020000	0.270000	0.051312	1.000000																							

$	12-Dec-2018	19:09:39.826	Ascend	Radiometer Message																																	SATFI40287,-0.0042,0.0019,-0.0070,-0.2216 			

$	12-Dec-2018	19:09:40.246	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.440000	151.157878	29.725000	170.500000																																

$	12-Dec-2018	19:09:40.358	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.2694050

$	12-Dec-2018	19:09:40.461	Ascend	MSC: Done Sampling.

$	12-Dec-2018	19:09:40.465	Ascend	MSC: OUT: SLP

$	12-Dec-2018	19:09:39.809	Ascend	Vel PID Values																		0.050000	0.000000	0.000069	-0.021596	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:39.809	Ascend	SBE41 Message						    3.44, 22.1048, 34.1672  	3.440000	4.601908	-0.035259	0.000000	0.010000	0.050012	1.000000																							

$	12-Dec-2018	19:09:39.925	Ascend	Optode Message																												4330,1532,264.625,22.492,29.161,29.091,36.133,7.042,537.6,466.2,110.6  								

$	12-Dec-2018	19:09:39.926	Ascend	Optode Message																												#  								

$	12-Dec-2018	19:09:41.137	Ascend	FLBB Message																											/99/99,99:99:99,695,2303,700,4130,532  									

$	12-Dec-2018	19:09:40.744	Ascend	Vel PID Values																		0.050000	0.010000	0.009967	0.009897	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:40.744	Ascend	SBE41 Message						    3.43, 22.1052, 34.1670  	3.430000	4.376465	0.011557	0.010000	0.005000	0.048624	1.000000																							

$	12-Dec-2018	19:09:41.591	Ascend	MSC: RX: ACK,SLP 

$	12-Dec-2018	19:09:41.994	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.430000	151.169643	29.725000	170.500000																																

$	12-Dec-2018	19:09:42.105	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.7412469

$	12-Dec-2018	19:09:41.791	Ascend	Vel PID Values																		0.050000	0.020000	0.019967	0.010000	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:41.791	Ascend	SBE41 Message						    3.41, 22.1054, 34.1670  	3.410000	4.051119	0.070832	0.020000	0.015000	0.047703	1.000000																							

$	12-Dec-2018	19:09:42.754	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.410000	151.178800	29.725000	170.500000																																

$	12-Dec-2018	19:09:42.741	Ascend	Vel PID Values																		0.050000	0.040000	0.039934	0.019967	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:42.741	Ascend	SBE41 Message						    3.37, 22.1060, 34.1669  	3.370000	3.738901	0.121087	0.040000	0.030000	0.047402	1.000000																							

$	12-Dec-2018	19:09:43.656	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.370000	151.153781	29.725000	170.500000																																

$	12-Dec-2018	19:09:43.768	Ascend	Vel PID Values																		0.050000	0.030000	0.030033	-0.009900	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:43.768	Ascend	SBE41 Message						    3.34, 22.1060, 34.1671  	3.340000	3.539128	0.149483	0.030000	0.035000	0.047341	1.000000																							

$	12-Dec-2018	19:09:44.656	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.340000	151.148126	29.725000	170.500000																																

$	12-Dec-2018	19:09:44.771	Ascend	Vel PID Values																		0.050000	0.030000	0.030000	-0.000033	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:44.771	Ascend	SBE41 Message						    3.31, 22.1064, 34.1670  	3.310000	3.453366	0.159850	0.030000	0.030000	0.047137	1.000000																							

$	12-Dec-2018	19:09:45.655	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.310000	151.158273	29.725000	170.500000																																

$	12-Dec-2018	19:09:45.769	Ascend	Vel PID Values																		0.050000	0.030000	0.030000	-0.000000	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:45.769	Ascend	SBE41 Message						    3.28, 22.1070, 34.1662  	3.280000	3.422028	0.162392	0.030000	0.030000	0.046819	1.000000																							

$	12-Dec-2018	19:09:46.657	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.280000	151.169330	29.725000	170.500000																																

$	12-Dec-2018	19:09:46.770	Ascend	Vel PID Values																		0.050000	-0.320000	-0.318835	-0.348835	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:46.770	Ascend	SBE41 Message						    3.60, 22.1060, 34.1590  	3.600000	3.399088	0.161392	-0.320000	-0.145000	0.046842	1.000000																							

$	12-Dec-2018	19:09:47.658	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.600000	151.159875	29.725000	170.500000																																

$	12-Dec-2018	19:09:47.770	Ascend	Vel PID Values																		0.050000	-0.200000	-0.200397	0.118438	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:47.770	Ascend	SBE41 Message						    3.80, 22.1070, 34.1669  	3.800000	3.372547	0.152062	-0.200000	-0.260000	0.047725	1.000000																							

$	12-Dec-2018	19:09:48.659	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.800000	151.163802	29.725000	170.500000																																

$	12-Dec-2018	19:09:48.769	Ascend	Vel PID Values																		0.050000	0.030000	0.029234	0.229631	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:48.769	Ascend	SBE41 Message						    3.77, 22.1070, 34.1670  	3.770000	3.352592	0.126622	0.030000	-0.085000	0.049571	1.000000																							

$	12-Dec-2018	19:09:49.658	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.770000	151.180878	29.725000	170.500000																																

$	12-Dec-2018	19:09:49.771	Ascend	Vel PID Values																		0.050000	0.030000	0.029999	0.000765	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:49.771	Ascend	SBE41 Message						    3.74, 22.1072, 34.1671  	3.740000	3.365786	0.084746	0.030000	0.030000	0.051920	1.000000																							

$	12-Dec-2018	19:09:50.661	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.740000	151.157528	29.725000	170.500000																																

$	12-Dec-2018	19:09:50.772	Ascend	Vel PID Values																		0.050000	-0.010000	-0.009867	-0.039866	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:50.772	Ascend	SBE41 Message						    3.75, 22.1075, 34.1670  	3.750000	3.437771	0.037239	-0.010000	0.010000	0.054107	1.000000																							

$	12-Dec-2018	19:09:51.661	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.750000	151.160248	29.725000	170.500000																																

$	12-Dec-2018	19:09:51.773	Ascend	Vel PID Values																		0.050000	0.010000	0.009934	0.019801	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:51.773	Ascend	SBE41 Message						    3.74, 22.1071, 34.1675  	3.740000	3.556928	-0.002195	0.010000	0.000000	0.055717	1.000000																							

$	12-Dec-2018	19:09:52.661	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.740000	151.173852	29.725000	170.500000																																

$	12-Dec-2018	19:09:52.773	Ascend	Vel PID Values																		0.050000	0.010000	0.010000	0.000066	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:52.773	Ascend	SBE41 Message						    3.73, 22.1082, 34.1668  	3.730000	3.670181	-0.026810	0.010000	0.010000	0.056707	1.000000																							

$	12-Dec-2018	19:09:53.661	Ascend	PTH Message														797.331299	23.666667	54.359955	24.664951																			

$	12-Dec-2018	19:09:53.864	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.730000	151.166551	29.775000	158.000000																																

$	12-Dec-2018	19:09:53.754	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.009967	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:53.754	Ascend	SBE41 Message						    3.73, 22.1075, 34.1669  	3.730000	3.733540	-0.038535	0.000000	0.005000	0.057151	1.000000																							

$	12-Dec-2018	19:09:54.682	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.730000	151.162721	29.775000	158.000000																																

$	12-Dec-2018	19:09:54.794	Ascend	Vel PID Values																		0.050000	0.010000	0.009967	0.009933	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:54.794	Ascend	SBE41 Message						    3.72, 22.1084, 34.1666  	3.720000	3.749671	-0.043380	0.010000	0.005000	0.056921	1.000000																							

$	12-Dec-2018	19:09:55.663	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.720000	151.176341	29.775000	158.000000																																

$	12-Dec-2018	19:09:55.775	Ascend	Vel PID Values																		0.050000	0.280000	0.279101	0.269135	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:55.775	Ascend	SBE41 Message						    3.44, 22.1085, 34.1669  	3.440000	3.745866	-0.046197	0.280000	0.145000	0.055801	1.000000																							

$	12-Dec-2018	19:09:56.663	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	3.440000	151.162602	29.775000	158.000000																																

$	12-Dec-2018	19:09:56.775	Ascend	Vel PID Values																		0.050000	0.760000	0.758401	0.479300	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:56.775	Ascend	SBE41 Message						    2.68, 22.1090, 34.1670  	2.680000	3.738486	-0.047745	0.760000	0.520000	0.053724	1.000000																							

$	12-Dec-2018	19:09:57.662	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.680000	151.182137	29.775000	158.000000																																

$	12-Dec-2018	19:09:57.774	Ascend	Vel PID Values																		0.050000	0.240000	0.241728	-0.516673	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:57.774	Ascend	SBE41 Message						    2.44, 22.1082, 34.1678  	2.440000	3.722400	-0.044112	0.240000	0.500000	0.050934	1.000000																							

$	12-Dec-2018	19:09:58.662	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.440000	151.154764	29.775000	158.000000																																

$	12-Dec-2018	19:09:58.774	Ascend	Vel PID Values																		0.050000	0.000000	0.000802	-0.240926	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:58.774	Ascend	SBE41 Message						    2.44, 22.1088, 34.1676  	2.440000	3.665010	-0.028103	0.000000	0.120000	0.047850	1.000000																							

$	12-Dec-2018	19:09:59.663	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.440000	151.177995	29.775000	158.000000																																

$	12-Dec-2018	19:09:59.775	Ascend	Vel PID Values																		0.050000	0.000000	0.000001	-0.000800	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:09:59.775	Ascend	SBE41 Message						    2.44, 22.1099, 34.1669  	2.440000	3.510717	0.005713	0.000000	0.000000	0.044959	1.000000																							

$	12-Dec-2018	19:10:00.664	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.440000	151.170269	29.775000	158.000000																																

$	12-Dec-2018	19:10:00.776	Ascend	Vel PID Values																		0.050000	0.010000	0.009967	0.009965	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:00.776	Ascend	SBE41 Message						    2.43, 22.1091, 34.1677  	2.430000	3.235499	0.053749	0.010000	0.005000	0.042745	1.000000																							

$	12-Dec-2018	19:10:01.665	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.430000	151.174054	29.775000	158.000000																																

$	12-Dec-2018	19:10:01.777	Ascend	Vel PID Values																		0.050000	0.010000	0.010000	0.000033	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:01.777	Ascend	SBE41 Message						    2.42, 22.1099, 34.1663  	2.420000	2.906116	0.102322	0.010000	0.010000	0.041494	1.000000																							

$	12-Dec-2018	19:10:02.666	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.420000	151.159137	29.775000	158.000000																																

$	12-Dec-2018	19:10:02.778	Ascend	Vel PID Values																		0.050000	0.040000	0.039900	0.029900	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:02.778	Ascend	SBE41 Message						    2.38, 22.1099, 34.1665  	2.380000	2.639760	0.137279	0.040000	0.025000	0.041307	1.000000																							

$	12-Dec-2018	19:10:03.666	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.380000	151.167266	29.775000	158.000000																																

$	12-Dec-2018	19:10:03.778	Ascend	Vel PID Values																		0.050000	0.020000	0.020067	-0.019834	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:03.778	Ascend	SBE41 Message						    2.36, 22.1104, 34.1673  	2.360000	2.495837	0.154639	0.020000	0.030000	0.042220	1.000000																							

$	12-Dec-2018	19:10:04.667	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.360000	151.162684	29.775000	158.000000																																

$	12-Dec-2018	19:10:04.782	Ascend	Vel PID Values																		0.050000	0.020000	0.020000	-0.000067	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:04.782	Ascend	SBE41 Message						    2.34, 22.1100, 34.1676  	2.340000	2.443446	0.160227	0.020000	0.020000	0.044189	1.000000																							

$	12-Dec-2018	19:10:05.667	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.340000	151.174799	29.775000	158.000000																																

$	12-Dec-2018	19:10:05.779	Ascend	Vel PID Values																		0.050000	0.030000	0.029967	0.009967	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:05.779	Ascend	SBE41 Message						    2.31, 22.1105, 34.1674  	2.310000	2.424516	0.160577	0.030000	0.025000	0.046930	1.000000																							

$	12-Dec-2018	19:10:06.667	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.310000	151.169956	29.775000	158.000000																																

$	12-Dec-2018	19:10:06.779	Ascend	Vel PID Values																		0.050000	-0.300000	-0.298902	-0.328869	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:06.779	Ascend	SBE41 Message						    2.61, 22.1111, 34.1672  	2.610000	2.407118	0.155608	-0.300000	-0.135000	0.049964	1.000000																							

$	12-Dec-2018	19:10:07.669	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.610000	151.179731	29.775000	158.000000																																

$	12-Dec-2018	19:10:07.781	Ascend	Vel PID Values																		0.050000	-0.150000	-0.150497	0.148404	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:07.781	Ascend	SBE41 Message						    2.76, 22.1110, 34.1672  	2.760000	2.386086	0.139096	-0.150000	-0.225000	0.052953	1.000000																							

$	12-Dec-2018	19:10:08.670	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.760000	151.179507	29.775000	158.000000																																

$	12-Dec-2018	19:10:08.782	Ascend	Vel PID Values																		0.050000	0.020000	0.019433	0.169931	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:08.782	Ascend	SBE41 Message						    2.74, 22.1109, 34.1675  	2.740000	2.371510	0.106842	0.020000	-0.065000	0.055752	1.000000																							

$	12-Dec-2018	19:10:09.669	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.740000	151.176490	29.775000	158.000000																																

$	12-Dec-2018	19:10:09.781	Ascend	Vel PID Values																		0.050000	0.010000	0.010032	-0.009401	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:09.781	Ascend	SBE41 Message						    2.73, 22.1108, 34.1677  	2.730000	2.386486	0.064253	0.010000	0.015000	0.058221	1.000000																							

$	12-Dec-2018	19:10:10.669	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.730000	151.188597	29.775000	158.000000																																

$	12-Dec-2018	19:10:10.781	Ascend	Vel PID Values																		0.050000	0.010000	0.010000	-0.000032	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:10.781	Ascend	SBE41 Message						    2.72, 22.1109, 34.1679  	2.720000	2.452344	0.023178	0.010000	0.010000	0.060043	1.000000																							

$	12-Dec-2018	19:10:11.670	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.720000	151.164696	29.775000	158.000000																																

$	12-Dec-2018	19:10:11.782	Ascend	Vel PID Values																		0.050000	0.000000	0.000033	-0.009967	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:11.782	Ascend	SBE41 Message						    2.72, 22.1118, 34.1673  	2.720000	2.557734	-0.007648	0.000000	0.005000	0.060805	1.000000																							

$	12-Dec-2018	19:10:12.671	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.720000	151.183575	29.775000	158.000000																																

$	12-Dec-2018	19:10:12.783	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000033	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:12.783	Ascend	SBE41 Message						    2.72, 22.1118, 34.1673  	2.720000	2.656679	-0.026368	0.000000	0.000000	0.060287	1.000000																							

$	12-Dec-2018	19:10:13.672	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.720000	151.176266	29.775000	158.000000																																

$	12-Dec-2018	19:10:13.785	Ascend	Vel PID Values																		0.050000	0.000000	0.000000	-0.000000	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:13.785	Ascend	SBE41 Message						    2.72, 22.1122, 34.1672  	2.720000	2.712246	-0.035593	0.000000	0.000000	0.058632	1.000000																							

$	12-Dec-2018	19:10:14.673	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.720000	151.177973	29.775000	158.000000																																

$	12-Dec-2018	19:10:14.785	Ascend	Vel PID Values																		0.050000	0.000000	-0.000000	-0.000000	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:14.785	Ascend	SBE41 Message						    2.72, 22.1123, 34.1673  	2.720000	2.727202	-0.039606	0.000000	0.000000	0.056201	1.000000																							

$	12-Dec-2018	19:10:15.675	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.720000	151.159346	29.775000	158.000000																																

$	12-Dec-2018	19:10:15.786	Ascend	Vel PID Values																		0.050000	0.260000	0.259135	0.259135	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:15.786	Ascend	SBE41 Message						    2.46, 22.1125, 34.1674  	2.460000	2.725150	-0.041957	0.260000	0.130000	0.053275	1.000000																							

$	12-Dec-2018	19:10:16.674	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	2.460000	151.198745	29.775000	158.000000																																

$	12-Dec-2018	19:10:16.786	Ascend	Vel PID Values																		0.050000	0.720000	0.718468	0.459333	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:16.786	Ascend	SBE41 Message						    1.74, 22.1127, 34.1672  	1.740000	2.720967	-0.043236	0.720000	0.490000	0.050008	1.000000																							

$	12-Dec-2018	19:10:17.675	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 25  Pres, BP, BatV, BatI: 	1.740000	151.176326	29.775000	158.000000																																

$	12-Dec-2018	19:10:18.138	Ascend	ERROR: Missed NO3 message 31

$	12-Dec-2018	19:10:18.242	Ascend	Sending Optode Do Sample

$	12-Dec-2018	19:10:18.246	Ascend	Sending Optode Wakeup

$	12-Dec-2018	19:10:18.861	Ascend	Sending FLBB run command

$	12-Dec-2018	19:10:18.865	Ascend	Sending radiometer pollSample

$	12-Dec-2018	19:10:17.787	Ascend	Vel PID Values																		0.050000	0.290000	0.291428	-0.427039	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:17.787	Ascend	SBE41 Message						    1.45, 22.1127, 34.1680  	1.450000	2.710105	-0.040030	0.290000	0.505000	0.046434	1.000000																							

$	12-Dec-2018	19:10:18.705	Ascend	Vel PID Values																		0.050000	0.010000	0.010934	-0.280494	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:18.705	Ascend	SBE41 Message						    1.44, 22.1133, 34.1676  	1.440000	2.660835	-0.025784	0.010000	0.150000	0.042807	1.000000																							

$	12-Dec-2018	19:10:19.661	Ascend	Radiometer Message																																	SATFI40287,-0.0045,0.0021,-0.0061,-0.2177 			

$	12-Dec-2018	19:10:19.759	Ascend	Vel PID Values																		0.050000	0.000000	0.000035	-0.010899	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:19.759	Ascend	SBE41 Message						    1.44, 22.1131, 34.1679  	1.440000	2.517913	0.004922	0.000000	0.005000	0.039692	1.000000																							

$	12-Dec-2018	19:10:20.322	Ascend	RecMode true, PID On, BE Ext'g, Valve Open, depthNum = 26  Pres, BP, BatV, BatI: 	1.440000	151.180513	29.775000	158.000000																																

$	12-Dec-2018	19:10:20.435	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.3029362

$	12-Dec-2018	19:10:20.538	Ascend	Starting ISUS sample

$	12-Dec-2018	19:10:20.641	Ascend	Starting MSC Sequence CTDValues (time, press, temp, salinity) 0, 0.00, 0.000, 0.00

$	12-Dec-2018	19:10:20.745	Ascend	MSC: OUT: W

$	12-Dec-2018	19:10:21.000	Ascend	MSC: OUT: W

$	12-Dec-2018	19:10:21.103	Ascend	Sending SBE41 stopProfile

$	12-Dec-2018	19:10:19.938	Ascend	Optode Message																												4330,1532,264.561,22.504,29.158,29.089,36.132,7.044,537.5,466.0,110.2  								

$	12-Dec-2018	19:10:19.941	Ascend	Optode Message																												#  								

$	12-Dec-2018	19:10:21.347	Ascend	FLBB Message																											/99/99,99:99:99,695,2301,700,4130,532  									

$	12-Dec-2018	19:10:20.744	Ascend	Vel PID Values																		0.050000	0.010000	0.009967	0.009932	NaN	NaN	NaN	0.000000	1.000000										

$	12-Dec-2018	19:10:20.744	Ascend	SBE41 Message						    1.43, 22.1130, 34.1681  	1.430000	2.254069	0.049781	0.010000	0.005000	0.038022	1.000000																							

$	12-Dec-2018	19:10:21.900	Ascend	MSC: RX: 

$	12-Dec-2018	19:10:22.002	Ascend	MSC: RX: 

$	12-Dec-2018	19:10:22.006	Ascend	MSC: RX: ACK 

$	12-Dec-2018	19:10:22.204	Ascend	MSC: Got ACK

$	12-Dec-2018	19:10:22.411	Ascend	Sending SBE41 stopProfile

$	12-Dec-2018	19:10:22.416	Ascend	Powering up GPS

$	12-Dec-2018	19:10:22.532	Ascend	Powering up BT Module

$	12-Dec-2018	19:10:22.633	Ascend	Normal State Exit

$	12-Dec-2018	19:10:22.735	Ascend	RecMode true, PID Off, Motor Stopped, Valve Open, depthNum = 26  Pres, BP, BatV, BatI: 	1.430000	151.179321	29.775000	158.000000																																

$	12-Dec-2018	19:10:22.948	Ascend	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.5075024

$	12-Dec-2018	19:10:22.953	Init Profile	State Entry

$	12-Dec-2018	19:10:23.056	Init Profile	Initializing SBE 41 LPF

$	12-Dec-2018	19:10:23.265	Init Profile	Sending SBE41 stopProfile

$	12-Dec-2018	19:10:24.270	Init Profile	Resetting SBE41 with qsr

$	12-Dec-2018	19:10:24.274	Init Profile	Sending qsr command

$	12-Dec-2018	19:10:21.786	Init Profile	SBE41 Message						    1.42, 22.1135, 34.1678  	1.420000	1.430000	0.000000	0.010000	0.005000	-0.000000	1.000000																							

$	12-Dec-2018	19:10:21.900	Init Profile	Unparsed SBE41 Message						profile stopped  																														

$	12-Dec-2018	19:10:22.527	Init Profile	Unparsed SBE41 Message						S>stopprofile  																														

$	12-Dec-2018	19:10:27.012	Init Profile	Normal State Exit

$	12-Dec-2018	19:10:27.016	Init Profile	RecMode true, PID Off, Motor Stopped, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.179008	29.775000	158.000000																																

$	12-Dec-2018	19:10:27.229	Init Profile	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:04.2759276

$	12-Dec-2018	19:10:27.236	Surface Ops Set Bellows	State Entry

$	12-Dec-2018	19:10:27.441	Surface Ops Set Bellows	Inflating bellows to Surface Ops position = 155.0 at 200000 cps

$	12-Dec-2018	19:10:23.382	Surface Ops Set Bellows	Unparsed SBE41 Message						?cmd S>stopprofile  																														

$	12-Dec-2018	19:10:24.478	Surface Ops Set Bellows	Unparsed SBE41 Message						?cmd S>qsr  																														

$	12-Dec-2018	19:10:24.478	Surface Ops Set Bellows	Unparsed SBE41 Message						powering down  																														

$	12-Dec-2018	19:10:26.710	Surface Ops Set Bellows	Unparsed SBE41 Message						S>SBE 41CP UW V 2.0  																														

$	12-Dec-2018	19:10:28.426	Surface Ops Set Bellows	Normal State Exit

$	12-Dec-2018	19:10:28.430	Surface Ops Set Bellows	RecMode true, PID Off, Motor Stopped, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.167035	29.775000	158.000000																																

$	12-Dec-2018	19:10:28.645	Surface Ops Set Bellows	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:01.4087545

$	12-Dec-2018	19:10:28.651	Dump CP Data	State Entry

$	12-Dec-2018	19:10:28.849	Dump CP Data	Starting valve close

$	12-Dec-2018	19:10:28.856	Dump CP Data	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:10:29.264	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Open, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.167654	29.775000	158.000000																																

$	12-Dec-2018	19:10:29.651	Dump CP Data	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:10:29.957	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.159361	29.775000	158.000000																																

$	12-Dec-2018	19:10:30.650	Dump CP Data	Didn't get CR/LF response, trying again

$	12-Dec-2018	19:10:28.958	Dump CP Data	Unparsed SBE41 Message						S>  																														

$	12-Dec-2018	19:10:31.065	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.161745	29.775000	158.000000																																

$	12-Dec-2018	19:10:31.651	Dump CP Data	Got CR/LF response

$	12-Dec-2018	19:10:31.858	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.165940	29.775000	158.000000																																

$	12-Dec-2018	19:10:32.652	Dump CP Data	Sending binaverage wait command

$	12-Dec-2018	19:10:32.657	Dump CP Data	Sending binaverage

$	12-Dec-2018	19:10:32.869	Dump CP Data	Trying again for binaverage response

$	12-Dec-2018	19:10:32.868	Dump CP Data	Unparsed SBE41 Message						S>binaverage  																														

$	12-Dec-2018	19:10:32.869	Dump CP Data	Unparsed SBE41 Message						samples = 428, maxPress = 25.66  																														

$	12-Dec-2018	19:10:32.972	Dump CP Data	Unparsed SBE41 Message						rd: 0  																														

$	12-Dec-2018	19:10:33.404	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.156813	29.775000	158.000000																																

$	12-Dec-2018	19:10:33.652	Dump CP Data	Sending binaverage wait command

$	12-Dec-2018	19:10:33.656	Dump CP Data	Trying again for binaverage response

$	12-Dec-2018	19:10:34.058	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.168019	29.775000	158.000000																																

$	12-Dec-2018	19:10:34.652	Dump CP Data	Sending binaverage wait command

$	12-Dec-2018	19:10:34.656	Dump CP Data	Trying again for binaverage response

$	12-Dec-2018	19:10:34.334	Dump CP Data	Unparsed SBE41 Message						rd: 128  																														

$	12-Dec-2018	19:10:35.068	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.167430	29.775000	158.000000																																

$	12-Dec-2018	19:10:35.652	Dump CP Data	Sending binaverage wait command

$	12-Dec-2018	19:10:35.657	Dump CP Data	Trying again for binaverage response

$	12-Dec-2018	19:10:35.754	Dump CP Data	Unparsed SBE41 Message						rd: 256  																														

$	12-Dec-2018	19:10:36.069	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.158132	29.775000	158.000000																																

$	12-Dec-2018	19:10:36.653	Dump CP Data	Sending binaverage wait command

$	12-Dec-2018	19:10:36.657	Dump CP Data	Trying again for binaverage response

$	12-Dec-2018	19:10:36.962	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.187256	29.775000	158.000000																																

$	12-Dec-2018	19:10:37.652	Dump CP Data	Sending binaverage wait command

$	12-Dec-2018	19:10:37.657	Dump CP Data	Trying again for binaverage response

$	12-Dec-2018	19:10:37.113	Dump CP Data	Unparsed SBE41 Message						rd: 384  																														

$	12-Dec-2018	19:10:37.587	Dump CP Data	Unparsed SBE41 Message						avg: 0  																														

$	12-Dec-2018	19:10:37.755	Dump CP Data	Unparsed SBE41 Message						  																														

$	12-Dec-2018	19:10:37.755	Dump CP Data	Unparsed SBE41 Message						done, nbins = 14  																														

$	12-Dec-2018	19:10:38.400	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.167594	29.775000	158.000000																																

$	12-Dec-2018	19:10:38.653	Dump CP Data	Sending binaverage wait command

$	12-Dec-2018	19:10:38.658	Dump CP Data	Got good binaverage response

$	12-Dec-2018	19:10:38.961	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.175514	29.775000	158.000000																																

$	12-Dec-2018	19:10:39.654	Dump CP Data	Sending da wait command

$	12-Dec-2018	19:10:39.658	Dump CP Data	Sending da command

$	12-Dec-2018	19:10:39.759	Dump CP Data	Trying again for da response

$	12-Dec-2018	19:10:39.861	Dump CP Data	Unparsed SBE41 Message						S>da  																														

$	12-Dec-2018	19:10:39.861	Dump CP Data	Unparsed SBE41 Message						    0.00,  0.0000,  0.0000, 0  																														

$	12-Dec-2018	19:10:39.965	Dump CP Data	Unparsed SBE41 Message						    2.31, 22.1111, 34.1674, 26  																														

$	12-Dec-2018	19:10:39.966	Dump CP Data	Unparsed SBE41 Message						    4.05, 22.1041, 34.1665, 39  																														

$	12-Dec-2018	19:10:39.967	Dump CP Data	Unparsed SBE41 Message						    5.98, 22.0946, 34.1658, 39  																														

$	12-Dec-2018	19:10:40.076	Dump CP Data	Unparsed SBE41 Message						    8.05, 22.0846, 34.1649, 43  																														

$	12-Dec-2018	19:10:40.077	Dump CP Data	Unparsed SBE41 Message						   10.05, 22.0740, 34.1639, 34  																														

$	12-Dec-2018	19:10:40.078	Dump CP Data	Unparsed SBE41 Message						   11.87, 22.0649, 34.1631, 36  																														

$	12-Dec-2018	19:10:40.079	Dump CP Data	Unparsed SBE41 Message						   13.68, 22.0537, 34.1622, 34  																														

$	12-Dec-2018	19:10:40.188	Dump CP Data	Unparsed SBE41 Message						   15.68, 22.0394, 34.1612, 47  																														

$	12-Dec-2018	19:10:40.189	Dump CP Data	Unparsed SBE41 Message						   17.96, 22.0238, 34.1600, 39  																														

$	12-Dec-2018	19:10:40.190	Dump CP Data	Unparsed SBE41 Message						   19.99, 22.0116, 34.1590, 20  																														

$	12-Dec-2018	19:10:40.301	Dump CP Data	Unparsed SBE41 Message						   22.02, 21.9979, 34.1582, 37  																														

$	12-Dec-2018	19:10:40.302	Dump CP Data	Unparsed SBE41 Message						   24.25, 21.9770, 34.1566, 28  																														

$	12-Dec-2018	19:10:40.303	Dump CP Data	Unparsed SBE41 Message						   25.65, 21.9709, 34.1562, 6  																														

$	12-Dec-2018	19:10:40.413	Dump CP Data	Unparsed SBE41 Message						upload complete  																														

$	12-Dec-2018	19:10:41.825	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.174329	29.775000	158.000000																																

$	12-Dec-2018	19:10:41.936	Dump CP Data	ERROR: State machine cycle time exceeded SMCyclePeriod = 00:00:02.2825315

$	12-Dec-2018	19:10:41.943	Dump CP Data	Sending da wait command

$	12-Dec-2018	19:10:42.142	Dump CP Data	Good da response; Upload sucessful

$	12-Dec-2018	19:10:42.446	Dump CP Data	RecMode true, PID Off, Motor Stopped, Valve Closed, depthNum = 0  Pres, BP, BatV, BatI: 	1.420000	151.160941	29.775000	158.000000																																

$	12-Dec-2018	19:11:13.956	Dump CP Data	Sending dah wait command

$	12-Dec-2018	19:11:13.960	Dump CP Data	Sending dah command

$	12-Dec-2018	19:11:14.063	Dump CP Data	Trying again for dah response

$	12-Dec-2018	19:11:14.166	Dump CP Data	Unparsed SBE41 Message						S>dah  																														
