$	Date	Time	State	Comment	Pressure	Bel Pos	Batt Volt	Batt Amps	CTD Temp	CTD Sal	Raw Press	LPF Press	LPF Vel	Diff Vel	Diff 2 Vel	Slope of LPF Press.	CTD Sample Period	LPS331 Press	LPS311 Temp	SHT21 Humid	SHT21 Temp	PID SP	PID PV	PID Y1	PID Y2	PID I	PID v	PID u	PID uCPS	PID h	FLBB	Topt	Tphase	Rphase	NO3	pH	OCR	GPS	Irid Modem	
$	20-Dec-2018	17:41:57.718	UnitTest	Reset Controller Status Register Value = 65792

$	20-Dec-2018	17:41:57.723	UnitTest	File Name = \SD\data\cpf001.000.2018-12-20T17-41-56.eng

$	20-Dec-2018	17:41:57.727	UnitTest	Volume Total Size:     7939817472

$	20-Dec-2018	17:41:57.732	UnitTest	Volume Free Space:     7861960704

$	20-Dec-2018	17:41:57.736	UnitTest	PlatVel PID K = 80

$	20-Dec-2018	17:41:57.740	UnitTest	PlatVel PID Ti = 50000

$	20-Dec-2018	17:41:57.744	UnitTest	PlatVel PID Td = 10

$	20-Dec-2018	17:41:57.748	UnitTest	PlatVel PID Tf = 10

$	20-Dec-2018	17:41:57.752	UnitTest	PlatVel PID Tt = 100

$	20-Dec-2018	17:41:57.756	UnitTest	PlatVel PID B = 1

$	20-Dec-2018	17:41:57.760	UnitTest	PlatVel PID uLow = -50

$	20-Dec-2018	17:41:57.764	UnitTest	PlatVel PID uHigh = 50

$	20-Dec-2018	17:41:57.768	UnitTest	PlatVel PID h = 3

$	20-Dec-2018	17:41:57.772	UnitTest	Engr Logger Initialized

$	20-Dec-2018	17:41:57.778	UnitTest	Start UT: Test Sample Sequence 

$	20-Dec-2018	17:41:57.781	UnitTest	Initializing MSC

$	20-Dec-2018	17:42:07.786	UnitTest	[MSC] RX: 

$	20-Dec-2018	17:42:07.790	UnitTest	[MSC] RX: 

$	20-Dec-2018	17:42:07.794	UnitTest	[MSC] RX:   MBARI Sensor Controller (ISUS/DURA) Initializing... 

$	20-Dec-2018	17:42:07.799	UnitTest	[MSC] RX:   MBARI Serial Number : 0777 

$	20-Dec-2018	17:42:07.803	UnitTest	[MSC] RX:   CF2 App: APF_MAIN.C Date: Sep 30 2015 14:08:04 

$	20-Dec-2018	17:42:07.808	UnitTest	[MSC] RX:   CF2 S/N: 14968 - BIOS:4.03 - PicoDOS:4.03 

$	20-Dec-2018	17:42:07.813	UnitTest	[MSC] RX:   CF Card: Size=256614400 Avail=98.92% 

$	20-Dec-2018	17:42:07.817	UnitTest	[MSC] RX:   CF2 Power Up  (GMT): Wed Feb  8 00:07:37 2012 

$	20-Dec-2018	17:42:07.822	UnitTest	[MSC] RX:   CF2 RTC Time  (GMT): Wed Feb  8 00:07:41 2012 

$	20-Dec-2018	17:42:07.828	UnitTest	[MSC] RX:   Adjusted Local Time: Tue Feb  7 16:07:41 2012 

$	20-Dec-2018	17:42:07.832	UnitTest	[MSC] RX:   TZ Offset (GMT - LOCAL)  : +8.00 hrs. 

$	20-Dec-2018	17:42:07.837	UnitTest	[MSC] RX:   RTC delta (CF2 - DS3234) : -1.00 secs. 

$	20-Dec-2018	17:42:07.841	UnitTest	[MSC] RX:   System Batt Voltage = 11.764 

$	20-Dec-2018	17:42:07.846	UnitTest	[MSC] RX:   System Load Current = 2.62e-02 

$	20-Dec-2018	17:42:07.851	UnitTest	[MSC] RX:   Housing Temperature = +20.81C = +69.47F 

$	20-Dec-2018	17:42:07.855	UnitTest	[MSC] RX:   Housing    Humidity =  55.7%  =  11.6C (DP) 

$	20-Dec-2018	17:42:07.860	UnitTest	[MSC] RX:   Power Cycle   Count = 218 

$	20-Dec-2018	17:42:07.865	UnitTest	[MSC] RX:   System Reset  Count = 222 

$	20-Dec-2018	17:42:07.869	UnitTest	[MSC] RX:   System Error  Count = 98 

$	20-Dec-2018	17:42:07.873	UnitTest	[MSC] RX:   Last System   Error = -110 

$	20-Dec-2018	17:42:07.878	UnitTest	[MSC] RX:   Last Error     Time = Tue Feb  7 07:37:19 2012 

$	20-Dec-2018	17:42:07.883	UnitTest	[MSC] RX:   ISUS Error    Count = 1 

$	20-Dec-2018	17:42:07.887	UnitTest	[MSC] RX:   Last ISUS     Error = -202 

$	20-Dec-2018	17:42:07.891	UnitTest	[MSC] RX:   Last Error     Time = Thu Jan 26 03:13:44 2012 

$	20-Dec-2018	17:42:07.896	UnitTest	[MSC] RX:   DURA Error    Count = 0 

$	20-Dec-2018	17:42:07.900	UnitTest	[MSC] RX:   Last DURA     Error = 0 

$	20-Dec-2018	17:42:07.905	UnitTest	[MSC] RX:   Last Error     Time = Thu Jan  1 00:00:00 1970 

$	20-Dec-2018	17:42:07.910	UnitTest	[MSC] RX:   Fiberlite  Odometer = 100:16:45 = 361005 secs 

$	20-Dec-2018	17:42:07.915	UnitTest	[MSC] RX:   Fiberlite    Cycles = 354 

$	20-Dec-2018	17:42:07.919	UnitTest	[MSC] RX:   Spectrometer Status = MBARI-ISUS 1.1, 04-Mar-15, 10:16:08 

$	20-Dec-2018	17:42:07.924	UnitTest	[MSC] RX:   Zeiss Coef File Found 4 coefs loaded. 

$	20-Dec-2018	17:42:07.929	UnitTest	[MSC] RX:   Ecoef File Found 256 coefs loaded. 

$	20-Dec-2018	17:42:07.933	UnitTest	[MSC] Starting ISUS sample

$	20-Dec-2018	17:42:07.939	UnitTest	[MSC] Starting MSC Sequence CTDValues (time, press, temp, salinity) 1545327727, 62.00, 15.000, 35.00

$	20-Dec-2018	17:42:07.944	UnitTest	[MSC] TX: W

$	20-Dec-2018	17:42:07.949	UnitTest	[MSC] RX:   There are 104 records in the error file! 

$	20-Dec-2018	17:42:07.954	UnitTest	[MSC] RX:   MBARI Sensor Controller (ISUS/DURA) Initialized 

$	20-Dec-2018	17:42:07.959	UnitTest	[MSC] RX:   Auto Shutdown in 10 hours at: Wed Feb  8 10:07:42 2012 

$	20-Dec-2018	17:42:07.964	UnitTest	[MSC] RX:   Entering Suspend Mode, Serial Input (/WAKE) will wake me, send M for Menu Mode  

$	20-Dec-2018	17:42:07.974	UnitTest	[MSC] RX: 

$	20-Dec-2018	17:42:08.181	UnitTest	[MSC] RX: 

$	20-Dec-2018	17:42:10.945	UnitTest	[MSC] Timeout for CommandWake

$	20-Dec-2018	17:42:10.950	UnitTest	[MSC] TX: W

$	20-Dec-2018	17:42:11.124	UnitTest	[MSC] RX: ACK 

$	20-Dec-2018	17:42:11.128	UnitTest	[MSC] Got ACK

$	20-Dec-2018	17:42:11.202	UnitTest	[MSC] TX: CTD,1545327727,15.0000,35.0000,62.0000

$	20-Dec-2018	17:42:11.466	UnitTest	[MSC] RX: ACK,CTD,1545327727,15.0000,35.0000,62.00 

$	20-Dec-2018	17:42:13.066	UnitTest	[MSC] TX: ACK

$	20-Dec-2018	17:42:14.471	UnitTest	[MSC] TX: MODE,1

$	20-Dec-2018	17:42:14.686	UnitTest	[MSC] RX: ACK,MODE,1 

$	20-Dec-2018	17:42:16.286	UnitTest	[MSC] TX: ACK

$	20-Dec-2018	17:42:17.692	UnitTest	[MSC] TX: TS

$	20-Dec-2018	17:42:17.894	UnitTest	[MSC] RX: ACK,TS,CMD 

$	20-Dec-2018	17:42:24.691	UnitTest	[MSC] RX: ACK,TS,DAT 

$	20-Dec-2018	17:42:24.696	UnitTest	[MSC] TX: SL

$	20-Dec-2018	17:42:25.354	UnitTest	[MSC] RX: 0x69F4,A,02/08/2012 00:07:59,0,-1.00,-1.0000,-1.0000,351,218,1,98,20.80,55.71,11.572,5.310e-01,10921.00,2273.00,1116.41,12.55,-20.47,-191.17,3.394e-02,17,87,4B134996475C44C0425040343EF23E8E3F364067428545B449D34F0B55185BF863176A4970D7767D7A7F7CAF7D5E7C15799875F0717B6D2E694065E86345618D60C460C361EC63A9663F69A96DE272AD77E97E16840989F98FCC950299569C4A9DEA9DBD9BD2987B93908DD8876C808579F2738F6DA0687063AA5F935C0A5911569254AE530151E6511C50D450DA,6831.4 

$	20-Dec-2018	17:42:25.368	UnitTest	[MSC] Data: 0x69F4,A,02/08/2012 00:07:59,0,-1.00,-1.0000,-1.0000,351,218,1,98,20.80,55.71,11.572,5.310e-01,10921.00,2273.00,1116.41,12.55,-20.47,-191.17,3.394e-02,17,87,4B134996475C44C0425040343EF23E8E3F364067428545B449D34F0B55185BF863176A4970D7767D7A7F7CAF7D5E7C15799875F0717B6D2E694065E86345618D60C460C361EC63A9663F69A96DE272AD77E97E16840989F98FCC950299569C4A9DEA9DBD9BD2987B93908DD8876C808579F2738F6DA0687063AA5F935C0A5911569254AE530151E6511C50D450DA,6831.4 

$	20-Dec-2018	17:42:25.381	UnitTest	[MSC] POST: -191.17

$	20-Dec-2018	17:42:25.384	UnitTest	[MSC] Done Sampling.

$	20-Dec-2018	17:42:25.388	UnitTest	[MSC] Starting pH sample

$	20-Dec-2018	17:42:25.392	UnitTest	[MSC] TX: MODE,2

$	20-Dec-2018	17:42:25.570	UnitTest	[MSC] RX: ACK,MODE,2 

$	20-Dec-2018	17:42:27.170	UnitTest	[MSC] TX: ACK

$	20-Dec-2018	17:42:28.576	UnitTest	[MSC] TX: TS

$	20-Dec-2018	17:42:28.779	UnitTest	[MSC] RX: ACK,TS,CMD 

$	20-Dec-2018	17:42:37.837	UnitTest	[MSC] RX: ACK,TS,DAT 

$	20-Dec-2018	17:42:37.843	UnitTest	[MSC] TX: SL

$	20-Dec-2018	17:42:38.199	UnitTest	[MSC] RX: 0x4353,pH,02/08/2012 00:08:13,0,-1.00,-1.0000,-1.0000,361,218,0,20.81,55.71,11.748,4.680e-02,7.1234,0.000,-1.035385,0.000970,-1.027888,0.001069,+4.838e-09,+1.606e-09 

$	20-Dec-2018	17:42:38.205	UnitTest	[MSC] Data: 0x4353,pH,02/08/2012 00:08:13,0,-1.00,-1.0000,-1.0000,361,218,0,20.81,55.71,11.748,4.680e-02,7.1234,0.000,-1.035385,0.000970,-1.027888,0.001069,+4.838e-09,+1.606e-09 

$	20-Dec-2018	17:42:38.214	UnitTest	MSC post: 7.1234

$	20-Dec-2018	17:42:38.218	UnitTest	[MSC] Done Sampling.
