# ESP Delta 2002 # MAY 27 03 # new Valve manifolds # rev 012 # dry filter between fix exchanges # ~67 ml sample # depressurization stroke 2ml # fills puck with fix after dry-out # fix delays at 20 min EACH = 60 min total fix time # mod 5/27/03 for HAB/SSU WCR # mod 8/10/04 for new syringe # this one for pressure # mod 8/13 based on 8/12 run notes PROGRAM: WCR SYRINGE PULL 7.5 Sample # prime puck DELAYSEC 20 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 30 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample # sample vol ~50ml DELAYSEC 45 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME #Done sampling SYRINGEJOG PULL 3 #DEPR. EXAHUST VALVEPORT Waste HOME #ISOLATE WASTE LINE DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste DELAYSEC 2 VALVEPORT Syringe HOME #NOW DO SAMPLE SIDE SYRINGEJOG PULL 3 DELAYSEC 5 VALVEPORT Sample HOME # depress complete DELAYSEC 15 VALVEPORT M3Bypass SET DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste DELAYSEC 2 SYRINGE PULL 1.5 M3Bypass DELAYSEC 2 SYRINGE PUSH ALL ToxicWaste # burp M1 loop DELAYSEC 2 SYRINGE PULL 2.5 Fixative DELAYSEC 5 VALVEPORT Fixative HOME SYRINGEJOG PULL 2 SYRINGE PUSH 4 Puck # RINSE FILTER WITH FIX VALVEPORT Puck HOME VALVEPORT Syringe HOME SYRINGEJOG PULL 4 DELAYSEC 5 SYRINGE PULL 2 Air # air out puck DELAYSEC 10 VALVEPORT M3Bypass SET # relieve vacuum DELAYSEC 2 SYRINGE PUSH ALL ToxicWaste # puck cleared SYRINGE PULL 2 Fixative DELAYSEC 5 VALVEPORT Fixative HOME SYRINGEJOG PULL 2.5 SYRINGE PUSH 1.5 Puck VALVEPORT Puck HOME SYRINGEJOG PUSH 2.25 # fill filter holder DELAYMIN 20 # 1st fix VALVEPORT Puck SET DELAYSEC 10 # RINSE FILTER WITH FIX VALVEPORT Puck HOME VALVEPORT Syringe HOME SYRINGEJOG PULL 3.5 DELAYSEC 5 SYRINGE PULL 3 Air # air out puck DELAYSEC 10 VALVEPORT M3Bypass SET # relieve vacuum DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste # puck cleared SYRINGE PULL 2 Fixative DELAYSEC 5 VALVEPORT Fixative HOME SYRINGEJOG PULL 2.5 SYRINGE PUSH 1.5 Puck VALVEPORT Puck HOME SYRINGEJOG PUSH 2.25 # fill filter holder DELAYMIN 20 # 2nd fix VALVEPORT Puck SET DELAYSEC 10 # RINSE FILTER WITH FIX VALVEPORT Puck HOME VALVEPORT Syringe HOME SYRINGEJOG PULL 3.5 DELAYSEC 5 SYRINGE PULL 3 Air DELAYSEC 10 VALVEPORT M3Bypass SET # relieve vacuum DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste # puck cleared SYRINGE PULL 2 Fixative DELAYSEC 5 VALVEPORT Fixative HOME SYRINGEJOG PULL 2.5 SYRINGE PUSH 1.5 Puck VALVEPORT Puck HOME SYRINGEJOG PUSH 2.25 # fill filter holder DELAYMIN 20 # 3rd fix; total fix t=60 min VALVEPORT Puck SET DELAYSEC 10 # RINSE FILTER WITH FIX VALVEPORT Puck HOME VALVEPORT Syringe HOME SYRINGEJOG PULL 3.5 DELAYSEC 5 SYRINGE PULL 3 Air DELAYSEC 10 VALVEPORT M3Bypass SET # relieve vacuum DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste # puck cleared SYRINGE PULL 5 Air # air out Puck x2 DELAYSEC 10 VALVEPORT M3Bypass SET # relieve vacuum in syringe DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 2 FlushM2 # rinse away Fix DELAYSEC 30 SYRINGE PULL 1 FlushM2 # rinse away Fix DELAYSEC 30 VALVEPORT FlushM2 HOME SYRINGEJOG PULL 4 SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 1.5 M2Bypass # pull in air to blow remaining fluid SYRINGE PUSH ALL ToxicWaste END