# ESP Delta 2002 # May 29,2003 # New Valve Manifolds # rev 001 # test tank # Sample archive for toxin analysis # mod 5/27/03 - temp not calibrated PROGRAM: TOXTST SYRINGE PULL 22.5 Sample DELAYSEC 20 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 20 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME #Done sampling total ~400 mL SYRINGEJOG PULL 2 #DEPR. EXAHUST VALVEPORT Waste HOME #ISOLATE WASTE LINE DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste VALVEPORT Syringe HOME #NOW DO SAMPLE SIDE SYRINGEJOG PULL 2 DELAYSEC 5 VALVEPORT Sample HOME # depress complete DELAYSEC 10 VALVEPORT M3Bypass SET # relieve vacuum DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste DELAYSEC 2 SYRINGE PULL 1.5 M3Bypass DELAYSEC 2 SYRINGE PUSH ALL ToxicWaste # burp M1 loop DELAYSEC 2 SYRINGE PULL 4 M3Bypass VALVEPORT M3Bypass HOME SYRINGEJOG PUSH 3 # INC PRESSURE IN PUCK SYRINGE PUSH ALL ToxicWaste DELAYSEC 2 VALVEPORT Syringe HOME SYRINGEJOG PULL 5 DELAYSEC 10 SYRINGE PULL 5 Air # PUCK DRY OUT COMPLETE DELAYSEC 20 VALVEPORT M3Bypass SET # relieve vacuum DELAYSEC 7 SYRINGE PUSH ALL ToxicWaste DELAYSEC 2 SYRINGE PULL 1.5 M3Bypass DELAYSEC 2 SYRINGE PUSH ALL ToxicWaste # burp M1 loop VALVEPORT Air SET # vent while heating HEAT 50 10 # Ck temp for 66C red puck DELAYMIN 13 # ck time delay END