# macro for creating sample lysate for sand hybe rxn used 6-9-03 # in situ ESP Delta unit - new valve manifold # lysis solun. pushed in from top; pressure push mix # rev 038 April 21 2003; mod 4/24 based on run notes # modified from rev 005/036 # the 5xSET rinse is with 0.005% IGEPAL # ~2 ml lysate # heating time 14 min # run time is ~26min PROGRAM: SSH1 SYRINGE PULL 5 Sample # prime puck DELAYSEC 10 SYRINGE PUSH ALL Waste SYRINGE PULL 22.5 Sample DELAYSEC 10 VALVEPORT Sample HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME #Done sampling SYRINGE PULL 4.0 SETrinse # 5x SET with 0.005% IGEPAL DELAYSEC 5 VALVEPORT SETrinse HOME SYRINGEJOG PULL 4 # clear manifold SYRINGE PUSH 7 Puck # clear syringe block mnfld DELAYSEC 10 VALVEPORT Puck HOME VALVEPORT Syringe HOME SYRINGEJOG PULL 7 DELAYSEC 10 SYRINGE PULL 3.0 Air DELAYSEC 10 VALVEPORT M3Bypass SET # relieve vacuum in syringe DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 5 Air # 2nd draw of air to dry out puck DELAYSEC 10 VALVEPORT M3Bypass SET # relieve vacuum in syringe DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 1.0 Lysis # premix lysis/sample buffer DELAYSEC 5 VALVEPORT Lysis HOME SYRINGEJOG PULL 3.5 # rinse syringe SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 1.5 M3Bypass SYRINGE PUSH ALL ToxicWaste # burp M1 loop SYRINGE PULL 2.25 Lysis # incl comp for backlash DELAYSEC 30 # increase from 5 to 30 sec VALVEPORT Lysis HOME SYRINGEJOG PULL 4.5 # should leave 2ml in syringe SYRINGE PUSH 1.5 Puck # prime top loop VALVEPORT Puck HOME # isolate puck SYRINGEJOG PUSH 4.7 # leaves syringe at .5 DELAYSEC 10 SYRINGEJOG PULL 2 #RELIEVE PRESSURE HEAT 57 14 # 85C DELAYMIN 2 SYRINGEJOG PUSH 2 DELAYSEC 10 SYRINGEJOG PULL 2 # PRESSURE PUSH DELAYMIN 3 SYRINGE PUSH 2 ToxicWaste # blow off SOME pressure VALVEPORT Syringe HOME SYRINGEJOG PULL 4 # draw vacuum; AT 4.5 DELAYSEC 10 SYRINGE PULL 2 Air # suck lysate in syringe; AT 6.5 DELAYSEC 15 VALVEPORT M3Bypass SET DELAYSEC 2 SYRINGE PUSH 1.5 Puck # recycle lysate 2nd pass VALVEPORT Puck HOME SYRINGEJOG PUSH 4.5 # syringe at .5 DELAYSEC 10 SYRINGEJOG PULL 2 DELAYMIN 3 SYRINGEJOG PUSH 2 # PRESSURE PUSH DELAYSEC 5 SYRINGEJOG PULL 2 DELAYMIN 3 SYRINGE PUSH 2 ToxicWaste # blow off SOME pressure VALVEPORT Syringe HOME SYRINGEJOG PULL 4 # draw vacuum; AT 4.5 DELAYSEC 10 SYRINGE PULL 2.5 Air # suck lysate in syringe; AT 7 DELAYSEC 20 VALVEPORT M3Bypass SET DELAYSEC 2 VALVEPORT Syringe SET # isolate lysate in syringe END