LOGFILE:/ata0/LOG/233_000.log
APPTITLE:ESP  VER:2.1 built Aug  5 2002 at 14:01:11
INSTRUMENT LOCATION:EMBEDDED ID:DELTA
15:10:22	Reading and parsing geno.cfg and geno.cal files
15:10:22	Initializing motion control
15:10:22	Initializing temperature control
15:10:22	Initializing valves and reagent maps
15:10:22	ERROR: Warning: No M3 ports are configured
15:10:22	Initializing camera
15:10:22	Initializing shuttle
15:10:24	Initializing syringe
15:10:25	Initializing carousel
15:10:27	Initializing filterclamp
15:10:29	Initializing load elevator
15:10:30	Initializing unload elevator
15:10:32	Initializing modem
15:10:32	Initializing com port 0
15:10:32	ERROR: Woken by external wakeup...now running in MANUAL MODE
15:10:32	ERROR: Check debug.log file for errors during last execution
15:10:32	Skipping homing macro during startup
15:10:33	ERROR: Water sense alarm 1385 [cnts]> 500 [cnts] max...beginning system shutdown
15:10:33	Application shutdown requested
15:10:33	Instrument NOT in SAFE state for shutdown