# lysate filtration # rev 003; Sept 12 2002 # Starts with ~2ml 5um lysate from sh1_28 macro # passes lysate from SH1 through 0.65um filter # Syringe begins @ pos 7 mL # valve config 001 PROGRAM: LYFIL HEAT 32 2 SYRINGE PUSH 1.5 Puck VALVEPORT Puck HOME SYRINGEJOG PUSH 4.5 #LEAVES SYRINGE AT 1ml DELAYSEC 30 VALVEPORT Syringe HOME SYRINGEJOG PULL 5 DELAYSEC 30 VALVEPORT Syringe SET DELAYSEC 2 VALVEPORT M3Bypass SET DELAYSEC 2 SYRINGE PULL 1 Air DELAYSEC 30 END