# This is the sh2vx.txt macros used in 102_002 modified to increase wash time # after the lysate incubation # will run with shrt1v.txt and qualtest # Brent 5/1/01 -- increased heat duration from 45 to 105 min # PROGRAM: SH2 HEAT 20.35 105 # puck T=~25C; check t delay=process t DELAYMIN 1 # pre-warm array puck SYRINGE PUSH 0.9 AirUpper # pushing in lysate from SH1 incrementally DELAYSEC 5 SYRINGE PUSH 0.9 AirUpper # total push = 1.7ml - check this and adj. DELAYMIN 4 SYRINGE PULL 2.O AirUpper SYRINGE PUSH 1.8 AirUpper DELAYMIN 4 # 2nd lysate push SYRINGE PULL 2.O AirUpper SYRINGE PUSH 1.8 AirUpper DELAYMIN 4 # 3rd lysate push; total hybe t=12 min SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 5.0 SHWash # assay wash, vol comp for draw down DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 3.5 AirUpper SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 2.5 SHProbe1 # check vol to make sure filter floods DELAYMIN 2.0 # 1st sig probe incubation SYRINGE PULL 0.25 SHProbe1 SYRINGE PULL 0.25 SHProbe1 DELAYMIN 2.0 # 2nd sig probe push SYRINGE PULL 0.25 SHProbe1 SYRINGE PULL 0.25 SHProbe1 DELAYMIN 2.0 # 3rd sig probe push SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 5 SHWash # assay wash, vol comp for draw down DELAYSEC 180 SYRINGE PULL 2.5 SHWash # assay wash, vol comp for draw down DELAYSEC 90 SYRINGE PULL 2.5 SHWash # assay wash, vol comp for draw down DELAYSEC 90 SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 3.5 AirUpper SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 2.5 Conjugate # check vol to make sure filter floods DELAYMIN 2.0 # 1st con cycle SYRINGE PULL 0.5 Conjugate # check vol to make sure filter floods DELAYMIN 3.0 # 2nd con push SYRINGE PULL 0.5 Conjugate # check vol to make sure filter floods DELAYMIN 2.0 SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 5 SHWash # assay wash, vol comp for draw down DELAYSEC 180 SYRINGE PULL 2.5 SHWash # assay wash, vol comp for draw down DELAYSEC 120 SYRINGE PULL 2.5 SHWash # assay wash, vol comp for draw down DELAYSEC 120 SYRINGE PULL 2.5 SHWash # assay wash, vol comp for draw down DELAYSEC 120 SYRINGE PULL 2.5 SHWash # assay wash, vol comp for draw down DELAYSEC 120 CAMERAPOWER ON # START CAMERA COOLING HEAT 20.35 1 # turn off heat in one minute SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 3.5 AirUpper SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 2.5 FlushLower SYRINGE PULL 2.0 AirLower SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 1.5 Air VALVEPORT Air CLOSE SYRINGE PUSH ALL ToxicWaste # clear syringe/loop of flush SYRINGE PULL 1.25 Sub1 SYRINGE PULL 1.25 Sub2 SYRINGE PULL 4.0 AirLower # mix sub 1&2 SYRINGE PUSH 2.5 AirUpper # 1st sub push SYRINGE PULL 2.5 AirUpper SYRINGE PUSH 2.5 AirUpper # 2nd sub push SYRINGE PULL 2.5 AirUpper ACQUIREIMAGE LIGHT, 2X2, 120000, "SH TEST 4" ACQUIREIMAGE LIGHT, 2X2, AUTO, "SH TEST 5" CAMERAPOWER OFF #SHUT CAMERA DOWN SYRINGE PUSH ALL ToxicWaste END