# macro for creating sample lysate for sand hybe rxn # in situ ESP Delta unit - new valve manifold # lysis solun. pushed in from top # rev 30 Oct 1, 2002 # modified Sept 30 2002 for full vol sample # full vol sample= 12 x 22.5ml # the 5xSET 0.01% IGEPAL rinse is included # ~2 ml lysate with swish during heating # heating time 15 min PROGRAM: SH1 SYRINGE PULL 22.5 Sample DELAYSEC 10 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 10 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 10 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 20 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 20 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 30 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 30 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 30 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 45 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 45 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 60 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample # sample= 12 x 22.5ml= 270ml DELAYSEC 60 SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME #Done sampling SYRINGEJOG PULL 3 #DEPR. EXAHUST VALVEPORT Waste HOME #ISOLATE WASTE LINE DELAYSEC 5 SYRINGE PUSH ALL Waste VALVEPORT Syringe HOME #NOW DO SAMPLE SIDE SYRINGEJOG PULL 3 DELAYSEC 30 VALVEPORT Air SET SYRINGEJOG PULL 5 DELAYSEC 5 VALVEPORT M3Bypass SET # relieve vacuum in syringe DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste #done depressurizing VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME #Done sampling SYRINGE PULL 4.0 SETrinse # 5x SET with 0.01% IGEPAL DELAYSEC 5 VALVEPORT SETrinse HOME SYRINGEJOG PULL 3 # clear manifold SYRINGE PUSH 6 Puck # clear syringe block mnfld DELAYSEC 10 VALVEPORT Puck HOME VALVEPORT Syringe HOME SYRINGEJOG PULL 7 DELAYSEC 10 SYRINGE PULL 3.0 Air DELAYSEC 10 VALVEPORT M3Bypass SET # relieve vacuum in syringe DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 5 Air # 2nd draw of air to dry out puck DELAYSEC 10 VALVEPORT M3Bypass SET # relieve vacuum in syringe DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 1.0 Lysis # premix lysis/sample buffer DELAYSEC 5 VALVEPORT Lysis HOME SYRINGEJOG PULL 3.5 # rinse syringe SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 1.5 M3Bypass SYRINGE PUSH ALL ToxicWaste # burp M1 loop HEAT 62.4 14 # 85C; check time SYRINGE PULL 2.25 Lysis DELAYSEC 5 VALVEPORT Lysis HOME SYRINGEJOG PULL 1.75 # clear mnfld DELAYSEC 2 SYRINGE PUSH 1.25 Puck VALVEPORT Puck HOME SYRINGEJOG PUSH ALL DELAYSEC 15 VALVEPORT Syringe HOME SYRINGEJOG PULL 0.25 DELAYSEC 3 SYRINGEJOG PULL 0.25 DELAYSEC 3 SYRINGEJOG PULL 0.4 DELAYSEC 3 SYRINGEJOG PUSH 0.2 DELAYSEC 3 SYRINGEJOG PUSH 0.2 DELAYMIN 3 # 1st lysis swish SYRINGEJOG PULL 0.4 DELAYSEC 3 SYRINGEJOG PUSH 0.2 DELAYSEC 3 SYRINGEJOG PUSH 0.2 DELAYMIN 3 # 2nd lysis swish SYRINGEJOG PULL 0.4 DELAYSEC 3 SYRINGEJOG PUSH 0.2 DELAYSEC 3 SYRINGEJOG PUSH 0.2 DELAYMIN 3 # 3rd lysis swish SYRINGEJOG PULL 0.4 DELAYSEC 3 SYRINGEJOG PUSH 0.2 DELAYSEC 3 SYRINGEJOG PUSH 0.2 DELAYMIN 2 # 4th lysis swish cook t= 11 min SYRINGEJOG PULL 4.5 # syringe @ 5ml DELAYSEC 10 VALVEPORT Syringe SET # ISOLATGE LYSATE IN SYRINGE VALVEPORT M3Bypass SET # RELIEVE VACUUM DELAYSEC 2 SYRINGE PUSH 1 Puck # recycle lysate 2nd pass; syringe @ 4 VALVEPORT Puck HOME SYRINGEJOG PUSH 3.5 # LEAVES SYRINGE AT 0.5ml DELAYSEC 10 VALVEPORT Syringe HOME SYRINGEJOG PULL 4.5 # draw vacuum; SYRNGE @5 DELAYSEC 5 VALVEPORT Syringe SET # ISOLATGE LYSATE IN SYRINGE VALVEPORT M3Bypass SET # RELIEVE VACUUM DELAYSEC 2 SYRINGE PULL 2 Air # suck in lysate; syringe @7 ml DELAYSEC 2O VALVEPORT M3Bypass SET # to relieve vacuum in syringe DELAYSEC 3 VALVEPORT M3Bypass HOME VALVEPORT Air HOME END