# macro for creating sample lysate for sand hybe rxn # in situ ESP Delta unit - new valve manifold # lysis solun. pushed in from top # rev 015 July 20, 2002 # modified from GOM macro for concnetrated sample # includes prime step for sample # the 5xSET 0.01% IGEPAL rinse is included # ~1.5 ml lysate # heating time 14 min no delays between steps # Modified to load filter make lysate and then push # out M3 for collection. PROGRAM: SH1 LOADFILTER 6 SYRINGE PULL 5 Sample DELAYSEC 15 SYRINGE PUSH ALL Waste # prime puck SYRINGE PULL 22.5 Sample DELAYSEC 10 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME #Done sampling SYRINGEJOG PULL 2 #DEPR. EXAHUST VALVEPORT Waste HOME #ISOLATE WASTE LINE VALVEPORT Syringe HOME #NOW DO SAMPLE SIDE SYRINGEJOG PULL 2 VALVEPORT Air SET SYRINGE PUSH ALL ToxicWaste #done depressurizing SYRINGE PULL 1.5 M3Bypass SYRINGE PUSH ALL ToxicWaste # burp fluid in M1 loop SYRINGE PULL 4.0 SETrinse # 5x SET with 0.01% IGEPAL VALVEPORT SETrinse HOME SYRINGEJOG PULL 3 # clear manifold SYRINGE PUSH ALL Puck # clear syringe block mnfld DELAYSEC 5 SYRINGE PULL 2 M2Bypass SYRINGE PULL 5.0 Air DELAYSEC 5 VALVEPORT M3Bypass SET # relieve vacuum in syringe SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 5 Air # 2nd draw of air to dry out puck DELAYSEC 5 VALVEPORT M3Bypass SET # relieve vacuum in syringe SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 1.0 Lysis # premix lysis/sample buffer VALVEPORT Lysis HOME SYRINGEJOG PULL 3.5 # rinse syringe SYRINGE PUSH ALL ToxicWaste SYRINGE PULL 1.5 M3Bypass SYRINGE PUSH ALL ToxicWaste # burp M1 loop HEAT 62.4 14 # 85C; check time SYRINGE PULL 1.7 Lysis # no comp for backlash VALVEPORT Lysis HOME SYRINGEJOG PULL 2 # should leave 1.5ml in syringe SYRINGE PUSH 1.25 Puck # prime top loop VALVEPORT Puck HOME # isolate puck SYRINGEJOG PUSH ALL # puck filled with lysis buffer SYRINGEJOG PULL 2.45 # relieve pressure DELAYMIN 3 SYRINGEJOG PUSH 2 SYRINGEJOG PULL 2 DELAYMIN 3 SYRINGEJOG PUSH 2 SYRINGEJOG PULL 2 DELAYMIN 3 SYRINGEJOG PUSH 2 SYRINGEJOG PULL 2 DELAYMIN 2 # lysate cook t= 11min SYRINGEJOG PUSH ALL VALVEPORT Syringe HOME SYRINGEJOG PULL 2.5 # draw vacuum SYRINGE PULL 2.5 Air # suck lysate in syringe DELAYSEC 5 SYRINGE PUSH 1.5 Puck # recycle lysate 2nd pass VALVEPORT Puck HOME SYRINGEJOG PUSH 3 VALVEPORT Syringe HOME SYRINGEJOG PULL 2.5 # draw vacuum SYRINGE PULL 4 Air # suck lysate in syringe; @5 ml DELAYSEC 5 VALVEPORT M3Bypass SET # to relieve vacuum in syringe VALVEPORT Syringe SET # isolate lysate in syringe SYRINGEJOG PUSH ALL #PUSH TO M3 TO COLLECT LYSATE #VALVEPORT M3Bypass HOME UNLOADFILTER END