#ESP Delta 2002 # August 19, 2002 #new Valve manifolds #rev 011 #dry filter between fix exchanges #~67 ml sample #depressurization stroke 2ml #fills puck with fix after dry-out #fix delays at 20 min EACH = 40 min total fix time PROGRAM: WCR SYRINGE PULL 22.5 Sample DELAYSEC 10 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 10 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME SYRINGE PULL 22.5 Sample DELAYSEC 10 VALVEPORT Intake HOME SYRINGE PUSH ALL Waste VALVEPORT Exhaust HOME #Done sampling need to depressurize SYRINGEJOG PULL 3 #DEPR. EXAHUST VALVEPORT Waste HOME #ISOLATE WASTE LIN DELAYSEC 5 SYRINGE PUSH ALL Waste VALVEPORT Syringe HOME #NOW DO SAMPLE SIDE SYRINGEJOG PULL 3 DELAYSEC 15 VALVEPORT Air SET SYRINGE PUSH ALL ToxicWaste #done depressurizing SYRINGE PULL 2.5 Fixative DELAYSEC 5 VALVEPORT Fixative HOME SYRINGEJOG PULL 2 SYRINGE PUSH ALL Puck # RINSE FILTER WITH FIX SYRINGE PULL 3 Air # air out puck DELAYSEC 5 VALVEPORT M3Bypass SET # relieve vacuum DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste # puck cleared SYRINGE PULL 2.5 Fixative DELAYSEC 5 VALVEPORT Fixative HOME SYRINGEJOG PULL 2 SYRINGE PUSH 1.5 Puck VALVEPORT Puck HOME SYRINGEJOG PUSH 2.25 # fill filter holder DELAYMIN 1 # 1st fix SYRINGE PUSH ALL Puck SYRINGE PULL 3 Air # air out puck DELAYSEC 5 VALVEPORT M3Bypass SET # relieve vacuum DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste # puck cleared SYRINGE PULL 2.5 Fixative DELAYSEC 5 VALVEPORT Fixative HOME SYRINGEJOG PULL 2 SYRINGE PUSH 1.5 Puck VALVEPORT Puck HOME SYRINGEJOG PUSH 2.25 # fill filter holder DELAYMIN 1 # 2nd fix; total fix t=40 min SYRINGE PUSH ALL Puck SYRINGE PULL 5 Air # air out puck DELAYSEC 10 VALVEPORT M3Bypass SET # relieve vacuum DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste # puck cleared SYRINGE PULL 5 Air # air out Puck x2 DELAYSEC 10 VALVEPORT M3Bypass SET # relieve vacuum in syringe DELAYSEC 5 SYRINGE PUSH ALL ToxicWaste END