Automated sensing of soil water content with capacitance methods is common due to the relative ease of installation and
monitoring at multiple sites. The resonant frequency of an inductance-capacitance circuit is a function of the dielectric
permittivity of the material surrounding the ring-capacitor sensor. However, limited relationships between resonant
frequency and permittivity in the soil water range have been reported. Furthermore, sensor readings and apparent water
contents have been shown to vary with temperature in laboratory and field studies in the opposite direction of that expected
for free water. We designed laboratory equipment and experiments to improve our fundamental knowledge about the behavior and
characterization of such a capacitance sensor (Sentek EnviroSMARTTM). Four sensors are used for resonant frequency
readings, while a fifth measures total capacitance directly on an HP Network Analyzer. A solvent-resistant container was
designed for water-dioxane mixtures, which provide a complete range of permittivity values from 2.2 to over 80. The sensor
readings are sensitive to permittivity changes in the range expected for soils from very dry to fully saturated. Variability
between four sensors is reduced by normalization to readings in air and water. Frequency-based permittivity estimates in free
water decreased linearly with temperature, as expected, substantiating the field-measured temperature dependence. Next, an
expanding metal cylinder was used to interfere with the electrical field in air, water and two dioxane-water mixtures. The
change in normalized readings with distance to the metal boundary is approximated by a negative exponential function with a
characteristic length of 11 mm. The laboratory results are confirmed with numerical experiments assuming axisymmetric
materials. Using the improved capacitance sensor characterization, water content can be estimated directly from permittivity
using a universal calibration, and there is now a stronger basis for addressing the temperature-dependence of measurements
in soil-air-water systems.