HR: 0800h
AN: MR11A-0888 [Abstracts]
TI: A New Laser Heating and Temperature Measurement System
AU: * Shahar, A
EM: ashahar@ess.ucla.edu
AF: University of California, Los Angeles, Department of Earth and Space Sciences, Los Angeles, CA 90095
AU: Kavner, A
EM: akavner@igpp.ucla.edu
AF: University of California, Los Angeles, Department of Earth and Space Sciences, Los Angeles, CA 90095
AB:
The laser-heated diamond anvil cell enables measurements of material properties in situ at high pressures and temperatures.
An ideal measurement consists of a sample subjected to uniform pressure throughout the sample chamber. However, due to the
small size of the samples and the high thermal conductivity of the diamond anvils temperature gradients are always present.
Therefore, the gradient must be measured to be able to interpret results from the laser heated diamond cell, and to evaluate
methods employed to reduce temperature gradients. In order to characterize these gradients, we have designed and installed a
new laser heating and temperature measurement system at UCLA. This system specializes in a precise, quantitative
measurement of the radial temperature gradient. Our design is anchored by a 20 Watt diode-pumped Nd:YLF laser for heating
and measures temperature by collecting spectral intensity information from the hotspot. The temperature and the associated
gradient are measured simultaneously using two cameras: 1) a high speed CCD camera coupled with a 150 mm imaging
spectroradiometer that measures temperature, 2) a high dynamic range CCD camera that measures two-dimensional intensity
distribution, at a variety of chosen spectral ranges. Information from both cameras is combined to provide the
two-dimensional temperature distribution of the hotspot. The spectral intensity of a Gaussian-shaped hotspot has a maximum in
the center part of the hotspot, and the sample temperature and therefore spectral intensity decreases radially from the
center as proscribed by the thermal properties of the experiment and Planck's law. The intensity is a much more sensitive
indication of temperature change. For example, in the case of a hotspot with a 2000K peak temperature, a 15% decrease in
intensity at a given wavelength (e.g. 500 nm) corresponds to a temperature drop of 20 K, a 1% decrease, well below most
estimates of precision in temperature measurement of laser heated spots. Therefore, relative spectral intensity provides a
more reliable estimate of two-dimensional temperature gradients in high P, T experiments. In addition, this method bypasses
most additional errors induced by chromatic aberration (a wavelength dependent focusing effect) of the objective lens. As a
test of the ability of this system to measure temperature gradients, we have measured the thermal diffusivity anisotropy of
highly oriented graphite. A graphite sample oriented perpendicular to the basal plane will create an elliptical hotspot from
a radially symmetric laser power input due to faster heat transport along the basal plane (D~1.10x10-3 m2/sec) compared with
the c direction (D~ 4.6x10-6 m2/sec). Therefore, the ellipticity of the hotspot is a direct measure of the systems ability
to measure thermal diffusivity. Side by side comparison of the high dynamic range camera and the imaging spectroradiometer
shows that the high dynamic range attains better measurements of the temperature gradients of a graphite hotspot. For
example, at 75% of the peak temperature, the high dynamic range camera measured an ellipticity of 2.8, while the imaging
spectroradiometer was smeared out to 2.2. We will present preliminary results from the laser-heating system including
calibration information new data on the high P,T behavior of a metals, silicates, oxides, and ices.
DE: 3994 Instruments and techniques
SC: Mineral and Rock Physics [MR]
MN: 2004 AGU Fall Meeting