HR: 1340h
AN: P53A-0998 [Abstracts]
TI: CheMin: A Definitive Mineralogy Instrument on the Mars Science Laboratory (MSL '09) Rover
AU: * Vaniman, D
EM: dvaniman@lanl.gov
AF: Los Alamos National Laboratory, MS D469, Hydrology, Geochemistry and Geology, Los
Alamos, NM 87545, United States
AU: Blake, D
EM: dblake@mail.arc.nasa.gov
AF: NASA Ames Research Center, MS 239-4, Moffett Field, CA 94035, United States
AU: Sarrazin, P
AF: In Xitu, Inc., 2551 Casey Ave., Suite A, Mountain View, CA 94043, United States
AU: Bish, D
AF: Indiana University, Departent of Geological Sciences, Bloomington, IN 47405, United
States
AU: Chipera, S
AF: Chesapeake Energy Corporation, P.O. Box 18496, Oklahoma City, OK 73154, United
States
AU: Ming, D
AF: NASA Johnson Space Center, 2101 NASA Parkway, Houston, TX 77058, United States
AU: Morris, D
AF: NASA Johnson Space Center, 2101 NASA Parkway, Houston, TX 77058, United States
AU: Yen, A
AF: California Institute of Technology, Jet Propulsion Laboratory, Pasadena, CA 91109, United
States
AB:
CheMin is a miniature X-ray diffraction / X-ray fluorescence (XRD/XRF) instrument that has been chosen for the
analytical laboratory of MSL. CheMin utilizes a microfocus source cobalt X-ray tube, a transmission sample cell
and an energy-discriminating X-ray sensitive CCD to produce simultaneous 2-D X-ray diffraction patterns and X-
ray fluorescence spectra from powdered or crushed samples.
The X-ray source utilizes a conventional tungsten cathode and a reflection type cobalt anode. A focusing grid in
the tube yields a 50 micron diameter photon source on the anode. The CCD detector is a 600X600 front
illuminated frame transfer device having 40 micron square pixels and a deep depletion zone for high QE of 7 KeV
X-rays (CoKa). Diffraction data are collected over the range 5-55 degrees 2-theta, with a resolution of 0.32
degrees 2-theta.
During the mission, MSL's sample delivery system will provide up to 74 crushed and sieved samples to the
CheMin instrument. Samples are delivered through a funnel to one of 26 reusable transmission sample cells on
the CheMin sample wheel. Samples are loaded at the top, analyzed, then discarded in a sump by rotating the
sample cell to the bottom of the wheel. Samples are analyzed for up to 10 hours during multiple sols. During
analysis, sample cells are shaken at sonic frequencies by a piezoelectric actuator that causes convection and
granular flow of the powdered material. The result is that all components of the sample are passed through
CheMin's ~50 micron diameter beam in random orientations during the course of an
analysis. In addition to the 26 reuseable sample cells, CheMin has 6 sealed cells containing XRF and XRD
reference standards. These standards can be analyzed at any time to evaluate the health of the instrument or to
generate calibrated XRD patterns.
For each analysis, CheMin will produce energy-selected 2-D CoKa and CoKb patterns as well as a histogram of
X-ray energy vs. number of photons. Diffraction data will be quantified using Rietveld refinement and
Fluorescence data will be quantified using a fundamental parameters technique. Downlinked data from Mars will
be validated in three ways: First, on-board standards will be used to calibrate unknowns and to control for non-
optimal analysis conditions during the lifetime of the mission. Second, laboratory-based CheMin instruments
will be used to analyze Mars analog materials. Third, a ray-tracing model has been developed that can simulate
the CheMin geometry and corroborate diffraction results from specific sample mineralogies.
DE: 3672 Planetary mineralogy and petrology (5410)
DE: 3694 Instruments and techniques
DE: 5220 Hydrothermal systems and weathering on other planets
DE: 5494 Instruments and techniques
DE: 6225 Mars
SC: Planetary Sciences [P]
MN: 2007 Fall Meeting