TI: High-Precision Double-Spike Sr Isotope Measurements: Applications to Geochemistry and Cosmochemistry
AU: * Parkinson, I J
EM: I.J.Parkinson@open.ac.uk
AF: The Open University, Department of Earth Sciences, Walton Hall, Milton Keynes, MK7 6AA,
United Kingdom
AU: Burton, K W
EM: K.W.Burton@open.ac.uk
AF: The Open University, Department of Earth Sciences, Walton Hall, Milton Keynes, MK7 6AA,
United Kingdom
AU: Burton, K W
EM: K.W.Burton@open.ac.uk
AF: Université de Toulouse, LMTG, Toulouse, F-31400, France
AU: Mokadem, F
EM: F.Mokadem@open.ac.uk
AF: The Open University, Department of Earth Sciences, Walton Hall, Milton Keynes, MK7 6AA,
United Kingdom
AU: Charlier, B L
EM: B.L.A.Charlier@open.ac.uk
AF: The Open University, Department of Earth Sciences, Walton Hall, Milton Keynes, MK7 6AA,
United Kingdom
AB:
Strontium isotope systematics consist of the stable 88Sr/86Sr ratio, the radiogenic 87Sr/86Sr
ratio and the 84Sr/86Sr ratio which is stable in terrestrial materials and may vary due to nuclear-
synthetic processes in some meteorites [1]. Growing interest in natural mass dependent 88Sr/86Sr
variations indicate that terrestrial samples vary by ~500 ppm and have been measured to a precision of 50-70
ppm by MC-ICP-MS [2, 3]. However, the precision is large relative to the likely variation in many terrestrial
systems. Furthermore, 84Sr/86Sr ratios are poorly determined due to spectral interferences and
therefore MC-ICP-MS techniques are not appropriate for high-precision analyses of all the Sr isotope ratios.
Recent advances in TIMS allow 87Sr/86Sr and 84Sr/86Sr ratios to be measured to an external
precision of ~5 and 60 ppm respectively with instrumental fractionation being corrected by internal normalisation
to a constant 88Sr/86Sr ratio. Following on from the pioneering work of [4] we use an 84Sr-
87Sr double-spike coupled with TIMS analyses. Optimal spiking, extended ion collection times and a 25-fold
improvement in the determination of the spike composition compared to [4] allow us to measure
87Sr/86Sr and 88Sr/86Sr ratios to better than 10 ppm external precision. The double-spike
determination consists of an un-spiked and spiked measurement, with spiking prior to chemical separation
because column separation produces a consistent and resolvable light fractionation effect of 21 to 93 ppm. The
un-spiked run allows us to assess any non-mass dependent fractionation effects. Double-spike deconvolution is
performed in 87Sr denominator space. Our preliminary high-precision data allows us to resolve small (25
ppm) mass-dependent shifts in seawater 88Sr/86Sr, which may reflect short-timescale Sr fluxes in the
oceans.
References
[1] D.A. Papanastassiou, G.J. Wasserburg, Geophys. Res. Lett. 5 (1978) 361-376.
[2] J. Fietzke, A. Eisenhauer, Geochem. Geophys. Geosyst., 7, (2006) Q08009, doi:10.1029/2006GC001243.
[3] G.M. Nowell, et al., Geochim. Cosmochim. Acta. 71 (2007) A725.
[4] P.J. Patchett, Nature, 283 (1980b) 438-441.
DE: 8410 Geochemical modeling (1009, 3610)
DE: 8494 Instruments and techniques
SC: Volcanology, Geochemistry, Petrology [V]
MN: 2007 Fall Meeting