Troubleshooting

Connecting

I can't find the FIDO's WiFi network.
The controller may still be starting. Wait a few minutes. Check the FIDO has power and the antenna is attached. See Power and Enclosure.
I'm on the FIDO's WiFi but the page won't load.
Make sure you typed https://10.42.0.1. Your device may have switched back to cellular because the hotspot "has no internet". Turn off cellular data, or tell the device to stay on this network. Accept the browser's security warning.
I can't sign in to the cloud.
Check your email address. Ask your organization's administrator to reset your password.
My FIDO isn't on the cloud Devices list.
Check that the right Organization is selected. If it still isn't there, the FIDO hasn't been adopted. See Remote Management.

Device status

Connectivity shows Offline.
The controller hasn't reached the cloud for over 5 minutes. It keeps running its schedule. On site, check Uplink Internet on the Management page for cell signal or WiFi.
System Status shows ERROR.
Read the latest lines on the Events page, fix the cause, and use Make Sampler Safe. See When Something Goes Wrong.
Scheduled samples aren't running.
Check Scheduled Events → Enabled on the Management page and that Next Scheduled shows a time. Check the event's End After date hasn't passed. Check there are stocked pucks.
Times look wrong.
Check Device Timezone on the Management page. If the sampler's own times in the logs are wrong, use Sync FIDO Clock on the Maintenance page.
Number of Stocked Pucks is wrong.
Run Restock Sampler with only Count pucks ticked.

Puck Scan Station

Puck Scan isn't in the side menu.
Check that you're connected to the Puck Scan Station, not to a FIDO. Puck Scan never appears on a FIDO. On the station, check that the reader is plugged in and wait a moment for the scanning software to start.
A puck "didn't scan as valid data".
Remove it and place it again, centered on the reader. If it still fails, set it aside and note its number.
Export to Thumbdrive doesn't find my drive.
The drive's name (volume label) must be exactly FIDO_DATA.

Getting help

When you contact support, include:

  • The device name and the READI-Net version (bottom of the side menu)
  • What you were doing and what you saw
  • A download from Download Logs on the Events page

To confirm

Who users should contact for support, and how.