Schedule Sampling¶
FIDO can sample on its own on a schedule you set. The schedule is stored on the controller, so it keeps running with no internet connection.
Add a scheduled event¶
- Open the Schedule page.
- Click the calendar at the day and time the first sample should run.
- In the dialog, choose an Action:
- Sample Only
- Sample, then Decontaminate
- Decontaminate only
- Check the Start date and time.
- Choose how it Repeats:
- None: once only
- Daily: every N days
- Weekly: on the days of the week you tick
- Monthly: on a chosen day of the month (1st – 28th)
- For a repeating event, set End After to the last date it should run.
- Click Save.
Scheduled samples use the Default Volume set on the Management page.
Sample, then Decontaminate
Unless you have a reason not to, use Sample, then Decontaminate. Cleaning the fluid path after each sample keeps one sample's DNA out of the next.
To confirm
The recommended default action and spacing between samples, and how long a sample plus decontamination takes (so users don't schedule events that overlap).
Turn scheduling on¶
Events on the calendar run only while scheduling is enabled.
- Open the Management page.
- Under Sample Management, tick Scheduled Events → Enabled.
- Click Save Changes.
- Check Next Scheduled under the checkbox shows the time you expect.
Change or delete an event¶
Click the event on the calendar. Change it and click Save, or click Delete. Past events are shown on the calendar for reference and can't be edited.
Pausing the schedule¶
To stop all scheduled sampling without deleting anything, for example while you service the sampler, untick Scheduled Events → Enabled on the Management page and click Save Changes. Remember to turn it back on when you're done.
Checking the schedule¶
- Upcoming Events on the Device page lists what's coming up and how long until it runs.
- The Number of Stocked Pucks must cover the samples scheduled before your next visit.